Quality Control in the PacBio HiFi Sequencing Workflow
Contenido
Overviewp. 2
Overview of HiFi sequencing QC goals and workflow
Procedurepp. 3-4
Sample preparation and measurement conditions for QC assays
Qualitypp. 5-6
QC results including sheared DNA sizing data
Procedurepp. 7-8
Fragment Analyzer runs and sizing evaluation
p. 9
Summary of QC findings and recommendations
p. 3table, 8 rows
Table 1. Analytical specifications of the Agilent automated electrophoresis systems and their respective kits for high molecular weight DNA quality control.
System
Femto Pulse
Fragment Analyzer
TapeStation
Kit name
Genomic DNA 165 kb kit (p/n FP-1002-0275)
HS Large Fragment Analysis kit (p/n DNF-464-0500)
Genomic DNA ScreenTape (p/n 5067-5365) and reagents (p/n 5067-5366)
The Femto Pulse is the only automated pulsed-field capillary electrophoresis system available, so it is the required choice for sizing high molecular weight input gDNA up to 165 kb.
Determines whether the lab must purchase a Femto Pulse instead of a constant-field system for HMW gDNA QC.
Source
“the Femto Pulse system is the only automated pulsed-field capillary electrophoresis (PFCE) system currently available”
Selection
Fragment Analyzer and TapeStation are constant-field systems limited to 60 kb, so they cannot assess input gDNA whose smear extends beyond that range.
Prevents selecting a constant-field instrument for the input gDNA QC step where HMW material exceeds 60 kb.
Source
“The Fragment Analyzer and TapeStation systems are both automated constant field electrophoresis systems that offer high throughput and are compatible with various assays that can size DNA up to 60 kb.”
Performance
The Femto Pulse requires only picogram amounts of DNA, which matters when input gDNA is scarce.
Allows QC of precious or low-yield gDNA samples without consuming significant material.
Source
“requiring only picogram amounts of DNA”
Operation
Smear analysis regions must be set between the lower and upper markers (or last ladder fragment) for valid average sizing; extending beyond the ladder relies on extrapolation.
Defines correct software setup for smear analysis to avoid inaccurate sizing results.
Source
“For best practice, this region should be set between the lower and upper markers (or the last ladder fragment for kits that do not use an upper marker)”
Performance
On the Fragment Analyzer, setting the smear upper limit above 48.5 kb limits resolution and the size is extrapolated by a linear fit, so results above the ladder are less reliable.
Warns engineers that Fragment Analyzer sizing above the ladder range is extrapolated and should not be trusted for HMW decisions.
Source
“When the upper limit of the smear range is set at a larger size then 48.5 kb, the separation resolution is limited and the size of the smear is extrapolated by the ProSize data analysis software using a linear line of best fit generated by the ladder.”
Performance
TapeStation software will not include any portion of the smear beyond 60 kb, so samples larger than 60 kb are truncated in the analysis.
Prevents using TapeStation for samples whose smear extends beyond 60 kb, as the reported size would be truncated.
Source
“Beyond 60 kb, the separation resolution is limited, and the TapeStation software will not include any portion of the smear that extends beyond this size range.”
Selection
PacBio recommends input gDNA meet GQN10kb ≥ 7.0 and GQN30kb ≥ 5.0 for whole genome sequencing, which is a pass/fail criterion for sample acceptance.
Provides the acceptance threshold used to decide whether gDNA proceeds to shearing and sequencing.
Source
“PacBio recommends that the input gDNA should have a GQN of at least 7.0 or higher at 10 kb (GQN10kb ≥ 7.0) and a GQN of 5.0 or higher at 30 kb (GQN30kb ≥ 5.0) for whole genome sequencing.”
Operation
For PacBio HiFi WGS, gDNA must be sheared to 10 to 25 kb, which is the target size window verified at the post-shearing QC step.
Sets the target size specification that the post-shearing QC must confirm before library preparation.
Source
“For PacBio HiFi WGS applications, the gDNA should be sheared to 10 to 25 kb.”
Selection
Any of the three Agilent systems can be used interchangeably for QC of sheared gDNA in the PacBio HiFi WGS workflow.
Allows selecting the lowest-cost or highest-throughput system for the sheared DNA QC step.
Source
“These results demonstrate that any of the Agilent automated electrophoresis systems can be used interchangeably for QC of sheared gDNA in the PacBio HiFi WGS library preparation workflow.”
Performance
The Femto Pulse gave sizing closest to sequencing results and the best precision (3 %CV) for sheared E. coli DNA, making it the optimal system for that example.
Supports choosing the Femto Pulse when sizing accuracy relative to actual read length is critical.
Source
“Sizing with the Femto Pulse was closest to the sequencing results, which showed an average read length of 19.0 kb. Additionally, the Femto Pulse system showed the best sizing precision of the three instruments, at 3 %CV.”
Selection
All three systems provided reliable sizing for the final library prior to sequencing, so any can be used for final library QC.
Confirms that final library QC does not require the Femto Pulse, enabling use of higher-throughput constant-field systems.
Source
“This example shows that any of the Agilent automated electrophoresis systems can provide reliable sizing analysis for QC of the library prior to sequencing.”
Selection
The Femto Pulse is the only automated system capable of sizing HMW input gDNA for HiFi WGS, while Fragment Analyzer and TapeStation suit smaller sheared gDNA and final libraries.
Summarizes the system-selection rule: Femto Pulse for input gDNA, constant-field systems for sheared DNA and libraries.
Source
“the Femto Pulse is the only automated electrophoresis system capable of sizing HMW input gDNA material for HiFi WGS, as well as assessing the sheared gDNA and the final library.”