Induztr
Resumen técnico Visor de documentos Modelos relacionados

Data Sheet Model BF 115

Technical Reference Summary

Covered Models & Document Scope

Incubadora estándar BINDER BF 115 (BF Avantgarde.Line, convección forzada), volumen interior 114 L. Dos variantes estándar: BF115-230V (Art.-No. 9010-0315, 230 V 1~ 50/60 Hz, enchufe CEE 7/7) y BF115UL-120V (Art.-No. 9010-0316, 120 V 1~ 60 Hz, enchufe NEMA 5-15). El documento también enumera accesorios y servicios opcionales.

Technical Data

Parameter Value Condición
Temperature range +8 °C above ambient temperature to 100 °C p.2 ↗ BF115-230V / BF115UL-120V
Temperature uniformity at 37°C 0.3 ± K p.2 ↗ BF115-230V / BF115UL-120V
Temperature uniformity at 100°C 1.6 ± K p.2 ↗ BF115-230V / BF115UL-120V
Temperature fluctuation at 37°C 0.1 ± K p.2 ↗ BF115-230V / BF115UL-120V
Temperature fluctuation at 100°C 0.2 ± K p.2 ↗ BF115-230V / BF115UL-120V
Heating up time to 100°C 65 min p.2 ↗ BF115-230V / BF115UL-120V
Heating up time to 37°C 8 min p.2 ↗ BF115-230V / BF115UL-120V
Recovery time after door was opened for 30 s at 37°C 3 min p.2 ↗ BF115-230V / BF115UL-120V
Rated Voltage 230 V p.2 ↗ BF115-230V
Rated Voltage 120 V p.2 ↗ BF115UL-120V
Power frequency 50/60 Hz p.2 ↗ BF115-230V
Power frequency 60 Hz p.2 ↗ BF115UL-120V
Nominal power 0,4 kW p.2 ↗ BF115-230V / BF115UL-120V
Unit fuse 6,3 A p.2 ↗ BF115-230V
Unit fuse 12,5 A p.2 ↗ BF115UL-120V
Phase (Nominal voltage) 1~ p.2 ↗ BF115-230V / BF115UL-120V
Interior volume 114 L p.2 ↗ BF115-230V / BF115UL-120V
Net weight of the unit (empty) 54 kg p.2 ↗ BF115-230V / BF115UL-120V
Load per rack 30 kg p.2 ↗ BF115-230V / BF115UL-120V
Permitted load 150 kg p.2 ↗ BF115-230V / BF115UL-120V
Wall clearance back 160 mm p.2 ↗ BF115-230V / BF115UL-120V
Wall clearance sidewise 100 mm p.2 ↗ BF115-230V / BF115UL-120V
Width net 710 mm p.2 ↗ BF115-230V / BF115UL-120V
Height net 725 mm p.2 ↗ BF115-230V / BF115UL-120V
Depth net 605 mm p.2 ↗ BF115-230V / BF115UL-120V

Operating Limits

  • Temperature range: +8 °C above ambient temperature to 100 °C p.2 ↗
  • Permitted load: 150 kg p.2 ↗
  • Load per rack: 30 kg p.2 ↗
  • Wall clearance back: 160 mm p.2 ↗
  • Wall clearance sidewise: 100 mm p.2 ↗

Type Code Decoder

Position Token Meaning
1-2 p.1 ↗ BF Series: Standard-Incubator with forced convection (BF Avantgarde.Line)
3-5 p.1 ↗ 115 Model / size (interior volume 114 L)
suffix p.1 ↗ UL UL variant for 120 V / 60 Hz supply (NEMA 5-15 plug)
suffix p.1 ↗ 230V 230 V 1~ 50/60 Hz variant (CEE 7/7 plug)
suffix p.1 ↗ 120V 120 V 1~ 60 Hz variant (NEMA 5-15 plug)

Ordering Data

Ordering Code Part No. Variant / Code Aplica a Note
BF115-230V p.1 ↗ 9010-0315 230 V 1~ 50/60 Hz, CEE 7/7 plug, Standard Standard model version
BF115UL-120V p.1 ↗ 9010-0316 120 V 1~ 60 Hz, NEMA 5-15 plug, Standard UL standard model version

Modelos relacionados