Data Sheet Model CB 56

Technical Reference Summary

Covered Models & Document Scope

Incubadoras de CO2 modelo CB 56 con esterilización por aire caliente y sensor de CO2 esterilizable por calor. Dos variantes estándar: CB056-230V (200...230 V, enchufe CEE 7/7, Art.-No. 9040-0197) y CB056UL-120V (100...120 V, enchufe NEMA 5-20, Art.-No. 9040-0198). Volumen interior 53 L.

Technical Data

Parameter Value Condición
Temperature range +4 °C above ambient temperature to 50 °C °C p.2 ↗ CB056-230V / CB056UL-120V
Temperature uniformity at 37 °C 0.3 ± K p.2 ↗ CB056-230V / CB056UL-120V
Temperature fluctuation at 37 °C 0.1 ± K p.2 ↗ CB056-230V / CB056UL-120V
Recovery time after door was opened for 30 s 5 min p.2 ↗ CB056-230V / CB056UL-120V
Humidity range 90 ...95 % RH p.2 ↗ CB056-230V / CB056UL-120V
CO2 range 0 ...20 Vol.-% CO2 p.2 ↗ CB056-230V / CB056UL-120V
CO2 measuring technology IR p.2 ↗ CB056-230V / CB056UL-120V
Rated voltage 200...230 V p.2 ↗ CB056-230V
Rated voltage 100...120 V p.2 ↗ CB056UL-120V
Power frequency 50/60 Hz p.2 ↗ CB056-230V / CB056UL-120V
Nominal power 0,8 kW p.2 ↗ CB056-230V / CB056UL-120V
Unit fuse 10,0 A p.2 ↗ CB056-230V
Unit fuse 16,0 A p.2 ↗ CB056UL-120V
Phase (nominal voltage) 1~ p.2 ↗ CB056-230V / CB056UL-120V
Interior volume 53 L p.2 ↗ CB056-230V / CB056UL-120V
Net weight of the unit (empty) 53 kg p.2 ↗ CB056-230V / CB056UL-120V
Load per rack 10 kg p.2 ↗ CB056-230V / CB056UL-120V
Permitted load 30 kg p.2 ↗ CB056-230V / CB056UL-120V
Wall clearance back 100 mm p.2 ↗ CB056-230V / CB056UL-120V
Wall clearance sidewise 50 mm p.2 ↗ CB056-230V / CB056UL-120V
Width net 528 mm p.2 ↗ CB056-230V / CB056UL-120V
Height net 674 mm p.3 ↗ CB056-230V / CB056UL-120V
Depth net 545 mm p.3 ↗ CB056-230V / CB056UL-120V
Sound-pressure level 41 dB(A) p.3 ↗ CB056-230V / CB056UL-120V
Energy consumption at 37 °C 45 Wh/h p.3 ↗ CB056-230V / CB056UL-120V

Operating Limits

  • Temperature range: +4 °C above ambient temperature to 50 °C p.2 ↗
  • Wall clearance back: 100 mm p.2 ↗
  • Wall clearance sidewise: 50 mm p.2 ↗
  • Permitted load: 30 kg p.2 ↗
  • Load per rack: 10 kg p.2 ↗

Type Code Decoder

Position Token Meaning
1-5 p.1 ↗ CB056 Model series CB, size 56
suffix p.1 ↗ UL UL variant with 100...120 V supply and NEMA 5-20 plug (absent = standard 200...230 V, CEE 7/7)
suffix p.1 ↗ 230V Mains voltage variant 200...230 V
suffix p.1 ↗ 120V Mains voltage variant 100...120 V

Ordering Data

Ordering Code Part No. Variant / Code Aplica a Note
CB056-230V p.1 ↗ 9040-0197 200...230 V 1~ 50/60 Hz, 10,0 A, CEE 7/7 plug, Standard Standard version
CB056UL-120V p.1 ↗ 9040-0198 100...120 V 1~ 50/60 Hz, 16,0 A, NEMA 5-20 plug, Standard UL standard version