Induztr
Resumen técnico Visor de documentos Modelos relacionados

Data Sheet Model CB S 260

Technical Reference Summary

Covered Models & Document Scope

Incubadoras de CO2 modelo CB-S 260 con esterilización por aire caliente. Dos variantes: CBS260-230V (Art.-No. 9040-0223, 200...230 V 1~ 50/60 Hz, enchufe CEE 7/7) y CBS260UL-120V (Art.-No. 9040-0224, 100...120 V 1~ 50/60 Hz, enchufe NEMA 5-20).

Technical Data

Parameter Value Condición
Interior volume 267 L p.1 ↗ CBS260-230V / CBS260UL-120V
Rated voltage 200...230 V p.2 ↗ CBS260-230V
Rated voltage 100...120 V p.2 ↗ CBS260UL-120V
Power frequency 50/60 Hz p.2 ↗ both variants
Phase (nominal voltage) 1~ p.2 ↗ both variants
Nominal power 1,5 kW p.2 ↗ both variants
Unit fuse 10,0 A p.2 ↗ CBS260-230V
Unit fuse 16,0 A p.2 ↗ CBS260UL-120V
Temperature range +6 °C above ambient temperature to 50 °C p.2 ↗ both variants
Temperature uniformity at 37°C 0.4 ± K p.2 ↗ both variants
Temperature fluctuation at 37°C 0.1 ± K p.2 ↗ both variants
Recovery time after door was opened 8 min p.2 ↗ both variants
Humidity range 90 ...95 % RH p.2 ↗ both variants
CO₂ range 0 ...20 Vol.-% CO₂ p.2 ↗ both variants
CO₂ measuring technology IR p.2 ↗ both variants
Net weight of the unit (empty) 116 kg p.2 ↗ both variants
Load per rack 10 kg p.2 ↗ both variants
Permitted load 40 kg p.2 ↗ both variants
Wall clearance back 100 mm p.2 ↗ both variants
Wall clearance sidewise 50 mm p.2 ↗ both variants
Width net 740 mm p.2 ↗ both variants
Height net 1,020 mm p.3 ↗ both variants
Depth net 785 mm p.3 ↗ both variants
Sound-pressure level 41 dB(A) p.3 ↗ both variants
Energy consumption at 37°C 55 Wh/h p.3 ↗ both variants

Operating Limits

  • Temperature range: +6 °C above ambient temperature to 50 °C p.2 ↗
  • Humidity range: 90 ...95 % RH p.2 ↗
  • CO₂ range: 0 ...20 Vol.-% CO₂ p.2 ↗
  • Permitted load: 40 kg p.2 ↗
  • Load per rack: 10 kg p.2 ↗
  • Wall clearance back: 100 mm p.2 ↗

Type Code Decoder

Position Token Meaning
1-6 p.1 ↗ CBS260 Model series CB-S, size 260
suffix p.1 ↗ UL UL variant with 100...120 V power supply and NEMA 5-20 plug (absent = standard 200...230 V, CEE 7/7)
suffix p.1 ↗ 230V Standard mains voltage variant 200...230 V 1~ 50/60 Hz, unit fuse 10,0 A
suffix p.1 ↗ 120V UL mains voltage variant 100...120 V 1~ 50/60 Hz, unit fuse 16,0 A

Ordering Data

Ordering Code Part No. Variant / Code Aplica a Note
CBS260-230V p.1 ↗ 9040-0223 200...230 V 1~ 50/60 Hz, 10,0 A, CEE 7/7 plug, Standard Standard model version
CBS260UL-120V p.1 ↗ 9040-0224 100...120 V 1~ 50/60 Hz, 16,0 A, NEMA 5-20 plug, Standard UL model version