Induztr
Resumen técnico Visor de documentos Modelos relacionados

Data Sheet Model FD S 115

Technical Reference Summary

Covered Models & Document Scope

Cámara de secado y calefacción modelo FD-S 115 con convección forzada (Solid.Line). Dos variantes: FDS115-230V (Art.-No. 9090-0024, 230 V 1~ 50/60 Hz, enchufe CEE 7/7) y FDS115UL-120V (Art.-No. 9090-0025, 120 V 1~ 60 Hz, enchufe NEMA 5-15).

Technical Data

Parameter Value Condición
Temperature range +10 °C above ambient temperature to 250 °C p.2 ↗ FDS115-230V
Temperature uniformity at 150°C 2.0 ± K p.2 ↗ FDS115-230V
Temperature fluctuation at 150°C 0.4 ± K p.2 ↗ FDS115-230V
Heating up time to 150°C 20 min p.2 ↗ FDS115-230V
Recovery time after door was opened for 30 s at 150°C 8 min p.2 ↗ FDS115-230V
Air change rate at 100°C 2 x/h p.2 ↗ FDS115-230V
Rated voltage 230 V p.2 ↗ FDS115-230V
Power frequency 50/60 Hz p.2 ↗ FDS115-230V
Nominal power 1,3 kW p.2 ↗ FDS115-230V
Unit fuse 6,3 A p.2 ↗ FDS115-230V
Phase (nominal voltage) 1~ p.2 ↗ FDS115-230V
Interior volume 106 L p.2 ↗ FDS115-230V
Net weight of the unit (empty) 48 kg p.2 ↗ FDS115-230V
Load per rack 20 kg p.2 ↗ FDS115-230V
Permitted load 75 kg p.2 ↗ FDS115-230V
Wall clearance back 160 mm p.2 ↗ FDS115-230V
Wall clearance sidewise 100 mm p.2 ↗ FDS115-230V
Width net 764 mm p.2 ↗ FDS115-230V
Height net 735 mm p.2 ↗ FDS115-230V
Depth net 616 mm p.2 ↗ FDS115-230V
Interior width 550 mm p.3 ↗ FDS115-230V
Interior height 550 mm p.3 ↗ FDS115-230V
Interior depth 350 mm p.3 ↗ FDS115-230V
Sound-pressure level 48 dB(A) p.3 ↗ FDS115-230V
Energy consumption at 150°C 340 Wh/h p.3 ↗ FDS115-230V

Operating Limits

  • Temperature range: +10 °C above ambient temperature to 250 °C p.2 ↗
  • Wall clearance back: 160 mm p.2 ↗
  • Wall clearance sidewise: 100 mm p.2 ↗
  • Permitted load: 75 kg p.2 ↗
  • Load per rack: 20 kg p.2 ↗

Type Code Decoder

Position Token Meaning
1-6 p.1 ↗ FDS115 Model series FD-S, size 115
7-9 p.1 ↗ UL- UL variant (120 V, NEMA 5-15 plug); absent for standard 230 V variant
10-13 p.1 ↗ 230V / 120V Rated voltage variant (230 V 50/60 Hz or 120 V 60 Hz)

Ordering Data

Ordering Code Part No. Variant / Code Aplica a Note
FDS115-230V p.1 ↗ 9090-0024 230 V 1~ 50/60 Hz, CEE 7/7 plug, Standard
FDS115UL-120V p.1 ↗ 9090-0025 120 V 1~ 60 Hz, NEMA 5-15 plug, Standard