Induztr
Resumen técnico Visor de documentos Modelos relacionados

Data Sheet Model MK 400

Technical Reference Summary

Covered Models & Document Scope

Cámara climática dinámica BINDER MK 400 para cambios rápidos de temperatura, modelo 400 (417 L). Dos variantes: MK400-400V (estándar, 400 V 3~/N/PE 50 Hz, Art.-No. 9020-0406) y MK400-480V-C (con convertidor de tensión y frecuencia, 480 V 3~ 60 Hz, Art.-No. 9020-0445).

Technical Data

Parameter Value Condición
Temperature range -40...180 °C p.2 ↗ MK400-400V and MK400-480V-C
Temperature uniformity dependent on set 0.1...1.2 ± K p.2 ↗ MK400-400V and MK400-480V-C
Temperature fluctuation dependent on set 0.1...0.5 ± K p.2 ↗ MK400-400V and MK400-480V-C
Cooling down time from 180°C to -40°C 115 min p.2 ↗ MK400-400V and MK400-480V-C
Rated voltage 400 V p.2 ↗ MK400-400V, 50 Hz
Rated voltage 480 V p.2 ↗ MK400-480V-C, 60 Hz
Power frequency 50 Hz p.2 ↗ MK400-400V
Power frequency 60 Hz p.2 ↗ MK400-480V-C
Nominal power 7 kW p.2 ↗ MK400-400V and MK400-480V-C
Unit fuse 16,0 A p.2 ↗ MK400-400V and MK400-480V-C
Phase (nominal voltage) 3~/N/PE p.2 ↗ MK400-400V
Phase (nominal voltage) 3~ p.2 ↗ MK400-480V-C
Interior volume 417 L p.2 ↗ MK400-400V and MK400-480V-C
Net weight of the unit (empty) 413 kg p.2 ↗ MK400-400V and MK400-480V-C
Load per rack 30 kg p.2 ↗ MK400-400V and MK400-480V-C
Permitted load 150 kg p.2 ↗ MK400-400V and MK400-480V-C
Wall clearance back 500 mm p.2 ↗ MK400-400V and MK400-480V-C
Wall clearance sidewise 300 mm p.2 ↗ MK400-400V and MK400-480V-C
Width net 1,115 mm p.2 ↗ MK400-400V and MK400-480V-C
Height net 1,710 mm p.2 ↗ MK400-400V and MK400-480V-C
Depth net 1,400 mm p.2 ↗ MK400-400V and MK400-480V-C
Interior width 735 mm p.3 ↗ MK400-400V and MK400-480V-C
Interior height 700 mm p.3 ↗ MK400-400V and MK400-480V-C
Interior depth 810 mm p.3 ↗ MK400-400V and MK400-480V-C
Sound-pressure level 65 dB(A) p.3 ↗ MK400-400V and MK400-480V-C

Operating Limits

  • Temperature range: -40...180 °C p.2 ↗
  • Wall clearance back: 500 mm p.2 ↗
  • Wall clearance sidewise: 300 mm p.2 ↗
  • Permitted load: 150 kg p.2 ↗
  • Load per rack: 30 kg p.2 ↗

Type Code Decoder

Position Token Meaning
1-5 p.1 ↗ MK400 Model series and size (MK dynamic climate chamber, size 400)
suffix p.1 ↗ -400V Standard variant, 400 V 3~/N/PE 50 Hz
suffix p.1 ↗ -480V-C Variant with voltage and frequency converter, 480 V 3~ 60 Hz

Ordering Data

Ordering Code Part No. Variant / Code Aplica a Note
MK400-400V p.1 ↗ 9020-0406 400 V 3~/N/PE 50 Hz, standard CEE 16 / 6 H plug 5-pin
MK400-480V-C p.1 ↗ 9020-0445 480 V 3~ 60 Hz, with voltage and frequency converter CEE 16 / 6 H plug 5-pin

Modelos relacionados