Induztr
Resumen técnico Visor de documentos Modelos relacionados

Data Sheet Model MKF 115

Technical Reference Summary

Covered Models & Document Scope

Cámara climática dinámica BINDER MKF 115 para cambios rápidos de temperatura con control de humedad. Dos variantes: MKF115-400V (estándar, 400 V 3~/N/PE 50 Hz, Art.-No. 9020-0379) y MKF115-480V-C (con convertidor de tensión y frecuencia, 480 V 3~ 60 Hz, Art.-No. 9020-0357).

Technical Data

Parameter Value Condición
Temperature range -40...180 °C p.3 ↗ MKF115-400V, 50 Hz
Temperature range -40...180 °C p.3 ↗ MKF115-480V-C, 60 Hz
Temperature range with humidity 10...95 °C p.3 ↗ MKF115-400V, 50 Hz
Humidity range 10...98 % RH p.3 ↗ MKF115-400V, 50 Hz
Temperature uniformity dependent on set 0.1...1.3 ± K p.3 ↗ MKF115-400V, 50 Hz
Temperature fluctuation dependent on set 0.1...0.6 ± K p.3 ↗ MKF115-400V, 50 Hz
Cooling down time from 180°C to -40°C 100 min p.3 ↗ MKF115-400V, 50 Hz
Max. heat compensation at 25°C and 90% RH 400 W p.3 ↗ MKF115-400V, 50 Hz
Rated voltage 400 V p.3 ↗ MKF115-400V, 50 Hz
Rated voltage 480 V p.3 ↗ MKF115-480V-C, 60 Hz
Power frequency 50 Hz p.3 ↗ MKF115-400V
Power frequency 60 Hz p.3 ↗ MKF115-480V-C
Nominal power 4,8 kW p.3 ↗ MKF115-400V, 50 Hz
Unit fuse 16,0 A p.3 ↗ MKF115-400V, 50 Hz
Phase (nominal voltage) 3~/N/PE p.3 ↗ MKF115-400V, 50 Hz
Phase (nominal voltage) 3~ p.3 ↗ MKF115-480V-C, 60 Hz
Interior volume 115 L p.3 ↗ MKF115-400V, 50 Hz
Net weight of the unit (empty) 280 kg p.3 ↗ MKF115-400V, 50 Hz
Load per rack 30 kg p.3 ↗ MKF115-400V, 50 Hz
Permitted load 60 kg p.3 ↗ MKF115-400V, 50 Hz
Sound-pressure level 62 dB(A) p.4 ↗ MKF115-400V, 50 Hz
Sound-pressure level 67 dB(A) p.4 ↗ MKF115-480V-C, 60 Hz
Energy consumption at 25°C and 60% RH 1,000 Wh/h p.4 ↗ MKF115-400V, 50 Hz
Width net 980 mm p.4 ↗ MKF115-400V, 50 Hz
Height net 1,725 mm p.4 ↗ MKF115-400V, 50 Hz

Operating Limits

  • Temperature range: -40...180 °C p.3 ↗
  • Temperature range with humidity: 10...95 °C p.3 ↗
  • Humidity range: 10...98 % RH p.3 ↗
  • Max. heat compensation at 25°C and 90% RH: 400 W p.3 ↗
  • Permitted load: 60 kg p.3 ↗
  • Load per rack: 30 kg p.3 ↗

Type Code Decoder

Position Token Meaning
1-3 p.1 ↗ MKF Series: dynamic climate chamber for rapid temperature changes with humidity control
4-6 p.1 ↗ 115 Model size / interior volume 115 L
suffix p.2 ↗ 400V Standard mains variant: 400 V 3~/N/PE 50 Hz
suffix p.2 ↗ 480V-C Option model with voltage and frequency converter: 480 V 3~ 60 Hz

Ordering Data

Ordering Code Part No. Variant / Code Aplica a Note
MKF115-400V p.2 ↗ 9020-0379 Standard, 400 V 3~/N/PE 50 Hz Standard model version
MKF115-480V-C p.2 ↗ 9020-0357 with voltage and frequency converter, 480 V 3~ 60 Hz Option model version