Induztr
Resumen técnico Visor de documentos Modelos relacionados

Data Sheet Model WIC 2

Technical Reference Summary

Covered Models & Document Scope

Modelo WIC 2 cámara walk-in para ensayos de estabilidad según ICH (Q1A); versión estándar WIC2-230V, Art.-No. 9800-0012, 230 V 1~ 50 Hz

Technical Data

Parameter Value Condición
Interior volume 18,523 L p.1 ↗ WIC 2
Rated voltage 230 V p.2 ↗ WIC 2, 50 Hz
Power frequency 50 Hz p.2 ↗ WIC 2
Phase (nominal voltage) 1~ p.2 ↗ WIC 2
Nominal power 2,7 kW p.2 ↗ WIC 2
Unit fuse 16,0 A p.2 ↗ WIC 2
Temperature range with humidity 5...50 °C p.2 ↗ WIC 2
Humidity range 20...90 % RH p.2 ↗ WIC 2
Temperature uniformity at 25°C and 60% RH 1.5 ± K p.2 ↗ WIC 2
Temperature uniformity at 40°C and 75% RH 1.5 ± K p.2 ↗ WIC 2
Temperature fluctuation at 25°C and 60% RH 0.5 ± K p.2 ↗ WIC 2
Temperature fluctuation at 40°C and 75% RH 0.5 ± K p.2 ↗ WIC 2
Humidity fluctuation at 25°C and 60% RH 2.5 ± % RH p.2 ↗ WIC 2
Humidity fluctuation at 40°C and 75% RH 2.5 ± % RH p.2 ↗ WIC 2
Net weight of the unit (empty) 98 kg p.2 ↗ WIC 2
Temperature accuracy ± 1.5 °C p.1 ↗ WIC 2
Relative humidity accuracy ± 2.5 % p.1 ↗ WIC 2
U value of room panels 0,21 W/m2K p.1 ↗ WIC 2
Surface load 50.000 N/m2 p.1 ↗ WIC 2
Refrigerant GWP 547 p.1 ↗ WIC 2

Operating Limits

  • Temperature range with humidity: 5...50 °C p.2 ↗
  • Humidity range: 20...90 % RH p.2 ↗
  • Unit fuse: 16,0 A p.2 ↗
  • Surface load: 50.000 N/m2 p.1 ↗

Type Code Decoder

Position Token Meaning
1-3 p.1 ↗ WIC Product family: walk-in chamber
5-8 p.1 ↗ 230V Mains voltage variant: 230 V, 1~, 50 Hz

Ordering Data

Ordering Code Part No. Variant / Code Aplica a Note
WIC2-230V p.1 ↗ 9800-0012 Standard, 230 V 1~ 50 Hz Standard model version

Modelos relacionados