Engineering Facts

14 facts · every claim linked to a page
Performance

The preparative system delivers 600 bar at 50 mL/min and 420 bar at 200 mL/min, so column pressure rating and particle size must be matched to the achievable flow/pressure envelope.

Determines which preparative columns and flow rates can be used without exceeding system pressure limits.

Source
“Power range of 600 bar at 50 mL/min and 420 bar at 200 mL/min for more resolution”
Selection

Flow rate up to 200 mL/min supports 50 mm id columns, so column diameter selection is directly tied to the pump head installed.

Guides matching pump head and column inner diameter during system configuration.

Source
“Flow rates up to 200 mL/min support 50 mm id columns, reducing number of required injections for more purified compound per unit time”
Selection

The 1260 preparative system reaches 50 mL/min and is limited to columns up to 30 mm id, capping the achievable scale.

Defines the maximum column size and throughput when choosing the 1260 platform.

Source
“Flow rates up to 50 mL/min for compatibility with columns up to 30 mm inner diameter”
Installation

Pump heads are exchangeable to extend the flow range, so the flow envelope can be changed later without replacing the whole system.

Affects initial configuration and planned upgrade path for higher flow rates.

Source
“Easy-to-change pump heads allow you to expand the flow range up to 200 mL/min”
Operation

Delay volumes are determined automatically by delay-sensor technology, so no extra fraction volume needs to be collected to compensate for delay.

Removes manual delay calibration and avoids unnecessary extra fraction collection.

Source
“Agilent delay-sensor technology determines delay volumes automatically and ensures that fractions are collected just-in-time—without the need to collect extra volumes.”
Performance

Low delay volume in the fraction collector reduces peak dispersion and carryover, which directly protects purity and recovery of collected fractions.

Justifies selecting low-delay fraction collectors and short delay coils for sharp separations.

Source
“Agilent fraction collectors are designed for low delay volumes to avoid peak dispersion and carryover between fractions, assuring high purity and recovery for your fractions.”
Operation

A recovery collector routes sample to safe locations instead of waste, protecting valuable material and improving throughput.

Important for high-value compounds where loss to waste is unacceptable.

Source
“Recovery collectors ensure that your valuable samples are not diverted to waste—everything goes to safe locations and can be recovered easily.”
Selection

The Open-Bed Sampler/Collector handles up to 864 samples or 432 fractions, and total fraction capacity can scale to 1296.

Sets the automation and throughput ceiling when sizing the sampler/collector.

Source
“Capacity to inject up to 864 samples or collect up to 432 fractions, or any combination, for full flexibility”
Selection

The 1260 valve-based fraction collector has one inlet, 11 collection ports and one waste port, limiting fraction routing per module.

Determines how many fractions can be routed per module and whether multiple modules are needed.

Source
“This module comprises one inlet port, 11 collection ports, and one waste port, facilitating easy compound isolation.”
Selection

The 1260 platform can collect up to 216 fractions per module and expand to 648 fractions within the same footprint.

Affects tube size selection and bench-space planning for fraction capacity.

Source
“Four outside diameters of fraction tubes enable collection of up to 216 fractions per module”
Installation

The preparative column compartment houses analytical and preparative columns, valves, delay coils and the MS Flow Modulator in one module.

Simplifies plumbing layout and bench-space planning for multi-column setups.

Source
“The 1290 Infinity II Preparative Column Compartment offers high flexibility for mounting and organization of your analytical and preparative LC column, valves, delay coils, and MS Flow Modulator—all within a single module.”
Performance

The MS Flow Modulator provides a controlled, reproducible split ratio for mass-selective fraction collection.

Needed for reliable mass-triggered fraction collection without variable split ratios.

Source
“the MS Flow Modulator delivers a controlled and reproducible split ratio”
Selection

Load & Lock columns use proprietary distribution plates giving 20% higher sample loading while minimizing backpressure and peak broadening.

Influences column choice when maximizing load per run is a priority.

Source
“Proprietary fluid and sample distribution plates enable 20% higher sample loading while minimizing backpressure and peak broadening.”
Installation

Load & Lock columns come standard with a stainless steel water jacket for temperature control.

Requires planning for water/coolant supply at the column installation site.

Source
“All our Load & Lock Columns are equipped as standard with a stainless steel water jacket for temperature control.”