Induztr
Resumen técnico Visor de documentos Modelos relacionados

Centrifugation Brochure Centrifuge 5804 5804 R 5810 5810 R Reliable Pe

Technical Reference Summary

Covered Models & Document Scope

Centrífuga Eppendorf 5804, 5804 R, 5810 y 5810 R (familia 58xx), incluidas las rotores de cesta oscilante, ángulo fijo, placas y microtubos compatibles y adaptadores.

Technical Data

Parameter Value Condición
Max. RCF with swing-bucket rotor S-4-72 3,234 × g p.4 ↗ 5804/5804 R, S-4-72, 4,200 rpm
Max. speed with swing-bucket rotor S-4-72 4,200 rpm p.4 ↗ 5804/5804 R, S-4-72
Max. RCF with plate rotor A-2-DWP 2,250 × g p.4 ↗ 5804/5804 R, A-2-DWP, 3,750 rpm
Max. speed with plate rotor A-2-DWP 3,750 rpm p.4 ↗ 5804/5804 R, A-2-DWP
Max. RCF with fixed-angle rotor FA-45-6-30 16,639 × g p.4 ↗ 5804/5804 R, FA-45-6-30, 11,000 rpm
Max. speed with fixed-angle rotor FA-45-6-30 11,000 rpm p.4 ↗ 5804/5804 R, FA-45-6-30
Max. RCF with fixed-angle rotor FA-45-30-11 20,800 × g p.4 ↗ 5804/5804 R, FA-45-30-11, 14,000 rpm
Max. speed with fixed-angle rotor FA-45-30-11 14,000 rpm p.4 ↗ 5804/5804 R, FA-45-30-11
Max. RCF with swing-bucket rotor S-4-104 3,214 × g p.5 ↗ 5810/5810 R, S-4-104, 3,900 rpm
Max. speed with swing-bucket rotor S-4-104 3,900 rpm p.5 ↗ 5810/5810 R, S-4-104
Max. RCF with fixed-angle rotor FA-45-6-30 20,130 × g p.5 ↗ 5810/5810 R, FA-45-6-30, 12,000 rpm
Max. speed with fixed-angle rotor FA-45-6-30 12,000 rpm p.5 ↗ 5810/5810 R, FA-45-6-30
Max. RCF with swing-bucket rotor S-4-104 3,214 × g p.10 ↗ S-4-104, 3,900 rpm
Max. RCF with swing-bucket rotor A-4-81 3,220 × g p.10 ↗ A-4-81, 4,000 rpm
Max. RCF with swing-bucket rotor A-4-62 3,220 × g p.10 ↗ A-4-62, 4,000 rpm
Max. RCF with swing-bucket rotor S-4-72 3,234 × g p.10 ↗ S-4-72, 4,200 rpm
Max. RCF with swing-bucket rotor A-4-44 4,500 × g p.10 ↗ A-4-44, 5,000 rpm
Max. RCF with plate rotor A-2-DWP-AT 3,486 × g p.11 ↗ A-2-DWP-AT, 4,500 rpm
Max. RCF with plate rotor A-2-DWP 2,250 × g p.11 ↗ A-2-DWP, 3,700 rpm
Max. RCF with fixed-angle rotor F-34-6-38 18,500/15,500 × g p.11 ↗ F-34-6-38, 12,000/11,000 rpm
Max. RCF with fixed-angle rotor FA-45-6-30 20,130/16,639 × g p.11 ↗ FA-45-6-30, 12,100/11,000 rpm
Max. RCF with fixed-angle rotor FA-45-48-11 19,083 × g p.11 ↗ FA-45-48-11, 13,000 rpm
Volume range with swing-bucket rotor S-4-104 0.2–750 mL p.10 ↗ S-4-104
Volume range with swing-bucket rotor A-4-81 0.2–500 mL p.10 ↗ A-4-81
Volume range with swing-bucket rotor A-4-62 0.2–250 mL p.10 ↗ A-4-62

Operating Limits

  • Max. loading height for plate rotor A-2-DWP: 89 mm p.4 ↗
  • Max. tube volume for swing-bucket rotor S-4-104: 750 mL p.10 ↗
  • Max. tube volume for swing-bucket rotor A-4-81: 500 mL p.10 ↗
  • Max. tube volume for swing-bucket rotor A-4-62: 250 mL p.10 ↗
  • Max. tube volume for swing-bucket rotor S-4-72: 250 mL p.10 ↗
  • Max. tube volume for swing-bucket rotor A-4-44: 100 mL p.10 ↗

Ordering Data

Ordering Code Part No. Variant / Code Aplica a Note
Adapter set of 2, 1.5/2.0 mL, Ø 11 mm p.12 ↗ 5825 740 009 Adapter, max RCF 3,197 × g
Adapter set of 2, 12 × 75 mm, Ø 12 mm p.12 ↗ 5825 747 003 Adapter, max RCF 3,112 × g
Adapter set of 2, 2.6–8 mL, Ø 13 mm p.12 ↗ 5825 738 004 Adapter, max RCF 3,044 × g
Adapter set of 2, 5.5–12 mL, Ø 16 mm p.12 ↗ 5825 736 001 Adapter, max RCF 3,061 × g
Adapter set of 2, 9 mL, Ø 17.5 mm p.12 ↗ 5825 743 008 Adapter, max RCF 3,044 × g
Adapter set of 2, 14 mL, Ø 17.5 mm p.12 ↗ 5825 748 000 Adapter, max RCF 3,146 × g
Adapter set of 2, 1.5/2.0 mL, Ø 11 mm p.18 ↗ 5804 794 001 Adapter, max RCF 3,136 × g
Adapter set of 2, 2.6–8 mL, Ø 13 mm p.18 ↗ 5804 789 008 Adapter, max RCF 3,136 × g
Adapter set of 2, 5.5–12 mL, Ø 16 mm p.18 ↗ 5804 791 002 Adapter, max RCF 3,096 × g
Adapter set of 2, 9 mL, Ø 17.5 mm p.18 ↗ 5804 792 009 Adapter, max RCF 3,116 × g