Selection
The operating temperature range depends on the cooling method chosen, so the required range must be matched to water cooling or a refrigerator.
Determines whether the base unit alone, water cooling, or an active refrigerator is needed for the process temperature.
Source
“Operating temperature range 28...300 °C with water cooling 20...300 °C with refrigerator -20...300 °C”
Installation
The electrical supply must be 200-240 V single or two-phase at 50/60 Hz and the branch circuit must be rated for the 16 A maximum current.
Defines the site power configuration and breaker/cabling capacity required before installation.
Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 16 A”
Safety
External branch circuit protection according to UL 489 is mandatory for this equipment.
The installer must provide a listed UL 489 protective device; omitting it is a compliance and safety defect.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage and frequency tolerances are each ±10% but must not deviate in opposite directions simultaneously.
Prevents selecting a supply point where combined voltage and frequency excursions exceed the allowed envelope.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation
Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at order time.
The plug type must be confirmed on the order, otherwise the unit may not connect to the site receptacle.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation
The ambient temperature at the installation site must stay between 5 °C and 40 °C.
Room conditions outside this window can affect performance and must be checked during site planning.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation
The housing is IP20, so the unit must be installed where it is protected from water and dust ingress.
Restricts placement to dry indoor locations and rules out washdown or wet environments.
Source
“Degree of Protection IP20”
Selection
The pump connection is M16x1 male, which dictates the mating hose or adapter to be ordered.
Ensures the correct external circuit connections and hoses are specified with the unit.
Source
“Pump connection M16x1 male”
Performance
The pressure pump delivers up to 25 l/min at 0.7 bar and the suction pump up to 18.5 l/min at 0.4 bar.
Used to verify that the external loop flow resistance and required flow rate are within pump capability.
Source
“max. delivery 25 l/min Delivery pressure (head) 0,7 bar Suction pump yes max. delivery (suction) 18,5 l/min Delivery suction pressure (head) 0,4 bar”
Selection
The maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s.
Limits the choice of heat transfer fluid, especially at low temperatures where viscosity rises.
Source
“max. permissible kin. viscosity 50 mm²/s”
Installation
The immersion depth is 175 mm and the telescopic arm extends up to 884 mm, which sets the bath mounting geometry.
Confirms the unit fits the bath depth and that the telescopic bridge can span the bath opening.
Source
“Immersion depth 175 mm”
Cumplimiento
The unit is classified Class A, Group 1 for electromagnetic disturbance to EN55011.
Class A equipment is intended for non-domestic environments and may require EMC measures in residential areas.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.
The overtemperature cut-out must be set for the application to protect the bath and medium from overheating.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Maintenance
The 4-year warranty requires registration.
Registration must be completed to preserve warranty coverage on the equipment.
Source
“4-year warranty - registration required.”