CC-508w Circulating Bath Datasheet

Contenido

  1. Overview p. 1
    Introduces the CC-508w circulating bath model
  2. Spec pp. 1-2
    Performance and technical data according to DIN 12876
  3. p. 2 table, 3 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
max. ambient temperature40 °C
min. ambient temperature5 °C
from Serial-No.:1.0/26

Technical data according to DIN 12876

ParameterValue
Operating temperature range-55...200 °C
Temperature stability at -10°C0,02 K
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power2,45 kW
Cooling power at 100°C1,5 kW
Cooling power at 20°C1,5 kW
Cooling power at 0°C1,5 kW
Cooling power at -20°C0,8 kW
Cooling power at -40°C0,25 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-1270 (A3, H220)
Global Warming Potential (GWP)0
UN-numberUN 3358
Pressure pumpyes
max. delivery25 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)18,5 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
max. permissible kin. viscosity50 mm²/s
Bath volume5 l
min. filling capacity4 l
Width bath opening WxD / bath depth120 x 110 / 160 mm
Height of bath opening562 mm
Overall dimensions WxDxH **410x480x764 mm
sound pressure level +/- 4 dB(A)57 dB(A)
Power supply requirement208-230V 1~/2~ 60Hz
Degree of ProtectionIP20

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit is water-cooled and requires cooling water intake between 5°C and 20°C, so the site must provide a suitable cooling water supply within this range.

Cooling water availability and temperature determine whether the unit can be installed at the site and how much cooling power will be available.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

If cooling water temperature falls below the condensation point, the cooling water pipes inside the unit must be insulated to prevent condensation.

Condensation on internal cooling water pipes can cause water damage and electrical risk, so insulation must be planned during installation.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase, so the cooling water circuit must be sized with margin.

The published cooling capacities are only valid under stated conditions; higher cooling water temperature reduces available cooling and may require more flow.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Incompatible water chemistry can corrode or damage these materials, so water quality must be checked before connection.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.

Supply quality must be verified against these combined tolerances to avoid invalid operating conditions and equipment malfunction.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified when ordering.

The correct plug type must be selected at order time to match the site's electrical outlet.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation

Three-phase devices below 63 A are supplied with cable but without plug, so a suitable plug must be provided on site.

Installation planning must include procurement and wiring of a plug for three-phase units under 63 A.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Cumplimiento

The equipment is classified as EN55011 Class A Group 1 for electromagnetic disturbance, which has implications for the intended environment.

Class A equipment may not be suitable for residential environments and may require additional EMC measures depending on the installation site.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety

The refrigerant is R-1270 with safety group A3 and GHS H220, meaning it is a flammable refrigerant requiring appropriate safety measures.

Flammable refrigerant classification affects site safety assessment, ventilation requirements and handling procedures.

Source
“R-1270 (A3, H220)”
Selection

The unit has a Global Warming Potential of 0, which is relevant for environmental and regulatory selection criteria.

Low-GWP refrigerant supports compliance with environmental regulations and corporate sustainability requirements.

Source
“Global Warming Potential (GWP) 0”
Selection

The pump connection is M16x1 male, so hoses and fittings must match this thread.

Matching the external circuit connections to M16x1 male avoids adapters and leakage risks.

Source
“Pump connection M16x1 male”
Operation

The maximum permissible kinematic viscosity of the pumped medium is 50 mm²/s, which limits the fluids that can be circulated.

Selecting a medium with higher viscosity than allowed can overload the pump and reduce circulation performance.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The unit has degree of protection IP20, so it must be installed in a dry environment protected from water ingress.

IP20 does not protect against water, so the installation location must avoid splashes and moisture.

Source
“Degree of Protection IP20”
Operation

The permissible ambient temperature range is 5°C to 40°C, which constrains the room conditions where the unit can operate.

Operating outside the ambient temperature range can affect performance, reliability and safety.

Source
“max. ambient temperature 40 °C”