CC-820w Circulating Bath Datasheet

Contenido

  1. Overview p. 1
    Introduces the CC-820w circulating bath and its main features
  2. Spec p. 2
    Lists technical data according to DIN 12876 standard
  3. p. 2 table, 22 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Consumption at water 15°C, flow 20°C78 l/h
Consumption at water 15°C, flow 0°C60 l/h
Consumption at water 15°C, flow -20°C60 l/h
Consumption at water 15°C, flow -40°C66 l/h
Consumption at water 15°C, flow -80°C30 l/h
min. cooling water differential pressure3 bar
max. cooling water pressure6 bar
Bath volume17 l
min. filling capacity15 l
Bath capacity with displacement rack10 l
Width bath opening WxD / bath depth270 x 150 / 200 mm
Height of bath opening903 mm
Overall dimensions WxDxH **539x629x1102 mm
Net weight150 kg
Power supply requirement (3 phase)400V 3~N 50Hz
max. current (3 Phase)12 A
Fuse (3 phase)3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
max. ambient temperature40 °C
min. ambient temperature5 °C
from Serial-No.:S544140

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-80...100 °C
Temperature stability at -10°C0,02 K
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power3 kW
Cooling power at 100°C1,2 kW
Cooling power at 20°C1,2 kW
Cooling power at 0°C1,2 kW
Cooling power at -20°C1,1 kW
Cooling power at -40°C0,9 kW
Cooling power at -60°C0,6 kW
Cooling power at -80°C0,14 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity0,8 kg
CO2 equivalent1,7 t
Refrigerant 2nd stage (ASHRAE, GHS)R-23 (A1, H280)
Global Warming Potential (GWP)14800
Refrigerant quantity 2nd stage0,57 kg
CO2 equivalent 2nd stage8,4 t
Pressure pumpyes
max. delivery25 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)18,5 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male

Order-No.: CC-820w; 2025.0002.01

Engineering Facts

13 facts · every claim linked to a page
Installation

The unit requires a three-phase 400 V supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply and protection device needed before installation.

Source
“Power supply requirement (3 phase) 400V 3~N 50Hz”
Installation

The cooling water circuit must be supplied at 3–6 bar differential pressure, so the building water system must be checked for adequate pressure.

Insufficient cooling water pressure will prevent the water-cooled refrigeration machine from operating correctly.

Source
“min. cooling water differential pressure 3 bar”
Installation

Cooling water intake must be between 5 °C and 20 °C, and pipes must be insulated if water can drop below the condensation point.

Defines the permissible cooling water supply conditions and required pipe insulation at the installation site.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops significantly at low temperatures, from 1.2 kW at 20 °C to only 0.14 kW at -80 °C, which must be considered for load sizing.

The available cooling capacity at the target low temperature determines whether the unit can handle the process heat load.

Source
“at -80°C 0,14 kW”
Installation

The unit is water-cooled and uses R-452A and R-23 refrigerants with a combined CO2 equivalent of about 10.1 t, which affects refrigerant reporting obligations.

Refrigerant charge and GWP determine leak-check and reporting requirements under F-gas regulations.

Source
“CO2 equivalent 1,7 t”
Operation

The pressure pump delivers up to 25 l/min at 0.7 bar and the suction pump up to 18.5 l/min, so external circuit pressure drop must be checked.

The pump head limits the permissible pressure drop in the external cooling circuit.

Source
“max. delivery pressure 0,7 bar”
Operation

The maximum permissible kinematic viscosity of the circulating fluid is 50 mm²/s, which constrains the choice of heat transfer medium at low temperatures.

Selecting a fluid with too high viscosity at operating temperature will overload the pump.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The bath requires a minimum filling capacity of 15 l within a 17 l bath volume, so the fluid charge must be planned accordingly.

Determines the minimum heat transfer fluid quantity needed for safe operation.

Source
“min. filling capacity 15 l”
Installation

The unit weighs 150 kg and measures 539x629x1102 mm, so floor loading and access clearance must be verified.

Needed for transport planning, doorway clearance and floor load assessment.

Source
“Overall dimensions WxDxH ** 539x629x1102 mm”
Operation

The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be installed in a wet or hot environment.

Defines the permissible ambient conditions and protection level at the installation location.

Source
“max. ambient temperature 40 °C”
Cumplimiento

Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.

The site power quality must be checked against these combined tolerances to avoid damage.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Three-phase devices below 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.

Affects the electrical installation scope and materials needed at commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL, relevant for flammable or sensitive media.

Confirms the safety concept required when selecting the unit for a given application.

Source
“With adjustable overtemperature protection according to DIN 12876.”