CC-K20 Refrigerated/Heating Circulator Datasheet

Contenido

  1. Overview p. 1
    Product introduction and key features of the CC-K20 circulator
  2. Spec pp. 1-2
    Technical data, performance figures and operating limits per DIN 12876
  3. p. 2 table, 11 rows
  4. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
sound pressure level +/- 4 dB(A)50 dB(A)
Power supply requirement220-240V 1~/2~ 50/60Hz
max. current refrigerated bath2 A
max. current immersion thermostat10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S395643

1.0/20

Technical data according to DIN 12876

ParameterValue
Operating temperature range-30...200 °C
Temperature stability at 70°C0,02 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at 20°C0,4 kW
Cooling power at 0°C0,35 kW
Cooling power at -10°C0,27 kW
Cooling power at -20°C0,16 kW
Cooling power at -30°C0,065 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,072 kg
CO2 equivalent0,0 t
Pressure / Suction pump max. delivery27 l/min
Pressure / Suction pump max. delivery pressure0,7 bar
Pressure / Suction pump max. delivery (suction)22 l/min
Pressure / Suction pump max. delivery pressure (suction)0,4 bar
Pump connenction (optional)M16x1 male
Bath volume17 l
Width bath opening WxD290x329 mm
Bath depth150 mm
Height of bath opening450 mm
Overall dimensions WxDxH **350x555x615 mm
Net weight41 kg

** см. сноску в оригинале (в тексте страницы пояснение к ** отсутствует)

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is classified as safety class III / FL per DIN 12876, which determines the permitted flammable-liquid use and protective measures.

Safety class governs whether and how flammable bath fluids may be used and what over-temperature protection is required.

Source
“Safety classification III / FL”
Selection

Cooling capacity drops from 0.4 kW at 20 °C to 0.065 kW at -30 °C, so the low-temperature working point must be checked against the actual heat load.

Sizing at the coldest set point prevents undersized cooling and long pull-down times.

Source
“at -30°C 0,065 kW”
Safety

The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which imposes ventilation and ignition-source restrictions on the installation site.

A3 flammability affects room ventilation, electrical zoning and site risk assessment.

Source
“R-290 (A3, H220)”
Installation

The unit is air-cooled, so sufficient clearance and cool intake air must be provided at the installation location.

Air-cooled condensers need unobstructed airflow; poor clearance reduces cooling capacity and can trip the compressor.

Source
“Refrigeration machine air-cooled, natural refrigerant”
Installation

The power supply must be 220-240 V single or two-phase 50/60 Hz with a fuse between 10 A and 16 A.

Correct supply and fuse rating are mandatory for safe connection and warranty compliance.

Source
“Power supply requirement 220-240V 1~/2~ 50/60Hz”
Installation

Voltage and frequency tolerances are ±5 % and ±2 % respectively, but the combination of -5 % voltage with +2 % frequency is not allowed.

Grid quality at the site must be verified; an out-of-tolerance combination can damage the compressor and electronics.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The enclosure is rated IP20, so the unit must be kept away from water spray and conductive dust.

IP20 limits where the bath can be placed and what cleaning methods are permitted.

Source
“Degree of Protection IP20”
Operation

Ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may fall as ambient temperature rises.

Room conditions directly affect achievable bath temperature and must be considered in the layout.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Operation

When a Huber pump adapter is used, polyglycol must not be used as the heat transfer fluid.

Using the wrong fluid with the pump adapter can damage the pump and seals.

Source
“Polyglycol is not permissible to be used as a heat transfer fluid”
Cumplimiento

The equipment is EMC class A, group 1 per EN55011, which is intended for industrial environments rather than residential use.

Class A equipment may cause interference in residential areas and affects where it may be installed.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

The plug type must be stated on the order to avoid a non-matching mains connection on site.

Source
“with power cable and country-specific plug (please specify when ordering)”
Selection

The pressure/suction pump delivers up to 27 l/min at 0.7 bar pressure and 22 l/min at 0.4 bar suction.

Pump flow and head determine whether an external circuit or reactor can be supplied adequately.

Source
“max. delivery 27 l/min”
Operation

The bath holds 17 l with a working opening of 290 x 329 mm and a depth of 150 mm, which sets the maximum sample size and fluid volume.

Bath volume and opening dimensions determine fluid fill quantity, thermal inertia and what can be immersed.

Source
“Bath volume 17 l”
Maintenance

A drain valve with cap is available as an accessory, which simplifies fluid changes and bath draining.

Planning the drain accessory at installation avoids awkward fluid handling later.

Source
“Drain valve with cap #6839”