CC-K6 Refrigerated/Heating Circulator Datasheet

Contenido

  1. Overview p. 1
    Product introduction and main features of CC-K6 circulator
  2. Spec pp. 1-2
    Technical data, performance and operating specifications per DIN 12876
  3. p. 2 table, 12 rows
  4. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
Net weight25 kg
sound pressure level +/- 4 dB(A)57 dB(A)
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current refrigerated bath1,35 A
max. current immersion thermostat10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S393652

Technical data according to DIN 12876

ParameterValue
Operating temperature range-25...200 °C
Temperature stability at 70°C0,02 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Cooling power at 20°C0,2 kW
Cooling power at 0°C0,15 kW
Cooling power at -10°C0,1 kW
Cooling power at -20°C0,05 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,047 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Pressure / Suction pump max. delivery27 l/min
max. delivery pressure0,7 bar
max. delivery (suction)22 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
Bath volume4,5 l
Width bath opening WxD140x120 mm
Bath depth150 mm
Height of bath opening376 mm
Overall dimensions WxDxH **210x400x546 mm

** см. сноску в оригинале (в тексте страницы пояснение к ** отсутствует)

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is classified as safety class III / FL per DIN 12876, which determines the permitted flammable-liquid use and required protection concept.

Safety class governs whether the bath may be used with flammable media and what protective measures are mandatory.

Source
“Safety classification III / FL”
Safety

The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, so the installation location must respect ignition-source and ventilation rules.

A3 flammability affects room classification, ventilation and prohibition of ignition sources near the unit.

Source
“R-290 (A3, H220)”
Installation

The refrigerating unit is air-cooled, so sufficient clearance and cool intake air must be provided at the installation site.

Air-cooled condensers need unobstructed airflow; poor clearance reduces cooling capacity and can trip the unit.

Source
“Refrigeration machine air-cooled, natural refrigerant”
Performance

Cooling capacity is only 0.05 kW at -20 °C, so the unit is unsuitable for sustained high heat loads at the low end of its range.

Knowing the low-temperature cooling capacity prevents oversizing expectations for exothermic reactions near -20 °C.

Source
“at -20°C 0,05 kW”
Installation

The power supply requirement is 208-240 V single or two-phase 50/60 Hz, so the site must provide the correct mains configuration.

Mismatched supply voltage or phase configuration prevents commissioning and may damage the unit.

Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz”
Installation

The circuit must be protected with a fuse between 10 A and 16 A, matching the maximum current draw of the immersion thermostat.

Correct upstream fusing is required for safe operation and to avoid nuisance trips or inadequate protection.

Source
“min. Fuse 10A max. Fuse 16A”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.

Sites with unstable mains must be checked against this combined tolerance rule before installation.

Source
“-5% voltage and +2% frequency -> not allowed!”
Operation

The permissible ambient temperature range is 5 °C to 40 °C, and cooling capacity may drop as ambient temperature rises.

Hot rooms degrade performance; the installation room must stay within the stated ambient limits.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Cumplimiento

The equipment is EMC class A, group 1 per EN55011, so it is intended for industrial environments and may cause interference in residential areas.

Class A equipment has different EMC obligations and may not be suitable for residential or light-industrial locations.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

The plug type must be stated on the order, otherwise the unit may arrive without a usable mains connection.

Source
“with power cable and country-specific plug (please specify when ordering)”
Safety

The degree of protection is IP20, so the unit must be kept away from water spray and conductive dust.

IP20 limits where the unit can be placed and rules out washdown or wet environments.

Source
“Degree of Protection IP20”
Maintenance

The refrigerant charge is only 0.047 kg with a GWP of 0.02, giving a CO2 equivalent of 0.0 t and minimal leak-reporting burden.

Small charge and low GWP simplify refrigerant record-keeping and leak-check obligations.

Source
“Refrigerant quantity 0,047 kg”
Operation

The pressure/suction pump delivers up to 27 l/min at 0.7 bar, which sets the external circuit pressure-drop limit.

External loop resistance must be checked against this pump curve to avoid insufficient circulation.

Source
“max. delivery 27 l/min”
Selection

The bath volume is 4.5 l with a 140x120 mm opening, which constrains the size of immersed vessels and samples.

Bath volume and opening size determine whether the intended sample load fits and how fast temperature changes occur.

Source
“Bath volume 4,5 l”