Contenido

  1. Overview p. 1
    Product introduction and general description of the CS 100 bath
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 1 table, 33 rows

CS 100 by Van der Heijden — Technical data according to DIN 12876

ParameterValue
Order-No.VDH1101104
Operating temperature range-20...15 °C
temperature set point / displaycolour LED Touchscreen
Internal temperature sensorPt100
Temperature stability at -10°C0,5 K
Interface digitalRS232
Safety classificationI / NFL
Cooling power at ambient temperature 20°C
at 15°C18 kW
at 10°C10 kW
at 0°C9,5 kW
at -10°C7,5 kW
at -20°C4,5 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Refrigerant quantity4,5 kg
CO2 equivalent6,29 t
Circulation pump:
max. delivery70 l/min
max. delivery pressure3,7 bar
Pump connectionG3/4 male
min. filling capacity160 l
max. filling capacity175 l
Overall dimensions WxDxH **800x850x1665 mm
Net weight265 kg
sound pressure level +/- 4 dB(A)70 dB(A)
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)12 A
Fuse (3 phase)3x16A
Degree of ProtectionIP52
min. ambient temperature5 °C
max. ambient temperature32 °C
from Serial-No.:1.0/24

** Overall dimensions WxDxH

Engineering Facts

13 facts · every claim linked to a page
Installation

The chiller requires a three-phase 400 V supply with a 3x16 A fuse, so the site must provide a suitable three-phase feed and protection.

Determines electrical supply and upstream protection design before installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so the installer must provide and fit the site plug.

Affects electrical installation scope and what must be ordered or prepared on site.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit must be operated within an ambient temperature window of 5 to 32 °C, so the plant room must be kept within this range.

Ambient conditions affect cooling performance and reliable operation.

Source
“min. ambient temperature 5 °C”
Performance

Cooling capacity is rated at 20 °C ambient and drops as ambient rises, so selection must account for actual room temperature.

Prevents undersizing when the chiller is installed in a warmer environment.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation

The chiller needs 160 to 175 litres of filling capacity, so the loop volume and expansion provisions must be planned accordingly.

Determines system volume, fluid quantity and expansion tank sizing.

Source
“min. filling capacity 160 l”
Operation

The recommended thermofluid is a 50:50 water-monoethylene glycol mixture, so the fluid must be selected to match this recommendation.

Fluid choice affects freeze protection, heat transfer and pump performance.

Source
“Recommended thermofluid: Water - Monoethylene Glycol 50:50”
Safety

The refrigerant is R-449A classified A1 with GHS H280, so refrigerant handling and ventilation precautions must follow this classification.

Needed for safety assessment, leak response and refrigerant documentation.

Source
“R-449A (A1, H280)”
Cumplimiento

The unit is EN55011 Class A Group 1, so it is intended for industrial environments and may require EMC measures in residential areas.

Affects EMC compliance and suitability of the installation location.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Voltage and frequency tolerances are ±5% and ±2% but not simultaneously in the unfavourable combination, so the supply quality must be checked.

Ensures the electrical supply stays within the allowed operating envelope.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The pump connection is G3/4 male, so mating hydraulic fittings must be G3/4 female.

Determines the connection hardware needed for the hydraulic circuit.

Source
“Pump connection G3/4 male”
Operation

Flow rate is adjusted via a manual bypass valve on the back of the chiller, so rear access must be kept clear.

Requires installation clearance at the rear for commissioning and adjustment.

Source
“The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.”
Installation

The unit produces about 70 dB(A) sound pressure, so acoustic treatment or separation from occupied areas may be needed.

Affects placement and noise control in the plant room.

Source
“sound pressure level +/- 4 dB(A) 70 dB(A)”
Cumplimiento

The equipment is compliant with US-SNAP for industrial process refrigeration, so its end use must match this application.

Confirms the permitted application for regulatory compliance.

Source
“The US-SNAP end-use for this equipment is the industrial process refrigeration.”