Grande Fleur

Contenido

  1. Overview p. 1
    Product introduction and main features of Grande Fleur
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 2 table, 16 rows
  4. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
max. delivery pressure0,9 bar
Pump connectionM24x1,5 male
max. permissible kin. viscosity50 mm²/s
min. filling capacity1,5 l
Filling capacity expansion tank3 l
Overall dimensions WxDxH **295x530x570 mm
Net weight53 kg
sound pressure level +/- 4 dB(A)61 dB(A)
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current11,5 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S393363

** Please respect space requirements. See operating conditions at www.huber-online.com

Grande Fleur Technical data according to DIN 12876

CharacteristicValue
Operating temperature range-40...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power at 240V2 kW
Heating power at 230V1,8 kW
Heating power at 220V1,6 kW
Heating power at 208V1,5 kW
Heating power convertible1,5 - 2 kW
Cooling power with Thermooil at 200°C0,6 kW
Cooling power with Thermooil at 150°C0,6 kW
Cooling power with Thermooil at 100°C0,6 kW
Cooling power with Thermooil at 50°C0,6 kW
Cooling power with Thermooil at 20°C0,6 kW
Cooling power with Ethanol at 0°C0,6 kW
Cooling power with Ethanol at -20°C0,35 kW
Cooling power with Ethanol at -30°C0,2 kW
Cooling power with Ethanol at -40°C0,04 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,099 kg
CO2-Äquvivalent0,0 t
UN-numberUN 3358
Circulation pump: max. delivery47 l/min

Engineering Facts

13 facts · every claim linked to a page
Selection

The unit is rated for an operating temperature range of -40 to 200 °C, so the application must stay within these limits.

Defines whether the circulator can cover the required process temperatures.

Source
“Operating temperature range -40...200 °C”
Performance

Cooling capacity drops sharply at low temperatures with ethanol, from 0.6 kW at 0 °C to only 0.04 kW at -40 °C.

Low-temperature heat loads may exceed available cooling capacity near the lower limit.

Source
“at -40°C 0,04 kW”
Safety

The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects installation and safety planning.

Flammable refrigerant classification requires appropriate site safety measures.

Source
“R-290 (A3, H220)”
Installation

The unit is air-cooled, so sufficient ventilation and respect for space requirements are needed at the installation site.

Air-cooled condensers need adequate airflow and clearances to maintain cooling performance.

Source
“Refrigeration machine air-cooled, natural refrigerant”
Installation

The pump connection is M24x1.5 male, so hoses or adapters must match this thread.

Prevents mismatch with existing pipework or flexible tempering hoses.

Source
“Pump connection M24x1,5 male”
Installation

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, limiting fluid selection at low temperatures.

Ensures the pump can circulate the fluid properly without excessive load.

Source
“max. permissible kin. viscosity 50 mm²/s”
Operation

The minimum filling capacity is 1.5 l and the expansion tank holds 3 l, which sets the required fluid charge.

Avoids underfilling or overflow during thermal expansion.

Source
“min. filling capacity 1,5 l”
Installation

The unit requires a 208-240 V single-phase or two-phase supply with a maximum current of 11.5 A and a 16 A fuse.

Determines electrical supply and protection sizing for the installation.

Source
“max. current 11,5 A”
Installation

The degree of protection is IP20, so the unit must be installed in a dry, protected environment.

IP20 is not water-protected and is unsuitable for wet or washdown areas.

Source
“Degree of Protection IP20”
Operation

The permissible ambient temperature range is 5 to 40 °C, and cooling capacity may drop if ambient temperature rises.

High ambient temperature reduces cooling performance and may limit operation.

Source
“max. ambient temperature 40 °C”
Cumplimiento

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Ensures the electrical supply stays within the allowed tolerance envelope.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified when ordering.

Prevents delivery of the wrong plug type for the installation country.

Source
“with power cable and country-specific plug (please specify when ordering)”
Cumplimiento

The equipment is classified as EN55011 Class A, Group 1 for electromagnetic disturbance.

Class A equipment may require EMC measures in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”