Product introduction and model identification for HB60
Specp. 2
Technical specifications according to DIN 12876
p. 1table, 24 rows
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
60...250 °C
Resolution of display
0,01 K
Temperature stability at 70°C
0,1 K
temperature set point / display
5,7" colour Touchscreen
Absolute accuracy
setup for calibration
Internal temperature sensor
Pt100
External sensor
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
II / FL
Heating power
6 kW
Circulation pump: max. delivery
90 l/min
max. delivery pressure
2,5 bar
Delivery at 0,5 bar
75 l/min
Pump connection
M30x1,5 male
min. filling capacity
3,5 l
Overall dimensions WxDxH
323x451x498 mm
Degree of Protection
IP20
Net weight
49 kg
Power supply (3 Phase)
400V 3~N 50Hz
max. current (3 Phase)
10,5 A
Fuse (3 phase)
3x16A
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S546850
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit is designed exclusively for external open operation and has no expansion tank, so the connected application must accommodate fluid filling and thermal volume changes.
Determines whether the unit can be used with a closed loop or requires an external open bath/application with volume compensation.
Source
“The machine is developed exclusively for external open operation, and does not possess an expansion tank.”
Installation
Three-phase devices below 63 A are supplied with cable but without a plug, so the installer must provide a suitable country-specific plug and connection.
Affects electrical installation planning and the bill of materials for commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The pump connection is M30x1.5 male, so hoses or piping must use a compatible union to DIN 3863.
Needed to select matching hoses/fittings and avoid leaks or adapters at installation.
Source
“Pump connection M30x1,5 male”
Installation
The pump connections use a 60° Y arrangement to DIN 3863 and pipework/flexible hoses must be fitted with a union to DIN 3863.
Ensures correct mechanical interface and sealing when connecting the circulation loop.
Source
“Pump connections: 60° "Y" arrangement to DIN 3863, pipework/flexible tempering hoses: With union to DIN 3863.”
Installation
The unit requires a 3-phase 400 V 3~N 50 Hz supply with a 3x16 A fuse, so the electrical supply must be sized accordingly.
Critical for electrical design, breaker selection and avoiding nuisance tripping.
Source
“Fuse (3 phase)
3x16A”
Installation
Voltage and frequency tolerances are ±10% only if they do not run in opposite directions, so supply quality must be checked before commissioning.
Prevents operation outside allowed supply conditions that could damage the unit or void performance.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Safety
The unit has adjustable overtemperature protection according to DIN 12876, which must be set appropriately for the application.
Important for safe operation and for setting a protective limit before running the process.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Safety
The unit is classified as Class A, Group 1 for electromagnetic disturbance to EN55011, which affects EMC planning in the installation environment.
Class A equipment may require additional EMC measures in residential or sensitive environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety
Leakage current from external temperature control must be considered, so protective earthing and RCD selection need attention.
Warns the installer that leakage current may affect protective device selection and earthing design.
Source
“Attention: leakage current externe Temperierung”
Operation
The minimum filling capacity is 3.5 l, so the external application and circuit must provide at least this fluid volume.
Prevents pump cavitation and ensures sufficient thermal buffer during operation.
Source
“min. filling capacity
3,5 l”
Operation
The ambient temperature must remain between 5 °C and 40 °C, so the installation location must be kept within this range.
Operating outside this range can reduce performance, trip protections or damage components.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Performance
The circulation pump delivers up to 90 l/min and 2.5 bar, with 75 l/min at 0.5 bar, so the external circuit pressure drop must be checked against this curve.
Determines whether the pump can overcome the pressure drop of the external application at the required flow.
Source
“max. delivery
90 l/min
max. delivery pressure
2,5 bar
Delivery at 0,5 bar
75 l/min”
Performance
The heating power is 6 kW, so the heat-up time and electrical supply capacity must be planned for this load.
Affects electrical sizing and achievable heating rates for the process.
Source
“Heating power
6 kW”
Cumplimiento
The unit is rated IP20, so it must be installed in a dry, protected location and not exposed to water or dust ingress.
IP20 limits the installation environment and requires additional enclosure protection if needed.