Introduces the CS 250HT OD unit and its main features
Specpp. 1-2
Lists technical data according to DIN 12876 standard
p. 2table, 4 rows
p. 1table, 40 rows
Technical data according to DIN 12876
Parameter
Value
Fuse (3 phase)
3x35A
min. ambient temperature
-20 °C
max. ambient temperature
50 °C
from Serial-No.:
1.0/20
Huber CS 250HT OD — Technical data according to DIN 12876
Parameter
Value
Order-No.
VDH32100077
Operating temperature range
10...22 °C
temperature set point / display
colour LED Touchscreen
Internal temperature sensor
Pt100
Temperature stability at 15°C
1,5 K
Interface digital
RS232
Safety classification
I / NFL
Kälteleistung bei Umgebungstemperatur 5°C
at 20°C
32 kW
at 15°C
28 kW
at 10°C
26 kW
Cooling power at ambient temperature 20°C
at 20°C
34 kW
at 15°C
31 kW
at 10°C
28 kW
Kälteleistung bei Umgebungstemperatur 35°C
at 20°C
32 kW
at 15°C
27 kW
at 10°C
25 kW
Kälteleistung bei Umgebungstemperatur 40°C
at 20°C
29 kW
at 15°C
26 kW
at 10°C
24 kW
Kälteleistung bei Umgebungstemperatur 50°C
at 20°C
23 kW
at 15°C
20 kW
at 10°C
19 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Refrigerant quantity
10 kg
CO2 equivalent
13,97 t
Circulation pump:
max. delivery
50 l/min
max. delivery pressure
6,5 bar
Pump connection
G3/4 male
min. filling capacity
160 l
max. filling capacity
175 l
Overall dimensions WxDxH **
1400x1000x1800 mm
Net weight
510 kg
sound pressure level +/- 4 dB(A)
78 dB(A)
** Overall dimensions WxDxH
Engineering Facts
13 facts · every claim linked to a page
Selection
The chiller's cooling capacity is rated at 34 kW at 20°C with 20°C ambient, but drops to 23 kW at 20°C with 50°C ambient, so sizing must account for the actual ambient temperature at the installation site.
Cooling capacity varies significantly with ambient temperature; selecting based on nominal 34 kW without derating for high ambient could lead to insufficient cooling.
Source
“Cooling power at ambient temperature 20°C . at 20°C 34 kW”
Installation
The unit is designed for outdoor mounting down to -20°C ambient and up to +50°C, with electrical components rated IP54, so it can be installed outdoors if those conditions are met.
Confirms outdoor installation is permissible and defines the ambient temperature limits for the installation location.
Source
“Suitable for outside mounting (min. environmental temperature -20°C)”
Installation
The operating panel is delivered as a remote control connected via a 20 m cable, so the panel can be mounted up to 20 m away from the chiller.
Affects control cabinet layout and cable routing during installation.
Source
“Delivered with remote control: operating panel connected via 20m cable”
Installation
The three-phase unit draws up to 27 A and requires 3x35 A fuses, so the electrical supply and protection must be sized accordingly.
Determines the required fuse rating and supply capacity for safe electrical installation.
Source
“Fuse (3 phase) 3x35A”
Installation
Three-phase devices with current consumption less than 63 A are supplied with a cable but without a plug, so the installer must provide a suitable plug or hard-wired connection.
Clarifies what is included in delivery and what the installer must supply for the electrical connection.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and the combination of -5% voltage with +2% frequency is not allowed, so the power supply quality must be verified.
Ensures the site power supply meets the allowed voltage/frequency envelope to avoid damage or malfunction.
Source
“-5% voltage and +2% frequency -> not allowed!”
Operation
The recommended thermofluid is a 50:50 mixture of water and monoethylene glycol, so the fluid must be prepared accordingly for proper operation.
Using the correct fluid ensures proper heat transfer and protects the system from freezing or corrosion.
Source
“Recommended thermofluid: Water - Monoethylene Glycol 50:50”
Operation
The flow rate can be adjusted via a manual bypass valve on the backside of the chiller, allowing the user to balance the circuit.
Provides a means to tune flow to the application without additional valves.
Source
“The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.”
Operation
The chiller has a minimum filling capacity of 160 l and a maximum of 175 l, so the external circuit volume must be within this range.
Ensures the system has enough fluid for proper pump operation and temperature stability.
Source
“min. filling capacity 160 l”
Performance
The circulation pump delivers up to 50 l/min at a maximum pressure of 6.5 bar, which defines the achievable flow and pressure for the application.
Used to verify the pump can meet the required flow and pressure drop of the external circuit.
Source
“max. delivery 50 l/min”
Safety
The refrigerant is R-449A with safety group A1 and GHS hazard H280, and the charge is 10 kg (13.97 t CO2 equivalent), so refrigerant handling and leak detection must comply with applicable regulations.
Informs about refrigerant safety classification and environmental impact for compliance and safety planning.
Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280)”
Cumplimiento
The equipment emits Class A, Group 1 disturbances per EN55011, so it may cause radio interference in residential areas and appropriate precautions may be needed.
Important for EMC compliance and installation in environments sensitive to electromagnetic interference.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance
The unit has a digital level indicator and an atmospheric open tank, so the fluid level should be monitored and maintained within the specified filling capacity.
Helps ensure proper fluid level for reliable operation and prevents pump cavitation or overflow.