Huber HS 120

Contenido

  1. Overview p. 1
    General description and key features of the HS 120
  2. Spec p. 1
    Technical data and performance parameters table
  3. p. 1 table, 18 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range35...85 °C
temperature set point / displaycolour LED Touchscreen
Internal temperature sensorPt100
Temperature stability at 30°C+/-0,1 K
Safety classificationI / NFL
Heating power12 kW
Circulation pump: max. delivery87 l/min
max. delivery pressure3,7 bar
Pump connectionG3/4 male
min. filling capacity24 l
max. filling capacity27 l
Overall dimensions WxDxH **580x900x830 mm
Net weight130 kg
Power supply (3 Phase)460V 3~ 60Hz
max. current (3 Phase)20 A
Fuse (3 phase)3x25A
Degree of ProtectionIP20
from Serial-No.215224

** Overall dimensions WxDxH

Engineering Facts

12 facts · every claim linked to a page
Selection

The unit's operating temperature range is 35 to 85 °C, so it cannot be used for sub-ambient or near-ambient cooling duties.

Defines the process window and excludes applications requiring temperatures below 35 °C.

Source
“Operating temperature range 35...85 °C”
Performance

Heating power is 12 kW, which sets the maximum heat-up rate and the achievable heating capacity at the process load.

Needed to size the heater against the thermal load and required ramp times.

Source
“Heating power 12 kW”
Performance

The circulation pump delivers up to 87 l/min at a maximum delivery pressure of 3.7 bar, which bounds the achievable flow in the external circuit.

Used to check whether the pump can overcome the pressure drop of the customer loop at the required flow.

Source
“max. delivery 87 l/min max. delivery pressure 3,7 bar”
Installation

The pump connection is G3/4 male, so the customer piping must be adapted to this thread size.

Determines the fittings and adapters required for connecting the unit to the process loop.

Source
“Pump connection G3/4 male”
Installation

The unit requires a 460 V 3-phase 60 Hz supply with a maximum current of 20 A and 3x25 A fusing.

Dictates the site power availability, cable sizing and upstream protection for the installation.

Source
“Power supply (3 Phase) 460V 3~ 60Hz max. current (3 Phase) 20 A Fuse (3 phase) 3x25A”
Installation

The unit has IP20 degree of protection, so it must be installed in a dry indoor location protected from water and dust ingress.

Restricts the installation environment and rules out washdown or outdoor siting.

Source
“Degree of Protection IP20”
Installation

The unit weighs 130 kg net with overall dimensions of 580x900x830 mm, which must be considered for floor loading and access routes.

Needed for planning the installation space, lifting equipment and floor load capacity.

Source
“Overall dimensions WxDxH ** 580x900x830 mm Net weight 130 kg”
Operation

The tank holds between 24 l minimum and 27 l maximum of filling capacity, so the expansion volume must be respected during filling.

Prevents overfilling or running dry and defines the usable expansion volume of the open tank.

Source
“min. filling capacity 24 l max. filling capacity 27 l”
Safety

The unit is classified as safety class I / NFL according to DIN 12876, which governs the safety requirements for the installation and operation.

Determines the applicable safety measures and compliance requirements for the installation.

Source
“Safety classification I / NFL”
Operation

The feed pump has integrated overtemperature protection, which safeguards the heater against overheating during operation.

Confirms a built-in safety function that reduces the need for external over-temperature protection.

Source
“Powerful feed pump with integrated overtemperature protection.”
Operation

The flow rate can optionally be adjusted via a manual bypass valve on the backside of the heater.

Allows the installer to trim the flow to the process requirement without additional hardware.

Source
“Optional the flow rate could be adjusted via the manual bypass valve on the backside of the heater.”
Operation

The control unit provides separate operating options for the feed pump and the cooling unit with convenient setpoint adjustment.

Affects how the operator configures pump and cooling independently during commissioning and operation.

Source
“Separate operating option for the feed pump and the cooling unit with convenient adjustment of the desired setpoint.”