Induztr
Resumen técnico Visor de documentos Modelos relacionados

KISS 212B Immersion Circulator

Contenido

  1. Overview p. 1
    Product introduction and main features of KISS 212B
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 1 table, 36 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range25...200 °C
with water cooling20...200 °C
with refrigerator-30...200 °C
Temperature stability at 70°C0,05 K
temperature set point / displaydigital
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Interface digitalUSB (Device), RS232
InterfaceAlarm message
optic, acoustic
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
max. delivery14 l/min
max. delivery pressure0,25 bar
max. delivery (suction)10,5 l/min
max. delivery pressure (suction)0,17 bar
Pump connenction (optional)M16x1 male
Bath volume10,5 l
min. filling capacity5,5 l
Height of bath opening205 mm
Width bath opening WxD/ bath depth290x152/ 150 mm
Overall dimensions WxDxH **350x375x375 mm
Net weight11 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryN/A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.393790

** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.

Engineering Facts

13 facts · every claim linked to a page
Selection

The operating temperature range depends on the cooling method: 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator.

Determines whether the bath can reach sub-ambient or near-ambient setpoints and what cooling utility or chiller must be specified.

Source
“Operating temperature range 25...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Performance

Heating power scales with supply voltage from 1.5 kW at 200 V to 2.1 kW at 240 V, so available heating capacity depends on the actual site voltage.

Affects heat-up time and whether the unit meets process heating requirements at the customer's supply voltage.

Source
“Heating power at 240V 2,1 kW Heating power at 230V 2 kW Heating power at 220V 1,8 kW Heating power at 208V 1,6 kW Heating power at 200V 1,5 kW”
Installation

The unit requires a 200–240 V single- or two-phase 50/60 Hz supply with a minimum 10 A fuse and maximum 16 A fuse.

Needed to size the branch circuit, protective device, and confirm the site supply matches the unit.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation

The enclosure is rated IP20, so it must be installed in a dry location protected from water ingress and access to live parts.

Dictates the installation environment and prohibits washdown or wet locations.

Source
“Degree of Protection IP20”
Installation

Ambient temperature must stay between 5 °C and 40 °C for proper operation.

Room conditions outside this range can affect performance, safety, and warranty.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Selection

The bath holds 10.5 l total but requires a minimum filling of 5.5 l, and the bath opening is 290 x 152 mm with 150 mm depth.

Determines whether sample vessels and immersion coolers fit and how much thermofluid must be ordered.

Source
“Bath volume 10,5 l min. filling capacity 5,5 l Height of bath opening 205 mm Width bath opening WxD/ bath depth 290x152/ 150 mm”
Selection

The pressure/suction pump delivers up to 14 l/min at 0.25 bar pressure and 10.5 l/min at 0.17 bar suction.

Used to check whether the pump can supply an external loop or jacketed vessel with adequate flow and head.

Source
“max. delivery 14 l/min max. delivery pressure 0,25 bar max. delivery (suction) 10,5 l/min max. delivery pressure (suction) 0,17 bar”
Safety

The unit is Safety Class III / FL with adjustable overtemperature protection per DIN 12876, providing fail-safe protection against overheating.

Important for risk assessment and for processes requiring fail-safe overtemperature protection.

Source
“Safety classification III / FL”
Operation

The plastic pump is not resistant to acetone or polyglycols, so water with glycerin or ethylene glycol is recommended for freeze protection.

Prevents chemical attack and pump failure when selecting the thermofluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Cumplimiento

Voltage tolerance is ±10% only if frequency tolerance does not run in the opposite direction; for example, -10% voltage with +3% frequency is not allowed.

Site power quality must be checked; combined out-of-tolerance voltage and frequency can damage the unit or void compliance.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation

An external branch circuit protection device according to UL 489 is required.

Must be included in the electrical installation design for UL-compliant installations.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Single- and two-phase devices ship with a country-specific plug that must be specified at order; three-phase devices under 63 A ship with cable but no plug.

Avoids ordering delays and ensures the correct plug for the destination country.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation

The controller has an autostart function after power failure, which may restart heating and pumping automatically.

Affects process safety and restart behavior after a power interruption; may need to be considered in risk assessment.

Source
“Autostart function for power failure”