TC45E-F Refrigerated/Heating Circulator Technical Data

Contenido

  1. Overview p. 1
    Introduces the TC45E-F unit and its main features
  2. Spec pp. 1-2
    Technical data table according to DIN 12876
  3. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterTC45E-F
Order-No.3003.0004.99
Operating temperature range-45...100 °C
Temperature stability at -10°C0,5 K
temperature set point / displaydigital
Temperature indicationdigital
Sensor external connectionPt100
Cooling power
at 0°C0,24 kW
at -20°C0,18 kW
at -30°C0,1 kW
at -40°C0,05 kW
Safety classificationI / NFL
Refrigeration machineair-cooled, natural
refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,041 kg
CO2 equivalent0,0 t
Nominal diameter probe13 mm
Length of probe600 mm
Length flexible connection1050 mm
Overall dimensions WxDxH **190x295x360 mm
Net weight16 kg
sound pressure level +/- 4 dB(A)54 dB(A)
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current1,4 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.S394130

** Overall dimensions WxDxH

Engineering Facts

12 facts · every claim linked to a page
Selection

The unit covers a working temperature range from -45 to 100 °C, so it must be matched to the process setpoint window.

Defines whether the cooler can reach and hold the required process temperature.

Source
“Operating temperature range -45...100 °C”
Performance

Cooling capacity drops sharply at low temperatures, from 0.24 kW at 0 °C to only 0.05 kW at -40 °C, so low-temperature duty needs careful load sizing.

Prevents oversizing expectations and ensures the heat load is removable at the coldest setpoint.

Source
“at 0°C 0,24 kW at -20°C 0,18 kW at -30°C 0,1 kW at -40°C 0,05 kW”
Safety

The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which drives ventilation and ignition-source requirements.

Affects room classification, ventilation and safe installation location.

Source
“R-290 (A3, H220)”
Installation

The unit is air-cooled and needs ambient temperature between 5 and 40 °C, so the installation room must stay within this band.

Out-of-range ambient air reduces cooling capacity and can trip the refrigeration circuit.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

The power supply must be 208-240 V single or two-phase at 50/60 Hz with a fuse between 10 A and 16 A.

Determines the electrical feed, breaker sizing and plug configuration at the site.

Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz max. current 1,4 A min. Fuse 10A max. Fuse 16A”
Installation

Voltage and frequency tolerances are linked: ±5 % voltage is only allowed with ±2 % frequency, and the combination -5 % voltage with +2 % frequency is not allowed.

Prevents compressor damage on sites with unstable mains supply.

Source
“-5% voltage and +2% frequency -> not allowed!”
Performance

Rated cooling output is valid at 20 °C room temperature and falls if ambient temperature rises.

Requires derating of capacity when the unit is installed in a warm room.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Operation

The refrigeration unit runs continuously and temperature is controlled by a solenoid valve in the cooling circuit, so the compressor is not cycled on/off.

Explains control behaviour and why the valve and sensor wiring must be correct.

Source
“The refrigeration unit works continuously. The temperature control actuates a solenoid valve in the cooling circuit.”
Installation

The flexible probe is 600 mm long with a 13 mm diameter and a 1050 mm flexible connection, which constrains how the probe reaches the bath or vessel.

Must be checked against the geometry of the vessel and the mounting position.

Source
“Nominal diameter probe 13 mm Length of probe 600 mm Length flexible connection 1050 mm”
Cumplimiento

The equipment is EMC class A group 1, so it is intended for industrial environments and may cause interference in residential areas.

Affects site selection and possible filtering when installed near sensitive electronics.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

Avoids a missing or wrong plug on delivery and delays at commissioning.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Selection

The unit is classified as pressure equipment category Art. 4.3 PED, which sets the regulatory regime for the refrigerant circuit.

Relevant for documentation and conformity assessment of the installed equipment.

Source
“Pressure equipment category Art. 4.3 PED”