Technical data according to DIN 12876 for the unit
Engineering Facts
12 facts · every claim linked to a page
Selection
The chiller is rated for 208-230 V single/two-phase 60 Hz supply with 14 A max current and a 16 A fuse, so the electrical supply must be sized accordingly.
Determines breaker, cable and supply configuration before ordering and installation.
Source
“Power supply requirement 208-230V 1~/2~ 60Hz”
Installation
The unit requires ambient temperature between 5 °C and 40 °C, and cooling capacity is valid at 20 °C room temperature and may drop if ambient rises.
Room ventilation and ambient conditions must be planned to avoid derated cooling.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Safety
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, so ignition sources and ventilation must be considered.
Flammable refrigerant classification affects site safety assessment and installation location.
Source
“R-290 (A3, H220)”
Safety
The plastic pump is not resistant to acetone and polyglycols, so these fluids must not be used without a more resistant pump option.
Prevents chemical attack and failure of the circulation pump when selecting heat transfer fluid.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Guides correct heat transfer fluid preparation for sub-zero operation.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and combining -5% voltage with +2% frequency is not allowed.
Ensures the site power quality is within the permitted combined voltage/frequency envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Single/two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.
Avoids missing or wrong plug type on delivery and simplifies electrical connection.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Cumplimiento
The equipment is EMC classified as Class A, Group 1 to EN55011, which has implications for use in residential environments.
Class A equipment may require additional EMC measures in residential or sensitive installations.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The unit has IP20 protection, so it must be installed in a dry indoor location protected from water ingress.
Prevents installation in wet or outdoor areas where IP20 is insufficient.
Source
“Degree of Protection IP20”
Performance
Cooling capacity is 2.5 kW at 15 °C, 2 kW at 0 °C and 1.2 kW at -10 °C, so capacity must be checked at the actual operating temperature.
Selection must use capacity at the required process temperature, not the maximum rating.
Source
“at 15°C 2,5 kW at 0°C 2 kW at -10°C 1,2 kW”
Installation
The pump connection is G3/4 male, so hoses and fittings must match this thread.
Ensures correct hose coupling and avoids adapters or leaks at commissioning.
Source
“Pump connection G3/4 male”
Operation
The controller has an autostart function for power failure, so the unit may restart automatically after power returns.
Affects risk assessment and process safety after a power interruption.