Introduces the Unichiller 060Tw-H chiller and its main features
Specpp. 1-2
Performance, temperature range, and operating data per DIN 12876
Specp. 2
Physical dimensions, weight, and process connection details
p. 2table, 15 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
Consumption at water 15°C, flow -20°C
380 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
5,9 l
Volume of expansion
7,5 l
Overall dimensions WxDxH **
600x600x1450 mm
Net weight
210 kg
Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
13 A
Fuse (3 phase)
3x16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S425010
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Unichiller 060Tw-H (Order-No. 3026.0119.01)
Parameter
Value
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power
2 kW
Cooling power at 15°C
6 kW
Cooling power at 0°C
5 kW
Cooling power at -10°C
3,1 kW
Cooling power at -20°C
1,7 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
1,15 kg
CO2 equivalent
1,6 t
Circulation pump
C3
max. delivery
80 l/min
max. delivery pressure
5,6 bar
Delivery at 0,35 bar
79 l/min
Delivery at 0,5 bar
76 l/min
Delivery at 1,0 bar
71 l/min
Delivery at 1,5 bar
66 l/min
Delivery at 2,0 bar
61 l/min
Delivery at 3,0 bar
49 l/min
Delivery at 4,0 bar
35 l/min
Delivery at 5,0 bar
16 l/min
Pump connection
G1 1/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 15°C
870 l/h
Consumption at water 15°C, flow 0°C
730 l/h
Consumption at water 15°C, flow -10°C
580 l/h
Engineering Facts
13 facts · every claim linked to a page
Installation
The chiller is water-cooled and requires a cooling water supply with a differential pressure between 1 and 6 bar.
Determines whether the site's cooling water loop can supply adequate pressure without overpressurizing the unit.
Source
“min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar”
Installation
The unit is designed for cooling water intake between 5°C and 20°C, and pipes must be insulated if water temperature can fall below the condensation point.
Cooling water outside this range will degrade performance or cause condensation, affecting installation design.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.
Engineers must size the cooling water supply for worst-case water temperature, not just nominal conditions.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Water chemistry must be checked to avoid corrosion or degradation of these wetted materials.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation
Three-phase devices with current consumption below 63 A are delivered with a cable but without a plug.
The installer must provide and fit a suitable plug or hard-wire the supply, which affects electrical installation planning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Cumplimiento
Voltage tolerance is ±5% only when frequency tolerance is simultaneously ±2%, and certain combinations are not allowed.
Supply quality must be verified against these limits to avoid invalidating operation or damaging the unit.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety
The unit has safety classification III / FL according to DIN 12876 and adjustable overtemperature protection.
Confirms the built-in safety concept and that overtemperature protection must be configured for the application.
Source
“With
adjustable overtemperature protection according to DIN 12876.”
Selection
The refrigerant is R-449A with GWP 1397 and a CO2 equivalent of 1.6 t, which matters for environmental and regulatory reporting.
Refrigerant charge and GWP affect compliance obligations and environmental impact assessment during selection.
Source
“Global Warming Potential (GWP)
1397”
Operation
The unit is intended for externally closed applications and has an atmospheric open expansion tank.
The external circuit must be closed; open-loop or unsealed external circuits are not suitable.
Source
“For externally closed applications.”
Installation
The unit has protection degree IP20 and must be operated within an ambient temperature of 5°C to 40°C.
Site ambient conditions and ingress protection must match the installation location.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Selection
The pump connection is G1 1/4 male and the cooling water connection is G1/2 male.
Thread sizes determine the hoses and fittings needed for installation.
Source
“Pump connection
G1 1/4 male
Cooling water connection
G1/2 male”
Performance
Cooling power is 6 kW at 15°C, 5 kW at 0°C, 3.1 kW at -10°C and 1.7 kW at -20°C.
Cooling capacity at the actual process temperature must be checked against the heat load.
Source
“at 15°C
6 kW
at 0°C
5 kW
at -10°C
3,1 kW
at -20°C
1,7 kW”
Installation
The unit requires a 400 V 3~ 50 Hz supply with a 3x16 A fuse and maximum current of 13 A.
Electrical supply, protection and cabling must be sized accordingly.
Source
“Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
13 A
Fuse (3 phase)
3x16A”