Unichiller 250T Technical Data

Contenido

  1. Overview p. 1
    Introduces the Unichiller 250T model and its intended use
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 1 table, 35 rows

Unichiller 250T — Technical data according to DIN 12876

CharacteristicValue
Order-No.3057.0001.01
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,5 K
Safety classificationI / NFL
Cooling power at 15°C25 kW
Cooling power at 0°C18 kW
Cooling power at -10°C11 kW
Cooling power at -20°C6 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pump:D3
max. delivery210 l/min
max. delivery pressure4,7 bar
Delivery at 1,1 bar202 l/min
Delivery at 1,5 bar193 l/min
Delivery at 2,0 bar179 l/min
Delivery at 2,5 bar159 l/min
Delivery at 3,0 bar140 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar77 l/min
Delivery at 4,5 bar31 l/min
Pump connectionG1 1/4 male
min. filling capacity14 l
Volume of expansion39 l
Overall dimensions WxDxH **905x2172x1900 mm
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C

** см. сноску в оригинале (Overall dimensions WxDxH **)

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller is air-cooled, so the installation site must provide adequate air flow and respect the specified space requirements.

Air-cooled units reject heat to the room; insufficient clearance or ventilation will reduce cooling capacity and can cause overheating.

Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Installation

The unit requires a 3-phase 400 V 3~ 50 Hz supply, and three-phase devices below 63 A are delivered with cable but without a plug.

The installer must plan for a suitable 3-phase supply and provide the correct plug/connection on site.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances are +/−5% and +/−2% respectively, but certain combinations such as −5% voltage with +2% frequency are not allowed.

Supply quality must be checked before commissioning; an out-of-tolerance combination can damage the compressor or pump.

Source
“-5% voltage and +2% frequency -> not allowed!”
Performance

Cooling capacity is rated at 20 °C room temperature and will drop if the ambient temperature rises.

Selection must account for the actual ambient temperature at the installation site, not just the nominal 25 kW at 15 °C.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Operation

The unit is intended for externally closed applications, so the external circuit must be closed and compatible with the stainless-steel pump and expansion tank.

Using an open external circuit can cause evaporation, contamination, or overflow of the atmospheric expansion tank.

Source
“For externally closed applications.”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280, and the unit has a GWP of 1397.

Refrigerant handling, leak detection, and disposal must follow H280 (pressurised gas) precautions, and high GWP affects environmental compliance and service planning.

Source
“R-449A (A1, H280)”
Cumplimiento

The equipment is EMC Class A, Group 1, which is intended for industrial environments and may cause interference in residential areas.

Class A equipment may not be suitable for residential or light-industrial locations without additional filtering or separation.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

The pump connection is G1 1/4 male, and a matching hose coupling is included.

External piping and hoses must be ordered or adapted to G1 1/4 male threads; the included coupling covers only one connection.

Source
“Hose coupling for G1 1/4 male”
Operation

The minimum filling capacity is 14 l and the expansion volume is 39 l, which sets the minimum heat-transfer fluid charge and expansion allowance.

Underfilling can cause pump cavitation and poor temperature control; the expansion volume must be respected when sizing the external loop.

Source
“min. filling capacity”
Operation

The pump delivers 210 l/min maximum and 4.7 bar maximum pressure, with flow falling as pressure rises.

The external circuit pressure drop must be checked against the pump curve; high-resistance loops will reduce flow below the required rate.

Source
“max. delivery pressure”
Installation

The unit has IP20 protection and must be installed where ambient temperature is between 5 °C and 40 °C.

IP20 offers no water protection, and operation outside 5–40 °C ambient may impair cooling or damage components.

Source
“min. ambient temperature”
Selection

The temperature stability at −10 °C is 0.5 K, which defines the achievable control accuracy for sensitive processes.

Processes requiring tighter than ±0.5 K stability need a different model or additional control measures.

Source
“Temperature stability at -10°C”