Unichiller 300Tw Technical Datasheet

Contenido

  1. Overview p. 1
    Introduces the Unichiller 300Tw chiller and its main features
  2. Spec p. 2
    Detailed performance and technical data per DIN 12876
  3. p. 2 table, 11 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Overall dimensions WxDxH **760x840x1767 mm
Net weight385 kg
sound pressure level +/- 4 dB(A)65 dB(A)
Power supply (3 Phase)460V 3~ 60Hz
max. current (3 Phase)23 A
Fuse (3 phase)3x25A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S582993

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at 15°C30 kW
Cooling power at 0°C18 kW
Cooling power at -10°C13 kW
Cooling power at -20°C8 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity3 kg
CO2 equivalent4,2 t
UN-numberUN 2857
Circulation pumpD3
max. delivery210 l/min
max. delivery pressure4,7 bar
Delivery at 1,1 bar202 l/min
Delivery at 1,5 bar193 l/min
Delivery at 2,0 bar179 l/min
Delivery at 2,5 bar159 l/min
Delivery at 3,0 bar140 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar77 l/min
Delivery at 4,5 bar31 l/min
Pump connectionG1 1/4 male
Cooling water connectionG3/4 male
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity15 l
Volume of expansion22 l

Order-No.: 3029.0058.01; Unichiller 300Tw

Engineering Facts

13 facts · every claim linked to a page
Installation

The chiller requires a three-phase 460 V 60 Hz supply with 23 A max current and 3x25 A fusing, so the site electrical infrastructure must be sized accordingly.

Determines breaker, cable and panel capacity needed before installation.

Source
“Power supply (3 Phase) 460V 3~ 60Hz”
Installation

External branch circuit protection per UL 489 is mandatory, so a listed breaker must be provided in the supply circuit.

Affects electrical design and compliance sign-off for the installation.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Three-phase devices under 63 A are delivered with cable but without a plug, so the customer must supply and fit the plug.

Impacts procurement and site wiring scope at commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Cooling water must be supplied at 1–6 bar with at least 1 bar differential pressure, so the facility water loop must meet this window.

Insufficient differential pressure will prevent rated cooling performance.

Source
“min. cooling water differential pressure 1 bar”
Operation

Cooling water inlet must stay between 5 °C and 20 °C, and pipes must be insulated if water can drop below the condensation point.

Out-of-range water temperature reduces cooling power and risks condensation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Rated cooling power (30 kW at 15 °C down to 8 kW at -20 °C) is valid only at 20 °C room, 15 °C water inlet and 1 bar differential, so actual capacity will differ otherwise.

Prevents overestimating capacity when site conditions deviate from rating basis.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Selection

The unit uses R-449A refrigerant with GWP 1397 and 4.2 t CO2 equivalent, which matters for refrigerant regulations and leak reporting.

Affects environmental compliance and service obligations for the refrigerant charge.

Source
“CO2 equivalent 4,2 t”
Safety

The refrigerant is classified A1 with H280 hazard statement, so handling and ventilation precautions apply during service.

Informs service safety procedures and labeling requirements.

Source
“R-449A (A1, H280)”
Installation

The unit is IP20 and must be installed where ambient stays between 5 °C and 40 °C, so a clean, temperature-controlled indoor location is required.

Rules out wet, dusty or uncontrolled outdoor environments.

Source
“Degree of Protection IP20”
Installation

The unit weighs 385 kg and measures 760x840x1767 mm, so floor loading and access routes must be planned for rigging.

Needed for transport, lifting and floor-load planning.

Source
“Net weight 385 kg”
Operation

Voltage and frequency tolerances are ±5% and ±2% but must not combine in the same direction, so supply quality must be verified.

Prevents operation outside the allowed supply envelope that could damage the compressor.

Source
“-5% voltage and +2% frequency -> not allowed!”
Cumplimiento

The equipment is EMC Class A Group 1, so it may cause radio interference in residential areas and is intended for industrial use.

Affects site selection and EMC coordination with sensitive equipment.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality may be used.

Guides water treatment to avoid corrosion and fouling of the circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”