Unichiller GL 750w

Contenido

  1. Overview p. 1
    Product introduction and main features of the chiller
  2. Spec pp. 1-2
    Technical data according to DIN 12876
  3. p. 2 table, 3 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/26

Unichiller GL 750w — Technical data according to DIN 12876

ParameterValue
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at 20°C75 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-744 (A1, H280)
Global Warming Potential (GWP)1
UN-numberUN 2857
max. delivery320 l/min
max. delivery pressure4 bar
Delivery at 0,5 bar305 l/min
Delivery at 1,0 bar285 l/min
Delivery at 1,5 bar260 l/min
Delivery at 2,0 bar240 l/min
Delivery at 2,5 bar215 l/min
Delivery at 3,0 bar185 l/min
Delivery at 3,5 bar150 l/min
Delivery at 4,0 bar110 l/min
Delivery at 4,5 bar50 l/min
Pump connectionG1 1/4 male
Cooling water connectionG1 1/4 male
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity17,5 l
Volume of expansion50 l
Overall dimensions WxDxH **918x963x1771 mm
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryII
Degree of ProtectionIP20

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Installation

The chiller requires cooling water with a minimum differential pressure of 1 bar and must not exceed 6 bar, so the site water supply must be verified before installation.

Insufficient differential pressure will prevent rated cooling performance and may trip the unit.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit is water-cooled and needs a cooling water supply between 5°C and 20°C; below the condensation point the cooling water pipes inside the unit must be insulated.

Out-of-range cooling water temperature reduces cooling power and risks condensation inside the unit.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Rated output data are valid only at 20°C room temperature, 15°C cooling water inlet and 1 bar differential pressure, so actual capacity will differ at other site conditions.

Selection must account for derating when site conditions deviate from the reference point.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Incompatible water chemistry can corrode or foul the circuit and void performance.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and the combination of -5% voltage with +2% frequency is not allowed.

Supply quality must be checked against this asymmetric tolerance before commissioning.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

The unit has a leakage current greater than 3.5 mA, which affects protective earthing and RCD selection.

Earth connection and residual current device rating must be sized accordingly.

Source
“Attention: leakage current > 3,5mA”
Installation

Three-phase devices below 63 A are delivered with a cable but without a plug, so the customer must provide the plug and connection.

Electrical installation scope and materials must be planned by the site.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The unit is IP20 and must be installed indoors protected from water ingress.

IP20 limits installation locations and requires protection from splashing or dripping water.

Source
“Degree of Protection IP20”
Selection

The refrigerant is R-744 (CO2) with GWP 1 and is classified A1/H280, which affects refrigerant handling and safety classification.

Refrigerant safety class and GWP influence site safety requirements and environmental compliance.

Source
“R-744 (A1, H280)”
Selection

The pump delivers 320 l/min maximum but only 110 l/min at 4.0 bar, so the system curve must be matched to the required pressure.

External circuit pressure drop determines the achievable flow and must be checked during selection.

Source
“Delivery at 4,0 bar 110 l/min”
Installation

The minimum filling capacity is 17.5 l and the expansion volume is 50 l, which sets the minimum external loop volume and expansion allowance.

External loop volume and expansion vessel sizing must respect these limits.

Source
“min. filling capacity 17,5 l”
Operation

The ambient temperature range is 5°C to 40°C, so the installation room must be kept within these limits.

Ambient conditions outside the range affect cooling performance and reliability.

Source
“max. ambient temperature 40 °C”
Cumplimiento

The equipment is EMC Class A Group 1, which is intended for industrial environments and may cause interference in residential areas.

Class A equipment has restrictions on use in residential environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety

The unit has adjustable overtemperature protection according to DIN 12876, which must be set for the application.

Overtemperature protection setting is a safety requirement for the connected process.

Source
“With adjustable overtemperature protection according to DIN 12876.”