Product introduction and main features of the chiller
Specpp. 1-2
Technical data and performance specifications per DIN 12876
p. 2table, 7 rows
p. 1table, 39 rows
Technical data according to DIN 12876
Parameter
Value
max. current
12,5 A
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.1/24
Technical data according to DIN 12876 — Unichiller P012w-H OLÉ (Order-No. 3009.0242.98)
Parameter
Value
Operating temperature range
-20...100 °C
temperature set point / display
digital
Internal temperature sensor
Pt100
Resolution of display
0,1 K
Interface digital
USB (Device), RS232
Interface
Temperature stability at -10°C
0,2 K
Alarm message
optic, acoustic
Safety classification
III / FL
Heating power
1,6 kW
Cooling power
at 15°C
1,2 kW
at 0°C
1 kW
at -10°C
0,7 kW
at -20°C
0,25 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,16 kg
CO2 equivalent
0,2 t
Circulation pump
B
at 0,5 bar
21 l/min
at 1,0 bar
17 l/min
at 1,5 bar
11 l/min
at 2,0 bar
6 l/min
max. delivery
25 l/min
max. delivery pressure
2,5 bar
Pump connection
G3/4 male
Consumption at water 15°C, flow 15°C
60 l/h
Consumption at water 15°C, flow 0°C
57 l/h
Consumption at water 15°C, flow -10°C
48 l/h
Cooling water connection
G1/2 male
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
3,8 l
expansion tank
1,7 l
Overall dimensions WxDxH **
350x496x622 mm
Net weight
61 kg
Power supply requirement
208V 2~ 60Hz
** Overall dimensions WxDxH
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller requires cooling water with a minimum differential pressure of 1 bar and a maximum pressure of 6 bar, so the site water supply must be sized accordingly.
Ensures the cooling water circuit can deliver adequate flow without overpressurizing the unit.
Source
“min. cooling water differential pressure 1 bar”
Installation
The unit is water-cooled and must be connected to a cooling water supply with G1/2 male connections.
Determines the piping and fittings needed at the installation site.
Source
“Cooling water connection G1/2 male”
Operation
Cooling water inlet temperature must be between 5°C and 20°C; below the condensation point the cooling water pipes in the unit must be insulated.
Operating outside this range risks condensation and reduced cooling performance.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.
Allows the engineer to account for derating and water supply capacity at higher site water temperatures.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Safety
The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided or a more resistant pump specified.
Prevents chemical attack and premature pump failure when selecting the heat transfer fluid.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Guides the choice of heat transfer fluid for sub-zero operation.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Cumplimiento
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Ensures the site power supply stays within the unit's acceptable operating envelope.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices drawing less than 63 A are supplied with a cable but without a plug, so a suitable plug must be provided on site.
Affects electrical installation planning and the need to source a matching plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The unit has IP20 protection, so it must be installed in a dry indoor environment protected from water ingress.
Restricts installation location and environmental conditions.
Source
“Degree of Protection IP20”
Operation
Ambient temperature must be kept between 5°C and 40°C for proper operation.
Defines the allowable room conditions at the installation site.
Source
“min. ambient temperature 5 °C”
Selection
The refrigerant is R-449A with GWP 1397 and a charge of 0.16 kg, giving a CO2 equivalent of 0.2 t.
Relevant for refrigerant regulations, leak reporting thresholds, and environmental compliance.
Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280)”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water should be used.
Prevents corrosion and scaling by matching water quality to wetted materials.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”