Unichiller P025-H OLÉ Datasheet

Contenido

  1. Overview p. 1
    Product introduction and general description of the Unichiller P025-H OLÉ
  2. Spec pp. 1-2
    Technical data, performance figures and operating parameters per DIN 12876
  3. p. 2 table, 3 rows
  4. p. 1 table, 39 rows

Technical data according to DIN 12876

ParameterValue
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S612190

Unichiller P025-H OLÉ — Technical data according to DIN 12876

ParameterValue
Operating temperature range-10...100 °C
temperature set point / displaydigital
Internal temperature sensorPt100
Resolution of display0,1 K
Interface digitalUSB (Device), RS232
Interface
Temperature stability at -10°C0,2 K
Alarm messageoptic, acoustic
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at 15°C2,5 kW
Cooling power at 0°C2 kW
Cooling power at -10°C1,2 kW
Refrigeration machineair-cooled, natural
refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,23 kg
CO2 equivalent0,0 t
Circulation pumpB
at 0,5 bar21 l/min
at 1,0 bar17 l/min
at 1,5 bar11 l/min
at 2,0 bar6 l/min
max. delivery25 l/min
max. delivery pressure2,5 bar
Pump connectionG3/4 male
min. filling capacity3,8 l
expansion tank1,7 l
Overall dimensions WxDxH **460x590x743 mm
Net weight81 kg
sound pressure level +/- 4 dB(A)64 dB(A)
Power supply requirement220-240V 1~ 50Hz
max. current16 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -10 to 100 °C, so it can serve both sub-ambient cooling and moderate heating tasks.

Defines whether the unit can cover the required process temperature window before any other sizing is done.

Source
“Operating temperature range -10...100 °C”
Selection

Cooling capacity falls from 2.5 kW at 15 °C to 1.2 kW at -10 °C, so the unit must be sized against the actual operating point, not the nominal figure.

Prevents oversizing assumptions at low temperature where available cooling is roughly halved.

Source
“Cooling power at 15°C 2,5 kW at 0°C 2 kW at -10°C 1,2 kW”
Safety

The refrigerant is R-290 (propane, A3 flammability class, H220), which imposes flammable-refrigerant handling and ventilation requirements on the installation.

A3 refrigerants require ignition-source control and room ventilation planning, and affect service and compliance approvals.

Source
“R-290 (A3, H220)”
Installation

The unit draws up to 16 A on a 220–240 V single-phase supply and is protected by a 16 A fuse, so the site circuit must be dedicated and adequately rated.

Determines cable cross-section, breaker rating and whether existing site power is sufficient.

Source
“Power supply requirement 220-240V 1~ 50Hz max. current 16 A Fuse 16 A”
Installation

The enclosure is rated IP20, so the chiller must be kept in a dry, clean location and cannot be washed down or exposed to splashing.

Constrains placement in the lab or plant and rules out wet or washdown environments.

Source
“Degree of Protection IP20”
Installation

The pump connection is G3/4 male, so hoses or adapters must match this thread size at the circuit interface.

Avoids ordering incompatible fittings and delays during commissioning.

Source
“Pump connection G3/4 male”
Performance

Pump delivery drops from 21 l/min at 0.5 bar to 6 l/min at 2.0 bar, so high-pressure-drop loops must be checked against the actual pump curve.

Ensures the required flow is still achievable in a restrictive external circuit.

Source
“at 0,5 bar 21 l/min at 1,0 bar 17 l/min at 1,5 bar 11 l/min at 2,0 bar 6 l/min”
Operation

Cooling capacity is specified at 20 °C room temperature and may drop as ambient rises, so the installation room must stay within the allowed ambient window.

Prevents underperformance when the chiller is placed in a warm or poorly ventilated room.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The permitted ambient temperature range is 5 to 40 °C, which sets the boundary for the room where the air-cooled unit is installed.

Defines site ventilation and climate requirements before installation.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant pump specified.

Prevents chemical attack and premature pump failure when selecting the heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection the manufacturer recommends water mixed with glysantine or ethylene glycol rather than pure water.

Guides the correct heat transfer fluid choice for sub-zero operation.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

Voltage and frequency tolerances are coupled: ±5 % voltage is only allowed with ±2 % frequency, and the combination -5 % voltage with +2 % frequency is not permitted.

Prevents operation on unstable or generator-supplied mains that could damage the compressor.

Source
“-5% voltage and +2% frequency -> not allowed!”
Cumplimiento

The unit is EMC-classified as Class A, Group 1 to EN55011, so it is intended for industrial environments rather than residential ones.

Affects where the chiller may be installed and what EMC measures are needed nearby.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single- and two-phase units are delivered with a power cable and country-specific plug, which must be specified at ordering.

Ensures the correct plug type is supplied and avoids rework at commissioning.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”