Introduces the Unichiller P025w chiller and its main features
Specpp. 1-2
Performance, temperature, and dimensional data per DIN 12876
p. 2table, 11 rows
p. 1table, 31 rows
Technical data according to DIN 12876
Parameter
Value
Overall dimensions WxDxH **
420x487x579 mm
Net weight
68 kg
sound pressure level +/- 4 dB(A)
60 dB(A)
Power supply requirement
208-230V 1~/2~ 60Hz
max. current
12,5 A
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.1/24
** Please respect space requirements. See operating conditions at www.huber-online.com
Unichiller P025w — Technical data according to DIN 12876
Characteristic
Value
Order-No.
3052.0031.01
Operating temperature range
-10...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at ambient temperature +20°C at 15°C
2,5 kW
Cooling power at ambient temperature +20°C at 0°C
2 kW
Cooling power at ambient temperature +20°C at -10°C
1,2 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,4 kg
CO2 equivalent
0,6 t
Circulation pump
B
max. delivery
25 l/min
max. delivery pressure
2,5 bar
Delivery at 0,5 bar
21 l/min
Delivery at 1,0 bar
17 l/min
Delivery at 1,5 bar
11 l/min
Delivery at 2,0 bar
6 l/min
Pump connection
G3/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 15°C
84 l/h
Consumption at water 15°C, flow 0°C
84 l/h
Consumption at water 15°C, flow -10°C
72 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
3,8 l
Volume of expansion
1,7 l
Значения приведены дословно по тексту страницы; строки 'Cooling power at ambient temperature +20°C' с подпунктами at 15°C / at 0°C / at -10°C объединены в одну характеристику с указанием подпункта.
Engineering Facts
13 facts · every claim linked to a page
Selection
The chiller is water-cooled, so the facility must provide a cooling water circuit with at least 1 bar differential pressure and no more than 6 bar.
Cooling water supply pressure must be verified before selection; too low a differential pressure will not achieve rated cooling.
Source
“min. cooling water differential pressure 1 bar”
Installation
Cooling water inlet temperature must be kept between 5 °C and 20 °C, and pipes must be insulated if condensation can occur.
Defines the permissible cooling water supply temperature range for correct installation and to avoid condensation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Rated cooling power is only valid at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure, so actual capacity will differ otherwise.
Prevents overestimating cooling capacity when site conditions differ from the rating basis.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Operation
Higher cooling water temperature reduces cooling power and may require a higher cooling water flow rate.
Affects expected performance and cooling water supply sizing in warm facilities.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate”
Safety
The plastic pump is not resistant to acetone or polyglycols, so these fluids must not be used without a more resistant pump option.
Critical for fluid selection and material compatibility; wrong fluid can damage the pump.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Guides heat transfer fluid preparation to prevent freezing at low temperatures.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
The unit requires a 208–230 V single/two-phase 60 Hz supply with a 16 A fuse and maximum current of 12.5 A.
Electrical supply and protection must be sized accordingly during installation.
Source
“Power supply requirement 208-230V 1~/2~ 60Hz”
Cumplimiento
Voltage and frequency tolerances must be applied simultaneously: ±5% voltage with ±2% frequency, and certain combinations are not allowed.
Ensures the site power quality is within the allowed envelope before commissioning.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
The unit has IP20 protection, so it must be installed in a dry environment protected from water ingress.
Determines suitable installation location and enclosure requirements.
Source
“Degree of Protection IP20”
Installation
Ambient temperature must remain between 5 °C and 40 °C for proper operation.
Site ambient conditions must be checked to avoid operating outside the allowed range.
Source
“min. ambient temperature 5 °C”
Selection
The refrigerant R-449A has a GWP of 1397 and a CO2 equivalent of 0.6 t, which matters for environmental compliance and reporting.
Relevant for refrigerant regulations and CO2-equivalent reporting at the facility.
Source
“Global Warming Potential (GWP) 1397”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides cooling water quality and material compatibility to prevent corrosion.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation
Three-phase devices below 63 A are supplied with cable but without a plug, so a suitable plug must be provided on site.
Affects electrical installation planning and required accessories.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”