Product introduction and general description of the unit
Specpp. 1-2
Technical data according to DIN 12876 and performance values
p. 2table, 14 rows
p. 1table, 26 rows
Technical data according to DIN 12876
Parameter
Value
Cooling water connection
G3/4 male
min. cooling water differential pressure
1 bar
max. cooling water differential pressure
4 bar
max. cooling water pressure
6 bar
max. System pressure
12 bar
min. filling capacity
8,9 l
Filling capacity expansion tank
19,9 l
Overall dimensions WxDxH **
730x804x1738 mm
Power supply requirement
400V 3~ 50Hz
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/24
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-45...150 °C
Temperature stability at -10°C
0,05 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Interface analog
optional
Alarm message
optic, acoustic, relay
Safety classification
I / NFL
Heating power
12 kW
Cooling power with Water/ Glysantin 40/60 at 20°C
12 kW
Cooling power with Water/ Glysantin 40/60 at 0°C
10 kW
at -20°C
7 kW
at -30°C
4,5 kW
at -40°C
2,5 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-744 (A1, H280)
Global Warming Potential (GWP)
1
Circulation pump:
MK pump
max. delivery
150 l/min
max. delivery pressure
5,3 bar
Pump connection
M38x1,5 male
max. permissible kin. viscosity
200 mm²/s
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit is designed for water-ethylene glycol mixtures with corrosion protection down to -45°C, so it suits automotive test applications but not plain water at low temperatures.
Determines the acceptable heat transfer fluid and the achievable low-temperature limit when selecting the unit.
Source
“The temperature control systems are constructed for operation with Water-ethylene glycol mixture incl. corrosion protection (for example Glysantin®) down to -45°C.”
Installation
Cooling water must be supplied at 1–4 bar differential pressure and no more than 6 bar static pressure, so the site water supply must be checked before installation.
Incorrect cooling water pressure can prevent rated cooling capacity or damage the water-cooled refrigeration circuit.
Source
“min. cooling water differential pressure 1 bar”
Installation
The unit requires a 400 V three-phase 50 Hz supply, so the electrical infrastructure must be verified before siting.
Three-phase power is mandatory; a single-phase supply cannot run this 12 kW heating/cooling unit.
Source
“Power supply requirement 400V 3~ 50Hz”
Installation
Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide and fit the appropriate plug.
Affects electrical installation planning and the scope of supply at commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so the cooling water must be compatible with these materials.
Incompatible cooling water chemistry can cause corrosion and clogging of the water-cooled condenser.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation
If cooling water temperature drops below the condensation point, the cooling water pipes inside the unit must be insulated.
Prevents condensation and corrosion inside the unit when using cold cooling water.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance
Cooling capacity is rated at 20°C room temperature, 15°C cooling water inlet and 1 bar differential pressure, so actual capacity will differ if site conditions deviate.
Needed to correctly interpret the published cooling capacity curves during selection.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -outlet.”
Performance
Cooling capacity falls as cooling water temperature rises, so higher site water temperatures reduce available cooling power.
Affects sizing margin and cooling water supply design.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Operation
The unit is intended for externally closed applications, so it should not be used with open baths.
Defines the permissible hydraulic circuit configuration for safe operation.
Source
“For externally closed applications.”
Operation
The maximum permissible kinematic viscosity of the circulating fluid is 200 mm²/s, which limits fluid choice at low temperatures.
Prevents pump overload and ensures adequate circulation when selecting heat transfer fluids.
Source
“max. permissible kin. viscosity 200 mm²/s”
Safety
Leakage current exceeds 3.5 mA, so protective earth and RCD selection must account for this.
Affects electrical safety design and choice of residual current protective device.
Source
“Attention: leakage current > 3,5mA”
Safety
The refrigerant is R-744 (CO2) with GHS hazard H280, so ventilation and pressure safety measures for the refrigerant must be considered.
Relevant for site risk assessment and handling of the natural refrigerant.
Source
“R-744 (A1, H280)”
Cumplimiento
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Ensures the supply quality at the installation site meets the unit's operating envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Cumplimiento
The equipment is EMC Class A Group 1, so it may cause radio interference in residential environments.
Affects installation location and EMC planning in sensitive environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”