Product introduction and general description of the Unistat 525w
Specpp. 1-2
Technical performance data and operating specifications
Specp. 2
Detailed parameters according to DIN 12876 standard
p. 2table, 20 rows
p. 1table, 32 rows
Technical data according to DIN 12876
Parameter
Value
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 20°C
792 l/h
Consumption at water 15°C, flow 0°C
430 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
5,1 l
Filling capacity expansion tank
18 l
Overall dimensions WxDxH **
540x604x1332 mm
Net weight
215 kg
sound pressure level +/- 4 dB(A)
61 dB(A)
Power supply requirement
400V 3~ 50Hz
max. current
13,5 A
Fuse
3x16A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S393240
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-55...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
6 kW
Cooling power with Thermooil at 250°C
10 kW
Cooling power with Thermooil at 200°C
10 kW
Cooling power with Thermooil at 100°C
10 kW
Cooling power with Thermooil at 20°C
10 kW
Cooling power with Ethanol at 0°C
7 kW
Cooling power with Ethanol at -20°C
4,2 kW
Cooling power with Ethanol at -40°C
1,5 kW
Cooling power with Ethanol at -50°C
0,65 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
1,44 kg
CO2 equivalent
3,1 t
Circulation pump: MK pump max. delivery
79 l/min
Circulation pump: MK pump max. delivery pressure
1,5 bar
Delivery at 0,25 bar
72 l/min
Delivery at 0,5 bar
63 l/min
Delivery at 0,75 bar
54 l/min
Delivery at 1,0 bar
43 l/min
Delivery at 1,25 bar
29 l/min
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply, cable and breaker rating needed before installation.
Source
“Power supply requirement 400V 3~ 50Hz max. current 13,5 A Fuse 3x16A”
Installation
Cooling water must be supplied at 5–20 °C with at least 1 bar differential pressure and no more than 6 bar, and pipes must be insulated if water can drop below the dew point.
Defines the cooling water utility requirements and condensation risk that affect piping design.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance
Cooling capacity is rated at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure, so higher water temperatures reduce available cooling.
Allows correct interpretation of the published cooling power figures for the actual site conditions.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -outlet.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Prevents corrosion or damage by selecting suitable water quality and materials.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Cumplimiento
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and the combination of -5% voltage with +2% frequency is not allowed.
Guides assessment of supply quality and whether the site power is acceptable.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.
Affects electrical installation planning and the need to procure a plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Pump connections are M30x1.5 male with bore shape Y (60°) per DIN 3863, and hoses must use ball socket per DIN 3863 and sleeve nut per DIN 3870.
Ensures correct mating fittings and hoses for leak-free connection.
Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Operation
The unit is rated for ambient temperatures from 5 °C to 40 °C, so the installation room must stay within this range.
Defines the allowable operating environment and ventilation/room requirements.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety
The refrigerant R-452A is classified A1 with GHS hazard H280, and the charge is 1.44 kg corresponding to 3.1 t CO2 equivalent.
Informs refrigerant safety classification and environmental reporting obligations.
Source
“R-452A (A1, H280)”
Selection
The unit provides 6 kW heating and 10 kW cooling with thermooil, with cooling dropping to 0.65 kW at -50 °C with ethanol.
Allows matching the unit to the required process load across the temperature range.
Source
“Cooling power with Ethanol at -50°C 0,65 kW”
Selection
The magnetic-coupled stainless steel pump delivers up to 79 l/min and 1.5 bar, with 43 l/min at 1.0 bar.
Supports checking whether the pump curve suits the external circuit pressure drop.