Product introduction and general description of the Unistat 815w unit
Specpp. 1-2
Technical data and performance specifications according to DIN 12876
p. 2table, 24 rows
p. 1table, 30 rows
Technical data according to DIN 12876
Parameter
Value
max. delivery
40 l/min
max. delivery pressure
0,9 bar
Delivery at 0,3 bar
27 l/min
Delivery at 0,5 bar
20 l/min
Delivery at 0,8 bar
5 l/min
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 0°C
130 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
3,2 l
Filling capacity expansion tank
3,4 l
Overall dimensions WxDxH **
460x604x1465 mm
Net weight
222 kg
sound pressure level +/- 4 dB(A)
60 dB(A)
Power supply requirement
400V 3~ 50Hz
max. current
13 A
Fuse
3x16A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
S466955
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-85...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
2 kW
Cooling power with Thermooil at 250°C
1,5 kW
Cooling power with Thermooil at 200°C
1,5 kW
Cooling power with Thermooil at 100°C
1,5 kW
Cooling power with Ethanol at 0°C
1,5 kW
Cooling power with Ethanol at -20°C
1,5 kW
Cooling power with Ethanol at -40°C
1,4 kW
Cooling power with Ethanol at -60°C
1,2 kW
Cooling power with Ethanol at -80°C
0,2 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
0,95 kg
CO2 equivalent
2,0 t
Refrigerant 2nd stage (ASHRAE, GHS)
R-23 (A1, H280)
Global Warming Potential (GWP)
14800
Refrigerant quantity 2nd stage
0,39 kg
CO2 equivalent 2nd stage
5,8 t
UN-number
UN 2857
Order-No.: 1053.0006.01; Unistat 815w
Engineering Facts
13 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply, cable and protection device needed before installation.
Source
“Power supply requirement 400V 3~ 50Hz”
Installation
The cooling water circuit must be supplied at 5–20 °C with at least 1 bar differential pressure and no more than 6 bar, and pipes must be insulated if water can drop below the dew point.
Defines the cooling water utility requirements and condensation risk at the installation site.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling capacity is rated at 1.5 kW down to -20 °C but falls to 0.2 kW at -80 °C, so low-temperature duty must be checked against the load.
Prevents oversizing expectations for deep-cold applications where available cooling power is small.
Source
“at -80°C 0,2 kW”
Performance
The circulation pump delivers 40 l/min maximum but only 5 l/min at 0.8 bar, so external circuit pressure drop must be checked against the pump curve.
Ensures the external loop resistance does not starve flow and degrade temperature control.
Source
“Delivery at 0,8 bar 5 l/min”
Safety
The unit is safety class III / FL with adjustable overtemperature protection per DIN 12876, which matters for flammable or sensitive fluids.
Confirms suitability and required protective settings for the intended heat transfer fluid.
Source
“Safety classification III / FL”
Safety
The refrigerant R-452A has GWP 2141 and 0.95 kg charge (2.0 t CO2e), and the second stage R-23 has GWP 14800 and 0.39 kg (5.8 t CO2e), so refrigerant regulations apply.
Affects environmental compliance, leak-check obligations and CO2-equivalent reporting.
Source
“Global Warming Potential (GWP) 2141”
Operation
The controller offers selectable internal/process control mode and a 10-program, 100-step programmer with ramp functions, enabling automated temperature profiles.
Determines whether the unit can run the required process ramps and external-sensor control without extra equipment.
Source
“selectable temperature control mode (Internal/Process), programmer with 10 programs (max. 100 steps), ramp function (linear and non-linear)”
Operation
The pump is a variable-speed soft-start unit with integrated pressure control and optional external pressure sensor, allowing adaptation to different external loops.
Supports matching pump performance to external circuit resistance and avoiding pressure shocks.
Source
“Powerful variable speed pump (soft start) with integrated pressure control with optional external pressure sensor.”
Installation
Pump connections are M30x1.5 male and cooling water connections are G1/2 male, so mating hoses and fittings must be ordered to match.
Prevents mismatch of process and cooling water plumbing during installation.
Source
“Pump connection M30x1,5 male”
Installation
The unit weighs 222 kg and measures 460x604x1465 mm, requiring adequate floor space and handling equipment.
Needed for site planning, access routes and floor loading.
Source
“Overall dimensions WxDxH ** 460x604x1465 mm”
Operation
Ambient temperature must stay between 5 °C and 40 °C and protection degree is IP20, so the location must be dry and within that range.
Out-of-range ambient or wet locations can cause malfunction or unsafe operation.
Source
“max. ambient temperature 40 °C”
Cumplimiento
Voltage and frequency tolerances are ±5% and ±2% but must not be combined in opposite directions, so supply quality must be verified.
Avoids invalid supply combinations that could damage the compressor or void compliance.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices below 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site.
Affects electrical installation scope and commissioning schedule.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”