Product introduction and general description of the Unistat 905w unit
Specpp. 1-2
Technical performance data and operating specifications per DIN 12876
p. 2table, 21 rows
p. 1table, 34 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 0,4 bar
36 l/min
Delivery at 0,6 bar
28 l/min
Delivery at 0,8 bar
17 l/min
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 20°C
580 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
3,2 l
Filling capacity expansion tank
5,3 l
Overall dimensions WxDxH **
540x654x1738 mm
Net weight
388 kg
Power supply requirement
208V 3~ 60Hz
max. current
34 A
Fuse
3x40A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S471454
** Please respect space requirements. See operating conditions at www.huber-online.com
Unistat 905w — Technical data according to DIN 12876
Parameter
Value
Order-No.
1054.0008.01
Operating temperature range
-90...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
6 kW
Cooling power with Thermooil at 250°C
4,5 kW
Cooling power with Thermooil at 200°C
4,5 kW
Cooling power with Thermooil at 100°C
4,5 kW
Cooling power with Ethanol at 0°C
4,5 kW
Cooling power with Ethanol at -20°C
4,5 kW
Cooling power with Ethanol at -40°C
4 kW
Cooling power with Ethanol at -60°C
2,5 kW
Cooling power with Ethanol at -80°C
0,7 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
1,4 kg
CO2 equivalent
3,0 t
Refrigerant 2nd stage (ASHRAE, GHS)
R-23 (A1, H280)
Global Warming Potential (GWP)
14800
Refrigerant quantity 2nd stage
0,75 kg
CO2 equivalent 2nd stage
11,1 t
Circulation pump:
MK pump
max. delivery
48 l/min
max. delivery pressure
0,9 bar
Delivery at 0,2 bar
43 l/min
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a water-cooled refrigeration machine, so a cooling water supply with at least 1 bar differential pressure and 580 l/h flow at 15°C must be available at the installation site.
Cooling water capacity and pressure determine whether the site can support the unit without additional pumps or chillers.
Source
“Consumption at water 15°C, flow 20°C 580 l/h”
Installation
The cooling water inlet temperature must be kept between 5°C and 20°C; below the condensation point the cooling water pipes must be insulated.
Operating outside this range risks condensation, reduced cooling power, or equipment damage.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops significantly at low temperatures, from 4.5 kW at -20°C to only 0.7 kW at -80°C, which must be considered when sizing for deep-cold applications.
The available cooling capacity at the target process temperature determines achievable pull-down and heat removal.
Source
“at -80°C 0,7 kW”
Installation
The electrical supply must be 208V 3~ 60Hz with a maximum current of 34 A and 3x40A fusing, and external branch circuit protection according to UL 489 is required.
The site electrical infrastructure and protection devices must be rated for the unit's current and comply with UL requirements.
Source
“External branch circuit protection according to UL 489 required.”
Safety
The unit has a leakage current greater than 3.5 mA, which may require additional protective measures such as a higher-rated RCD.
Standard 30 mA RCDs may nuisance-trip; the electrical design must account for this leakage current.
Source
“Attention: leakage current > 3,5mA”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but they must not be applied in opposite directions simultaneously.
Supply quality must be verified; a combination such as -5% voltage with +2% frequency is not permitted.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices with current consumption less than 63A are delivered with cable but without a plug, so a suitable plug must be provided on site.
The installer must plan for a compatible plug and receptacle, as none is included.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Selection
The pump delivers 48 l/min at zero head but only 17 l/min at 0.8 bar, so external circuit pressure drop must be checked against the pump curve.
Insufficient flow at the required pressure can cause poor temperature control in external closed or open applications.
Source
“Delivery at 0,8 bar 17 l/min”
Selection
The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, which limits the choice of fluid at low temperatures.
Using a fluid that is too viscous at the operating temperature can overload the pump and reduce performance.
Source
“max. permissible kin. viscosity 50 mm²/s”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so the cooling water must be compatible with these materials.
Aggressive or scaling cooling water can corrode or foul these materials, reducing cooling performance and causing leaks.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Safety
The refrigerant R-452A has a GWP of 2141 and the second-stage R-23 has a GWP of 14800, giving a total CO2 equivalent of about 14.1 t, which matters for leak reporting and environmental compliance.
High-GWP refrigerants are subject to F-gas regulations, leak checks, and end-of-life recovery obligations.
Source
“Global Warming Potential (GWP) 14800”
Operation
The unit is rated IP20 and must be operated between 5°C and 40°C ambient temperature, so it cannot be installed in wet or dusty environments or outside this ambient range.
Exceeding the ambient limit or exposing the unit to moisture/dust can cause overheating or electrical hazards.
Source
“max. ambient temperature 40 °C”
Selection
The minimum filling capacity is 3.2 l and the expansion tank holds 5.3 l, which sets the minimum external circuit volume and expansion allowance.
External circuits smaller than the minimum filling capacity may not allow stable temperature control or proper pump operation.
Source
“min. filling capacity 3,2 l”
Cumplimiento
The equipment is classified as EN55011 Class A, Group 1, so it is intended for industrial environments and may cause radio interference in residential areas.
Class A equipment may not be suitable for residential use without additional EMC measures.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”