Unistat 925w Technical Data

Contenido

  1. Overview p. 1
    Introduces the Unistat 925w unit and its main features
  2. Spec pp. 1-2
    Performance, temperature range, and technical data per DIN 12876
  3. p. 2 table, 23 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
max. delivery pressure2,5 bar
Delivery at 0,5 bar168 l/min
Delivery at 1,0 bar126 l/min
Delivery at 1,5 bar100 l/min
Delivery at 2,0 bar66 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C1300 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity12 l
Filling capacity expansion tank29 l
Overall dimensions WxDxH **950x1205x1650 mm
Net weight992 kg
Power supply requirement460V 3~ 60Hz
max. current45,2 A
Fuse3x63A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S603469

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-90...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power12 kW
Cooling power with Thermooil at 200°C16 kW
Cooling power with Thermooil at 100°C16 kW
Cooling power with Thermooil at 20°C16 kW
Cooling power with Ethanol at 0°C16 kW
Cooling power with Ethanol at -20°C16 kW
Cooling power with Ethanol at -40°C15 kW
Cooling power with Ethanol at -60°C13,5 kW
Cooling power with Ethanol at -80°C3,5 kW
Cooling power with Ethanol at -90°C1,8 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity5 kg
CO2 equivalent10,7 t
Refrigerant 2nd stage (ASHRAE, GHS)R-23 (A1, H280)
Global Warming Potential (GWP)14800
Refrigerant quantity 2nd stage2,2 kg
CO2 equivalent 2nd stage32,6 t
UN-numberUN 2857
Circulation pump:MK pump
max. delivery168 l/min

Order-No.: 1081.0022.01; Unistat 925w

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 460 V 3-phase 60 Hz supply with a 3x63 A fuse, so the electrical infrastructure must be sized accordingly.

Determines site power availability, cable sizing and breaker selection before purchase.

Source
“Power supply requirement 460V 3~ 60Hz”
Installation

The cooling water supply must be between 5 °C and 20 °C with at least 1 bar differential pressure and no more than 6 bar.

Cooling water conditions directly affect achievable cooling power and must be verified on site.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

Cooling water pipes must be insulated if the water temperature can fall below the condensation point.

Prevents condensation damage and corrosion in the cooling water circuit.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality may be used.

Guides water treatment and material compatibility to avoid corrosion and fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Defines acceptable supply quality and warns against out-of-tolerance combinations that could damage the unit.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

Leakage current exceeds 3.5 mA, so external branch circuit protection per UL 489 is mandatory.

Requires proper protective device selection and grounding to meet electrical safety.

Source
“Attention: leakage current > 3,5mA”
Installation

Three-phase devices drawing less than 63 A are supplied with cable but without a plug.

The customer must plan for on-site plug/connection hardware and qualified electrician work.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Selection

Cooling power drops significantly at low temperatures, from 16 kW at -20 °C to 1.8 kW at -90 °C.

Critical for sizing the unit for low-temperature process loads.

Source
“at -90°C 1,8 kW”
Selection

The refrigerant R-452A has a GWP of 2141 and 5 kg charge, giving 10.7 t CO2 equivalent; the second stage R-23 adds 32.6 t.

Affects environmental compliance, refrigerant handling regulations and total CO2 footprint.

Source
“CO2 equivalent 10,7 t”
Operation

The pump delivers 168 l/min at 0.5 bar but only 66 l/min at 2.0 bar, so system pressure drop must be matched to the pump curve.

Ensures adequate circulation flow through the external circuit at the actual pressure drop.

Source
“Delivery at 2,0 bar 66 l/min”
Operation

Maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s.

Limits fluid selection at low temperatures to avoid pump overload and poor heat transfer.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

Minimum filling capacity is 12 l and the expansion tank holds 29 l, which sets the minimum fluid charge and expansion allowance.

Needed to plan fluid volume, expansion volume and external loop volume.

Source
“min. filling capacity 12 l”
Operation

The unit is rated IP20 and must operate between 5 °C and 40 °C ambient.

Defines the allowable room environment and protection level for installation planning.

Source
“max. ambient temperature 40 °C”
Cumplimiento

The unit is classified as EMC Class A, Group 1, which is intended for industrial environments.

May restrict use in residential areas and requires consideration of nearby sensitive equipment.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”