Unistat GL 550w Technical Data

Contenido

  1. Overview p. 1
    Product introduction and general description of the Unistat GL 550w unit
  2. Spec pp. 1-2
    Technical performance data, dimensions, and operating parameters
  3. Spec p. 2
    Detailed specifications according to DIN 12876 standard
  4. p. 2 table, 20 rows
  5. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C3100 l/h
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **918x963x1771 mm
Net weight800 kg
Power supply requirement400V 3~ 50Hz
max. current61,5 A
Fuse3x63A
Pressure equipment categoryII
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S581643

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

CharacteristicValue
Operating temperature range-50...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power24 kW
Cooling power with Thermooil at 200°C41 kW
Cooling power with Thermooil at 100°C41 kW
Cooling power with Thermooil at 20°C41 kW
Cooling power with Ethanol at 0°C37 kW
Cooling power with Ethanol at -20°C22 kW
Cooling power with Ethanol at -40°C10 kW
Cooling power with Ethanol at -50°C6 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-744 (A1, H280)
Global Warming Potential (GWP)1
Refrigerant quantity8,7 kg
CO2 equivalent0,0 t
UN-numberUN 2857
Circulation pump: MK pump
max. delivery196 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar175 l/min
Delivery at 1,0 bar152 l/min
Delivery at 1,5 bar123 l/min
Delivery at 2,0 bar86 l/min

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 3x63 A fuse and draws up to 61.5 A, so the site electrical infrastructure must be sized accordingly.

Determines the electrical supply, cable and protection device needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz max. current 61,5 A Fuse 3x63A”
Installation

The cooling water circuit must be supplied at 5–20 °C with 1–4 bar differential pressure and no more than 6 bar pressure, and pipes must be insulated if water can drop below the dew point.

Defines the cooling water utility requirements and condensation risk at the installation site.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

The cooling water connection is G3/4 male and the pump connection is M38x1.5 male, so matching hoses and adapters must be procured.

Needed to order correct hoses, fittings and adapters for commissioning.

Source
“Pump connection M38x1,5 male”
Installation

The unit is 918x963x1771 mm and weighs 800 kg, so transport routes, floor loading and clearance must be planned.

Affects rigging, doorway clearance and structural floor requirements.

Source
“Overall dimensions WxDxH ** 918x963x1771 mm Net weight 800 kg”
Installation

The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be installed in a wet or condensing environment.

Constrains the installation location regarding moisture and ambient temperature.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety

The equipment produces leakage current above 3.5 mA, which requires appropriate protective earthing and RCD selection.

Affects electrical protection design and compliance with local wiring regulations.

Source
“Attention: leakage current > 3,5mA”
Installation

Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so a suitable plug must be supplied on site.

Prevents commissioning delays because the customer must provide the plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Cumplimiento

Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.

Must be checked against the site power quality to avoid invalid operating conditions.

Source
“-5% voltage and +2% frequency -> not allowed!”
Operation

The unit uses R-744 (CO2) refrigerant with GWP 1 and a charge of 8.7 kg, which simplifies refrigerant regulations compared with high-GWP fluids.

Relevant for environmental compliance and refrigerant handling requirements.

Source
“Refrigerant (ASHRAE, GHS) R-744 (A1, H280) Global Warming Potential (GWP) 1 Refrigerant quantity 8,7 kg”
Performance

Cooling capacity depends strongly on the heat transfer fluid and temperature, e.g. 41 kW with thermooil at 20–200 °C but only 6 kW with ethanol at -50 °C.

Needed to verify the unit meets the process heat load at the required operating temperature.

Source
“Cooling power with Ethanol at -50°C 6 kW”
Performance

The circulation pump delivers 196 l/min maximum and 86 l/min at 2.0 bar, so system pressure drop must be checked against the pump curve.

Ensures adequate flow through the external circuit at the actual pressure drop.

Source
“max. delivery 196 l/min max. delivery pressure 2,5 bar Delivery at 2,0 bar 86 l/min”
Operation

Minimum filling capacity is 8.9 l and the expansion tank holds 19.9 l, which sets the heat transfer fluid charge and expansion allowance.

Determines fluid volume to purchase and expansion volume available.

Source
“min. filling capacity 8,9 l Filling capacity expansion tank 19,9 l”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and treatment to avoid corrosion and fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento

The unit is EMC classified as Class A, Group 1 to EN55011, which may require additional filtering in residential or sensitive environments.

Affects EMC planning and suitability for the intended electromagnetic environment.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”