Unistat P515w Technical Data

Contenido

  1. Overview p. 1
    Product introduction and general description of the Unistat P515w unit
  2. Spec pp. 1-2
    Technical data and performance specifications according to DIN 12876
  3. p. 2 table, 21 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 1,5 bar74 l/min
Delivery at 2,0 bar55 l/min
Delivery at 2,5 bar33 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 0°C504 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity4,1 l
Filling capacity expansion tank3,7 l
Overall dimensions WxDxH **460x554x1453 mm
Net weight181 kg
Power supply requirement400V 3~ 50Hz
max. current14,5 A
Fuse3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S429820

** Please respect space requirements. See operating conditions at www.huber-online.com

Unistat P515w — Technical data according to DIN 12876

ParameterValue
Order-No.1071.0004.01
Operating temperature range-50...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Cooling power with Thermooil at 250°C7 kW
Cooling power with Thermooil at 200°C7 kW
Cooling power with Thermooil at 100°C7 kW
Cooling power with Thermooil at 20°C7 kW
Cooling power with Ethanol at 0°C5,3 kW
Cooling power with Ethanol at -20°C2,8 kW
Cooling power with Ethanol at -40°C0,9 kW
Cooling power with Ethanol at -50°C0,25 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity1,45 kg
CO2 equivalent3,1 t
Circulation pump:MK pump
max. delivery119 l/min
max. delivery pressure3 bar
Delivery at 0,5 bar106 l/min
Delivery at 1,0 bar91 l/min

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and protection device needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

The cooling water circuit must be supplied at 1 bar minimum differential pressure and must not exceed 6 bar.

Cooling water supply pressure must be verified to avoid insufficient cooling or overpressure damage.

Source
“min. cooling water differential pressure 1 bar”
Installation

Cooling water intake temperature must stay between 5 °C and 20 °C, and pipes must be insulated if condensation can occur.

Defines the permissible cooling water conditions and condensation protection required at the site.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity is rated at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure, so site conditions affect achievable performance.

Performance figures must be corrected if site cooling water or ambient conditions differ from the rating basis.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Performance

Higher cooling water temperature reduces cooling power and may require a higher cooling water flow rate.

Prevents overestimating cooling capacity when the available cooling water is warmer than the rating basis.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides selection of cooling water quality and treatment to avoid corrosion or deposits.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Defines acceptable supply quality limits for reliable operation and warranty compliance.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

The equipment has a leakage current greater than 3.5 mA, which affects protective earthing and RCD selection.

Requires suitable earthing and residual current protection design for safe installation.

Source
“Attention: leakage current > 3,5mA”
Installation

Three-phase devices below 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.

Affects electrical installation planning and the need to procure a matching plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Pump connections use M30x1.5 male threads with bore shape Y (60°) per DIN 3863, and hoses must use ball socket and sleeve nut per DIN 3870.

Ensures correct selection of hoses and fittings for leak-free connection.

Source
“Pump connection M30x1,5 male”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s.

Limits the choice of heat transfer fluid and operating temperature range to protect the pump.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The unit must be operated within an ambient temperature range of 5 °C to 40 °C.

Defines the required room conditions and space planning for reliable operation.

Source
“min. ambient temperature 5 °C”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.

Confirms built-in safety protection relevant for risk assessment and compliance.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Selection

The P models use circulation pumps with high discharge pressure for applications with high pressure drops.

Helps confirm the pump is suitable for high-pressure-drop external circuits.

Source
“Unistat „P“ Models: Circulating pumps with a high discharge pressure for applications with high pressure drops”