Product introduction and general description of Unistat P625w
Specpp. 1-2
Technical data according to DIN 12876 including performance parameters
p. 2table, 17 rows
p. 1table, 31 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 5,0 bar
42 l/min
Pump connection
M38x1,5 male
max. permissible kin. viscosity
100 mm²/s
Cooling water connection
G3/4 male
min. cooling water differential pressure
1 bar
max. cooling water differential pressure
4 bar
max. cooling water pressure
6 bar
max. System pressure
6 bar
min. filling capacity
8,9 l
Filling capacity expansion tank
19,9 l
Overall dimensions WxDxH **
918x963x1771 mm
Power supply requirement
400V 3~ 50Hz
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/24
** Please respect space requirements. See operating conditions at www.huber-online.com
Unistat P625w — Technical data according to DIN 12876
Characteristic
Value
Order-No.
5017.0101.01
Operating temperature range
-60...200 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Interface analog
optional
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
12 kW
Cooling power with Thermooil at 200°C
16 kW
Cooling power with Thermooil at 100°C
16 kW
Cooling power with Thermooil at 20°C
16 kW
Cooling power with Ethanol at 0°C
16 kW
Cooling power with Ethanol at -20°C
15 kW
Cooling power with Ethanol at -40°C
6,4 kW
Cooling power with Ethanol at -60°C
1,7 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Circulation pump
MK pump
max. delivery
201 l/min
max. delivery pressure
5,3 bar
Delivery at 1,0 bar
179 l/min
Delivery at 2,0 bar
154 l/min
Delivery at 3,0 bar
124 l/min
Delivery at 4,0 bar
87 l/min
Engineering Facts
13 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase supply, so the site must provide a 3~ 50 Hz feed sized for the 12 kW heater and 16 kW cooling load.
Determines electrical infrastructure, cable sizing and whether the facility can support the unit at all.
Source
“Power supply requirement
400V 3~ 50Hz”
Installation
Cooling water must be supplied at 1–4 bar differential and no more than 6 bar, so the building cooling loop needs a pressure regulator or check.
Prevents damage to the water-cooled condenser and ensures rated cooling performance.
Source
“min. cooling water differential pressure
1 bar
max. cooling water differential pressure
4 bar
max. cooling water pressure
6 bar”
Installation
The unit is designed for cooling water intake between 5 °C and 20 °C, and pipes must be insulated if water can drop below the condensation point.
Defines the acceptable cooling water supply range and the need for condensation insulation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling capacity is rated at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential, so actual capacity will differ if site conditions deviate.
Allows correct comparison of quoted cooling power against real site conditions.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Installation
The pump connection is M38x1.5 male with a Y-shaped bore per DIN 3863, so mating hoses must use ball socket DIN 3863 and sleeve nut DIN 3870.
Ensures correct fittings are ordered and leak-free connections are made.
Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Operation
The pump delivers 201 l/min at zero head but only 42 l/min at 5 bar, so high-pressure-drop loops must be checked against the pump curve.
Confirms whether the circulation pump can overcome the pressure drop of the external circuit.
The refrigerant R-452A has GWP 2141 and is classified A1/H280, so leak detection and refrigerant handling procedures must account for a high-GWP, flammable-classified substance.
Affects environmental compliance, leak monitoring and service procedures.
The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be installed in a washdown or hot environment.
Sets the environmental envelope for the installation location.
Source
“Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Installation
The unit is 918x963x1771 mm, so space and access for a machine of this footprint must be planned before delivery.
Determines room layout, door widths and lifting/rigging requirements.
Source
“Overall dimensions WxDxH **
918x963x1771 mm”
Installation
Three-phase devices below 63 A are supplied with cable but without a plug, so the customer must provide a suitable plug and connection.
Avoids a missing-plug surprise at commissioning and clarifies scope of supply.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Cumplimiento
Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.
Prevents operation outside the permitted supply envelope that could damage the compressor or electronics.
Source
“-5% voltage and +2% frequency -> not allowed!”
Operation
The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s, so fluid selection must respect this limit at operating temperature.
Guides selection of a suitable heat transfer fluid and avoids pump overload.
Source
“max. permissible kin. viscosity
100 mm²/s”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Prevents corrosion and fouling that would degrade cooling performance.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”