Induztr
Resumen técnico Visor de documentos Modelos relacionados

Unistat P630w Technical Data

Contenido

  1. Overview p. 1
    Introduces the Unistat P630w unit and its main features
  2. Spec pp. 1-2
    Lists performance, temperature, and physical specifications per DIN 12876
  3. p. 2 table, 17 rows
  4. p. 1 table, 30 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 5,0 bar42 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **918x963x1771 mm
Power supply requirement400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/24

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unistat P630w (Order-No.: 5018.0101.01)

ParameterValue
Operating temperature range-60...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power24 kW
Cooling power with Thermooil at 200°C22 kW
Cooling power with Thermooil at 100°C22 kW
Cooling power with Thermooil at 20°C22 kW
Cooling power with Ethanol at 0°C21 kW
Cooling power with Ethanol at -20°C20 kW
Cooling power with Ethanol at -40°C10,5 kW
Cooling power with Ethanol at -60°C2,5 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Circulation pump: MK pump
max. delivery201 l/min
max. delivery pressure5,3 bar
Delivery at 1,0 bar179 l/min
Delivery at 2,0 bar154 l/min
Delivery at 3,0 bar124 l/min
Delivery at 4,0 bar87 l/min

Значения приведены дословно; строки 'Cooling power with Thermooil' и 'Cooling power with Ethanol' в оригинале разбиты на подстроки по температурам.

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a three-phase 400 V supply and is delivered without a plug, so the customer must provide a suitable connection.

Electrical installation planning must include a 400 V 3~ supply and a plug/connection provided on site.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Cooling water must be supplied at 1–4 bar differential pressure and no more than 6 bar, with intake temperature between 5 °C and 20 °C.

The site cooling water system must meet these pressure and temperature limits or cooling performance and safety are compromised.

Source
“min. cooling water differential pressure 1 bar”
Installation

If cooling water is colder than the condensation point, the cooling water pipes inside the unit must be insulated.

Prevents condensation and corrosion/electrical risk when using cold cooling water.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling capacity is rated at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure, and drops as cooling water temperature rises.

Selection must account for actual site cooling water conditions, otherwise available cooling power will be lower than catalog values.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -outlet.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Water quality must be checked for corrosion and scaling to protect these materials.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Cumplimiento

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Site power quality must be verified; some voltage/frequency combinations exceed the permitted operating envelope.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit is IP20 and must be operated between 5 °C and 40 °C ambient temperature.

Installation location must be dry and within the ambient temperature range; IP20 offers no water protection.

Source
“Degree of Protection IP20”
Installation

Pump connections are M38x1.5 male and require ball socket fittings per DIN 3863 and sleeve nuts per DIN 3870.

Correct mating fittings must be procured to avoid leaks and ensure proper connection.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Selection

The pump delivers 201 l/min max and 5.3 bar max, with 42 l/min still available at 5.0 bar, making it suitable for high pressure drop applications.

Confirms suitability for flow-through chemistry or semiconductor applications with high pressure losses.

Source
“Circulating pumps with a high discharge pressure for applications with high pressure drops”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.

Safety classification and overtemperature protection are relevant for flammable fluids and process safety assessment.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

The refrigerant is R-452A with GWP 2141, which matters for environmental compliance and leak management.

High GWP refrigerant requires attention to leak prevention and regulatory reporting.

Source
“R-452A (A1, H280)”
Maintenance

The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s.

Fluid selection at low temperatures must respect this limit to avoid pump damage and poor circulation.

Source
“max. permissible kin. viscosity 100 mm²/s”
Selection

Minimum filling capacity is 8.9 l and the expansion tank holds 19.9 l.

Determines the required heat transfer fluid volume and expansion allowance for the system.

Source
“min. filling capacity 8,9 l”
Installation

The unit requires space per the overall dimensions 918x963x1771 mm and the manufacturer's space requirements must be respected.

Room layout and service access must accommodate the unit footprint and clearance requirements.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”