Introduces the Unistat P950w unit and its main features
Specp. 2
Lists technical data according to DIN 12876 standard
p. 2table, 19 rows
p. 1table, 31 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 2,5 bar
173 l/min
Delivery at 3,0 bar
153 l/min
Delivery at 4,0 bar
103 l/min
Delivery at 4,5 bar
68 l/min
Pump connection
flange DN40 PN10/40 EN1092-1
max. permissible kin. viscosity
100 mm²/s
Cooling water connection
G1 1/4 male
Consumption at water 15°C, flow 0°C
5000 l/h
min. cooling water differential pressure
3 bar
max. cooling water pressure
6 bar
min. filling capacity
30 l
Filling capacity expansion tank
90 l
Overall dimensions WxDxH **
2630x1300x1980 mm
Power supply requirement
400V 3~ 50Hz
Pressure equipment category
IV
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/23
** Please respect space requirements. See operating conditions at www.huber-online.com
Unistat P950w — Technical data according to DIN 12876
Parameter
Value
Order-No.
1065.0005.01
Operating temperature range
-90...200 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
36 kW
Cooling power with Thermooil at 200°C
36 kW
Cooling power with Thermooil at 100°C
36 kW
Cooling power with Thermooil at 20°C
36 kW
Cooling power with Ethanol at 0°C
36 kW
Cooling power with Ethanol at -20°C
36 kW
Cooling power with Ethanol at -40°C
36 kW
Cooling power with Ethanol at -60°C
25 kW
Cooling power with Ethanol at -80°C
10 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-507 (A1, H280)
Global Warming Potential (GWP)
3985
Refrigerant 2nd stage (ASHRAE, GHS)
R-23 (A1, H280)
Global Warming Potential (GWP)
14800
Refrigerant quantity 2nd stage
7 kg
CO2 equivalent 2nd stage
103,6 t
Circulation pump:
MK pump
max. delivery
260 l/min
max. delivery pressure
4,8 bar
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase 50 Hz supply, so the electrical installation must be sized for a high-power three-phase load.
Determines the site power distribution, cable sizing and protection device selection before installation.
Source
“Power supply requirement
400V 3~ 50Hz”
Installation
The cooling water circuit must be supplied at 3–6 bar with a minimum differential pressure of 3 bar, so the site cooling water system must be verified before commissioning.
Insufficient cooling water pressure will prevent the water-cooled refrigeration machine from operating correctly.
Source
“min. cooling water differential pressure
3 bar
max. cooling water pressure
6 bar”
Installation
The unit is designed for cooling water intake between 5 °C and 20 °C, and cooling water pipes must be insulated if the water temperature can fall below the condensation point.
Cooling water outside this range reduces cooling capacity and can cause condensation and electrical risk.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops significantly at low temperatures, from 36 kW at -40 °C to 25 kW at -60 °C and 10 kW at -80 °C, so the process heat load must be checked against the actual operating temperature.
Sizing only on the 36 kW headline figure will under-estimate the required capacity at deep cooling temperatures.
Source
“at -60°C
25 kW
at -80°C
10 kW”
Installation
The pump connections are DN40 PN10/40 flanges, so the customer pipework must be prepared with matching flanges.
Mismatched flanges will prevent connection and may require adapters or rework of the piping.
Source
“Pump connection
flange DN40 PN10/40
EN1092-1”
Installation
The unit requires a minimum filling capacity of 30 l and has a 90 l expansion tank, so the heat transfer fluid volume and expansion allowance must be planned.
Under-filling or insufficient expansion volume can cause pump cavitation and poor temperature control.
Source
“min. filling capacity
30 l
Filling capacity expansion tank
90 l”
Safety
The leakage current exceeds 3.5 mA, so protective earthing and residual current device selection must account for this.
Standard 30 mA RCDs may nuisance-trip; earth connection and RCD type must be chosen accordingly.
Source
“Attention: leakage current > 3,5mA”
Cumplimiento
Voltage and frequency tolerances are ±5 % and ±2 % respectively, but the combination of -5 % voltage with +2 % frequency is not allowed.
Site power quality must be checked; an out-of-tolerance supply can damage the compressor and electronics.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Three-phase devices drawing less than 63 A are delivered with cable but without a plug, so the customer must provide the final connection.
The electrical contractor must supply and install the appropriate plug or direct connection.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The unit must be installed in an ambient temperature between 5 °C and 40 °C and has IP20 protection, so it must be kept in a dry indoor location.
IP20 offers no water protection, and ambient temperature outside the range affects performance and reliability.
Source
“Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Selection
The refrigerant charges are R-507 (GWP 3985) and R-23 (GWP 14800, 7 kg, 103.6 t CO2 equivalent), which affects F-gas compliance and leak monitoring obligations.
High-GWP refrigerants above the CO2-equivalent threshold trigger periodic leak checks and record-keeping.
Source
“Refrigerant quantity 2nd stage
7 kg
CO2 equivalent 2nd stage
103,6 t”
Operation
The pump delivers 260 l/min at maximum with 4.8 bar maximum delivery pressure, and flow falls as pressure rises, so the external circuit pressure drop must be matched to the pump curve.
A circuit with high pressure drop may not achieve the required flow, affecting heat transfer and control stability.