Unistat T320w HT Temperature Control System Datasheet

Contenido

  1. Overview p. 1
    Product introduction and general description of the Unistat T320w HT unit
  2. Spec pp. 1-2
    Technical data, performance figures and operating parameters per DIN 12876
  3. p. 2 table, 9 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Net weight179 kg
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)20 A
Fuse (3 phase)3x20A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
max. ambient temperature40 °C
min. ambient temperature5 °C
from Serial-No.:S591845

Technical data according to DIN 12876

ParameterValue
Operating temperature range65...300 °C
Minimum temperature with water cooling15 °C
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power12 kW
Cooling power at 300°C10 kW
Cooling power at 200°C10 kW
Cooling power at 100°C10 kW
Circulation pump: MK pump
max. delivery96 l/min
max. delivery pressure3,5 bar
Delivery at 0,5 bar82 l/min
Delivery at 1,0 bar72 l/min
Delivery at 1,5 bar60 l/min
Delivery at 2,0 bar45 l/min
Delivery at 2,5 bar33 l/min
Delivery at 3,0 bar20 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water consumption at water temp. 15°C640 l/h
Cooling water connectionG3/4 male
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity7 l
Filling capacity expansion tank24 l
Overall dimensions WxDxH **540x749x1158 mm

** см. сноску в оригинале (в тексте страницы пояснение к сноске отсутствует)

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a three-phase 400 V supply fused at 3x20 A, so the electrical installation must be sized for a 20 A per-phase load.

Determines the site power supply, cable cross-section and breaker rating needed before installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

The unit is rated IP20, so it must be installed in a dry indoor location protected from water and dust ingress.

Affects placement and enclosure requirements at the installation site.

Source
“Degree of Protection IP20”
Installation

Ambient temperature at the installation site must stay between 5 °C and 40 °C for correct operation.

Room conditions must be verified during site selection and ventilation planning.

Source
“max. ambient temperature 40 °C”
Installation

Cooling water must be supplied at 1 bar minimum differential pressure and no more than 6 bar, with a G3/4 male connection.

Defines the cooling water supply system and fittings required for commissioning.

Source
“min. cooling water differential pressure 1 bar”
Installation

Cooling water intake temperature must be between 5 °C and 20 °C, and pipes must be insulated if it can drop below the dew point.

Prevents condensation damage and ensures rated cooling performance.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

The stated cooling power of 10 kW is only valid at 20 °C room temperature, 15 °C cooling water and 1 bar differential pressure.

Prevents overestimating cooling capacity when site conditions differ from the reference values.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Selection

This model has no active refrigeration, so reaching 65 °C depends on adequate heat loss from the application or on water cooling.

Critical for deciding whether the unit suits the process or an external cooling source is needed.

Source
“These devices do not have an active refrigeration system.”
Operation

The pump delivers 96 l/min at zero head but only 20 l/min at 3.0 bar, so the system pressure drop must be checked against the pump curve.

Ensures the circulation loop resistance does not starve flow at the required rate.

Source
“Delivery at 3,0 bar 20 l/min”
Installation

Pump connections are M30x1.5 male with Y-shaped bore per DIN 3863, requiring matching ball-socket hoses and sleeve nuts per DIN 3870.

Specifies the correct hose and fitting standard to avoid leaks or mismatched connections.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so a suitable plug must be sourced on site.

Affects electrical commissioning planning and procurement.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

Voltage and frequency tolerances of ±10% apply only if they do not run in opposite directions, e.g. -10% voltage with +3% frequency is not allowed.

Defines acceptable grid quality for reliable operation of the electronics and pump.

Source
“-10% voltage and +3% frequency -> not allowed !”
Safety

The unit is safety class III / FL with adjustable overtemperature protection per DIN 12876.

Relevant for risk assessment and compliance with laboratory safety requirements.

Source
“Safety classification III / FL”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and treatment to avoid corrosion and fouling of the heat exchanger.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Operation

The expansion tank is not thermoregulated and must be locked for open bath operation.

Affects setup and safe operation depending on whether a closed or open bath is used.

Source
“Expansion tank (not thermoregulated) for closed systems, lockable for open baths.”