Unistat tango wl — Temperature Control System

Contenido

  1. Overview p. 1
    Introduces the Unistat tango wl unit and its purpose
  2. Spec p. 2
    Technical data per DIN 12876 for the unit
  3. p. 2 table, 19 rows
  4. p. 1 table, 28 rows

Technical data according to DIN 12876

ParameterValue
Consumption at water 15°C, flow 0°C66 l/h
Consumption at water 15°C, flow -20°C30 l/h
Consumption at water 25°C, flow 0°C102 l/h
Verbrauch b. Wasser 25°C, Vorlauf -20°C54 l/h
min. cooling water differential pressure0,5 bar
max. cooling water pressure6 bar
min. filling capacity1,5 l
Filling capacity expansion tank2,8 l
Overall dimensions WxDxH **426x327x631 mm
Net weight61 kg
sound pressure level +/- 4 dB(A)56 dB(A)
Power supply requirement220-240V 1~/2~ 50/60Hz
max. current15,5 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S585454

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-45...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power3 kW
Cooling power with Thermooil at 250°C0,7 kW
Cooling power with Thermooil at 200°C0,7 kW
Cooling power with Thermooil at 100°C0,7 kW
Cooling power with Ethanol at 0°C0,7 kW
Cooling power with Ethanol at -20°C0,4 kW
Cooling power with Ethanol at -40°C0,06 kW
Refrigeration machineair- water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-1270 (A3, H220)
Global Warming Potential (GWP)0
Refrigerant quantity0,17 kg
CO2 equivalent0,0 t
Circulation pump: max. delivery55 l/min
Circulation pump: max. delivery pressure0,9 bar
Pump connectionM24x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is a hydraulically sealed refrigerated heating circulator suitable for both externally closed and externally open applications, which determines whether an isolation sleeve or expansion-tank blocking is needed.

Application type dictates accessories and whether the expansion tank must be blocked off for open baths.

Source
“As well as for externally closed and also externally open applications.”
Installation

For external open baths the expansion tank is blocked off, making the thermostat atmospherically sealed so it can be placed below or above the application level.

This allows flexible physical placement relative to the open bath without atmospheric head constraints.

Source
“For external open baths the expansion tank will be blocked off. This means that the thermostat is atmospherically sealed and can be located below or above the level of the application.”
Installation

The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Incompatible water chemistry can corrode or foul the cooling circuit and reduce performance.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Installation

The unit is designed for cooling water temperatures between 5 °C and 20 °C, and if the cooling water is below the dew point the cooling water pipes inside the unit must be insulated.

Condensation on internal cooling water pipes can cause electrical or corrosion problems if not insulated.

Source
“The temperature control unit is designed for cooling water temperatures between 5 °C and 20 °C. If the cooling water temperature is below the dew point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Output data are valid only at room temperature 20 °C, cooling water inlet 15 °C and 0.5 bar differential pressure, so higher cooling water or ambient temperatures will reduce cooling capacity.

Specified cooling power cannot be expected outside these reference conditions.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 0,5 bar differential pressure between cooling water inlet and -outlet.”
Installation

The minimum cooling water differential pressure is 0.5 bar and the maximum cooling water pressure is 6 bar, which sets the allowable supply pressure window.

Too low differential pressure gives insufficient cooling; too high pressure may damage the circuit.

Source
“min. cooling water differential pressure 0,5 bar”
Cumplimiento

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and a -5% voltage with +2% frequency is explicitly not allowed.

Supply quality outside this combined envelope may cause malfunction or damage and affects site power assessment.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Single/two-phase devices are supplied with a power cable and country-specific plug that must be specified at ordering, while three-phase devices below 63 A come with cable but no plug.

The plug type must be stated at order and the site must provide the matching receptacle or a qualified connection.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety

The refrigerant is R-1270 (A3, H220) with GWP 0 and a charge of 0.17 kg, which is a flammable natural refrigerant requiring appropriate handling and ventilation awareness.

A3 flammability classification affects installation location, ventilation and service procedures.

Source
“R-1270 (A3, H220)”
Installation

The pump connection is M24x1.5 male and the cooling water connection is G1/2 male, so mating hoses and fittings must match these threads.

Correct thread matching avoids adapters and leaks during commissioning.

Source
“Pump connection M24x1,5 male”
Installation

Pump connections use bore shape Y (60°) per DIN 3863 with ball socket per DIN 3863 and sleeve nut per DIN 3870, defining the required hose end geometry.

Non-conforming hose ends will not seal properly on the M24x1.5 pump connection.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, which limits fluid selection at low temperatures.

Too-viscous fluid at low temperature can overload the pump and reduce circulation.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The minimum filling capacity is 1.5 l and the expansion tank capacity is 2.8 l, which sets the minimum system fill and expansion allowance.

Underfilling can cause pump cavitation and poor temperature control.

Source
“min. filling capacity 1,5 l”
Installation

The unit has IP20 protection and must operate between 5 °C and 40 °C ambient, so it must be installed in a dry, protected location within that ambient range.

IP20 offers no water protection, so washdown or wet locations are unsuitable.

Source
“Degree of Protection IP20”