Rev n protector xstream laboratory fume hoods
Technical Reference Summary
Covered Models & Document Scope
Covers Labconco Protector XStream high-performance laboratory fume hoods in 4', 5', 6' and 8' widths, with standard vertical-rising sash and optional A-Style Combination Sash models. Eco Energy Saving Air Foil and Flush Air Foil variants are tabulated. Optional accessories covered include Guardian airflow monitors, service fixtures, duplex outlets, work surfaces, ceiling enclosures, rear panels, distillation grids and 8' dual exhaust adapter. Appendix D covers the Protector I-S Hood with Intelli-Sense Blower 3-speed operation.
Section-by-Section Technical Breakdown
The exhaust connection is designed for 12-inch nominal duct (12.75 in OD) to minimize static pressure loss at 50-100 fpm face velocity. Tables give volumetric airflow and hood static pressure for 4', 5', 6' and 8' hoods at face velocities from 40 to 100 fpm, at full-open sash (28 in) and 62.5% open sash (18 in), for both Eco Energy Saving and Flush air foils.
- 12 in nominal (12.75 in OD) exhaust connection; install a reducer after the outlet if transport velocities of 1000-2500 fpm are required.
- Airflow/static pressure tabulated at face velocities 40, 50, 60, 80 and 100 fpm.
- Two sash conditions: full open at 28 in and 62.5% open at 18 in.
- Separate tables for Eco Energy Saving Air Foil and Flush Air Foil.
- Blower sizing requires exhaust volume plus total system static pressure loss; contact Labconco Customer Service for assistance.
Hoods are factory-wired for 115 V, 50/60 Hz, 20 A or 230 V, 50/60 Hz, 10 A service; the correct voltage must be verified on the I.D. plate behind the front panel. All internal wiring for LED light, light switch, blower switch and duplex outlets terminates at a single-point junction box for connection by a licensed electrician.
- 115 V models: Phase (Hot) black, Neutral white. 230 V models: Phase 1 brown, Phase 2 blue.
- Building supply must include overload protection; a marked disconnecting switch or circuit breaker must be close to the equipment and within easy reach of the operator (NEC-2002).
- Dedicated branch circuits recommended; if circuits are combined, maximum load is the sum of individual outlets plus the unit rating (i.e. 2 Amps).
- Ground the hood to MAINS protective earth using a ring terminal for a 10-24 machine screw on the marked grounding lug; no other conductors on this lug.
- Grounding connection shall not be made to the terminal box cover; shielded conduit is usually required.
- LED light is mounted outside the top liner panel and sealed from hood vapors; bulb replacement requires removing the front panel and knockout plugs.
Service lines should be 1/4-inch OD copper (brass for natural gas) with an accessible shut-off valve. Recommended operating pressure is 40 PSI with a maximum allowable pressure of 200 PSI. Fixtures are pre-plumbed to hose connector or gooseneck; tubing may enter from above, the back or through the work surface.
- Recommended operating pressure 40 PSI; maximum allowable 200 PSI; consider a pressure regulator.
- Approved services: Air, Hot Water, Vacuum, Cold Water, Natural Gas.
- Do not use oxygen with any standard service fixture; contact Labconco for oxygen fixture information.
- Natural gas only in the fixture pre-plumbed with brass tubing; sulfur content can deteriorate standard copper lines.
- Inspect all fittings for leakage and tighten slightly if needed.
- All hoods are factory prepared for up to 8 service fixtures.
Hoods weigh 400-800 lbs (182-363 kg) and must be installed permanently and stationary on a base cabinet and chemically resistant work surface. The work surface must be at least as wide as the hood, flush with the back of the hood, and supported at the rear by a cross support installed after leveling the cabinets and before placing the hood.
- Do not install on a cart, dolly or mobile bench; all installations must be permanent and stationary.
- Do not lift by the front air foil; move on the shipping skid with a floor jack or furniture dolly, never by tilting onto a hand truck.
- Work surface should be 36 in deep for the high-performance XStream; Eco-Foil performs best with a 1.0 in leading radius.
- Work surface should be nonporous, chemical-resistant (e.g. epoxy resin) and have a dished recess for primary spills.
- Sash weight is individually matched to its hood and must not be exchanged with another unit.
- Seal the hood to the work surface with silicone sealant after ducting, wiring and plumbing.
The XStream uses a by-pass airflow design with aerodynamic sash handle, air foil, upper dilution air supply and Opti-Zone rear downflow dual baffle to produce horizontal airflow and reduce turbulence. The sash provides 37.5 in visibility while limiting travel to 28 in, with sash stops at 18 in to reduce energy use.
- Sash travel limited to 28 in operating height; sash stops at 18 in off the work surface.
- Interior height 48 in; removable primary and secondary baffles for cleaning.
- Energy-efficient LED T8 lighting rated 50,000 hours, using about half the power of fluorescent bulbs.
- Factory prepared for up to 8 service fixtures; duplex receptacles factory-wired to the single-point junction box.
- Right corner post factory prepared for Guardian Digital Airflow Monitor or Guardian Airflow Monitor (sold separately).
- Sash options: smooth anti-racking cable, chain, auto-return or Intelli-Sash.
Face velocity must be measured across the sash opening per the Industrial Ventilation Manual. Labconco recommends an average face velocity of 50-100 fpm for Protector XStream hoods. Factory ASHRAE 110-1995 testing achieves an as-manufactured rating of less than 0.05 ppm at 4 lpm.
- Recommended average face velocity at sash opening: 50-100 fpm.
- Factory tested per ASHRAE 110-1995; as-manufactured rating AM <0.05 ppm at 4 lpm.
- Field-use ASHRAE testing available through Labconco Sales Engineering or a certified on-site contractor.
- Performance at 50-60 fpm validated by Labconco under adverse conditions.
Chapter 6 provides a routine maintenance schedule and routine service operations for the fume hood. Service safety precautions apply before maintenance work.
- Follow the routine maintenance schedule in Chapter 6.
- Perform routine service operations as listed in Chapter 6.
- Observe service safety precautions before servicing.
Chapter 7 covers field installation of work surfaces, ceiling enclosures above the hood, rear panels behind the hood, additional service fixtures, Guardian airflow monitors, distillation grids, an electrical duplex outlet and the 8' dual exhaust adapter.
- Installing work surfaces.
- Installing ceiling enclosures above the fume hood.
- Installing rear panels behind the fume hood.
- Installing additional service fixtures and Guardian airflow monitors.
- Distillation grids - field installation.
- Installing an electrical duplex outlet and the 8' dual exhaust adapter.
Appendix B provides complete dimensional diagrams for all hood models. Appendix C provides electrical requirements, environmental conditions and wiring diagrams. Appendix G gives VAV open-area guidelines for programming variable air volume systems at various sash positions.
- Appendix B: Protector XStream Dimensions.
- Appendix C: Protector XStream Specifications, including electrical requirements and wiring diagrams.
- Appendix E: Serial Number Tag Description with current rating code.
- Appendix G: VAV Open Area Guidelines for variable air volume systems.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Aplica a |
|---|---|---|
| Exhaust connection is 12 in nominal pipe (12.75 in OD). p.10 ↗ | Size ductwork and blower for this outlet; install a reducer after the outlet if transport velocities of 1000-2500 fpm are required. | All Protector XStream models |
| Recommended average face velocity at the sash opening is 50-100 fpm. p.17 ↗ | Certify the installed hood to this range; re-certify if face velocity is improper. | All Protector XStream models |
| At 100 fpm with Eco air foil and sash full open at 28 in: 4' hood 705 CFM at 0.26 in w.c.; 5' hood 930 CFM at 0.32 in w.c.; 6' hood 1150 CFM at 0.41 in w.c.; 8' hood 1600 CFM at 0.29 in w.c. p.11 ↗ | Use these values with total system static pressure loss to select the remote blower. | XStream with Eco Energy Saving Air Foil |
| At 100 fpm with Eco air foil and sash at 62.5% open (18 in): 4' hood 440 CFM at 0.10 in w.c.; 5' hood 580 CFM at 0.12 in w.c.; 6' hood 720 CFM at 0.16 in w.c.; 8' hood 1000 CFM at 0.11 in w.c. p.11 ↗ | Sash stops at 18 in reduce required exhaust volume and energy use. | XStream with Eco Energy Saving Air Foil |
| At 100 fpm with Flush air foil and sash full open at 28 in: 4' hood 740 CFM at 0.28 in w.c.; 5' hood 975 CFM at 0.34 in w.c.; 6' hood 1205 CFM at 0.44 in w.c.; 8' hood 1675 CFM at 0.31 in w.c. p.11 ↗ | Flush air foil requires slightly higher airflow and static pressure than Eco air foil at the same face velocity. | XStream with Flush Air Foil |
| Electrical service: 115 V, 50/60 Hz, 20 A or 230 V, 50/60 Hz, 10 A. p.18 ↗ | Verify voltage on the I.D. plate behind the front panel before connecting supply. | All Protector XStream models |
| If circuits are wired together, maximum load allowed is the sum of individual outlets plus the rating of the unit (i.e. 2 Amps). p.18 ↗ | Dedicated branch circuits are recommended; combined circuits must respect this load limit. | All Protector XStream models |
| Wire color codes: 115 V Phase (Hot) black, Neutral white; 230 V Phase 1 brown, Phase 2 blue. p.18 ↗ | Connect MAINS supply conductors to hood supply wires per these codes. | All Protector XStream models |
| Grounding lug accepts a ring terminal sized for a 10-24 machine screw; only MAINS ground conductors may be connected to it. p.18 ↗ | Do not connect the grounding conductor to the terminal box cover. | All Protector XStream models |
| Service lines: 1/4 in OD copper (brass for natural gas) with accessible shut-off valve; recommended operating pressure 40 PSI, maximum allowable 200 PSI. p.12 ↗ | Consider a pressure regulator to reduce line pressure to 40 PSI; check local codes. | Hoods with service fixtures |
| Hood weight is 400 to 800 lbs (182-363 kg). p.13 ↗ | Supporting structure and work surface must safely support the combined weight of hood and related equipment; use mechanical lift or floor jack. | All Protector XStream models |
| Work surface should be 36 in deep for the high-performance XStream; Eco-Foil performs best with a 1.0 in leading radius. p.15 ↗ | Select a work surface with the correct depth and radius to promote aerodynamic airflow at low velocities. | XStream with Eco-Foil |
| Sash travel is limited to 28 in operating height; sash stops are located 18 in off the work surface. p.23 ↗ | Exhaust volume and blower sizing are based on the 18 in sash height to maximize energy savings. | All Protector XStream models |
| Factory ASHRAE 110-1995 as-manufactured rating: less than 0.05 ppm at 4 lpm (AM <0.05). p.17 ↗ | Use as baseline containment performance; field-use ASHRAE testing can be arranged through Labconco. | All Protector XStream models |
| LED T8 bulbs are rated 50,000 hours and use approximately half the power of fluorescent bulbs. p.24 ↗ | Lighting maintenance interval is long; replacement requires removing the front panel and knockout plugs. | All Protector XStream models |
| All hoods are factory prepared for up to 8 service fixtures. p.24 ↗ | Plan service fixture additions within this capacity. | All Protector XStream models |
| Do not use oxygen with any standard service fixture; natural gas only in the fixture pre-plumbed with brass tubing. p.20 ↗ | Contact Labconco for oxygen fixture information and to confirm compatibility of any other service. | Hoods with service fixtures |
| Lower baffle is 12.9 in tall and middle baffle is 27.4 in tall; baffles ship loose behind the hood. p.14 ↗ | Install baffles per Figure 1-1, Figure 3-1 and Figure B-1 before operating the hood. | All Protector XStream models |
Faults & Troubleshooting
| Código / Síntoma | Possible Cause | Remedy | Note |
|---|---|---|---|
| Remote blower and lights won’t operate p.43 ↗ | Wires not connected at junction boxes or switches. | Check connection of switches. Check connection to control box on top of unit. | Chapter 8 |
| Remote blower and lights won’t operate p.43 ↗ | Circuit breakers tripped in building electrical supply. | Reset circuit breakers. | Chapter 8 |
| Remote blower won’t operate, but lights work p.43 ↗ | Blower wiring is disconnected. Belt broken. | Inspect blower wiring and switch. Replace belt. | Chapter 8 |
| Remote blower won’t operate, but lights work p.43 ↗ | Blower motor is defective. | Replace blower motor. | Chapter 8 |
| Fume hood blower operates but lights will not operate p.43 ↗ | Lamp not installed correctly. | Inspect lamp installation. | Chapter 8 |
| Fume hood blower operates but lights will not operate p.43 ↗ | Lamp is defective. | Replace lamp. | Chapter 8 |
| Fume hood blower operates but lights will not operate p.43 ↗ | Lamp circuit breaker in building is tripped. | Reset the lamp circuit breaker. | Chapter 8 |
| Fume hood blower operates, but lights will not operate p.44 ↗ | Lamp wiring is disconnected. | Inspect lamp wiring. | Chapter 8 |
| Fume hood blower operates, but lights will not operate p.44 ↗ | Defective lamp ballasts. | Replace lamp ballasts. | Chapter 8 |
| Contaminants outside of fume hood p.44 ↗ | Improper user techniques for the fume hood. | See “Certifying the Hood” Chapter 3 and “Safety Precautions” Chapter 4 sections in the manual. (Ref. Appendix E) | Chapter 8 |
| Contaminants outside of fume hood p.44 ↗ | Restriction of the baffle air slots or – blockage of the exhaust outlet. | Remove baffles to ensure that all air slots, and the exhaust outlet are unobstructed. | Chapter 8 |
| Contaminants outside of fume hood p.44 ↗ | External factors are disrupting the fume hood airflow patterns or acting as a source of contamination. | See “Location Requirements” Chapter 2, “Certifying the Hood” Chapter 3, and “Safety Precautions” Chapter 4 sections of this manual. (Ref. Appendix E) | Chapter 8 |
| XStream Fume hood has improper face velocity. p.44 ↗ | Improper face velocity. | Have fume hood re-certified and check remote blower exhaust system. XStream Hood should have average face velocity of 50-100 fpm. | Chapter 8 |
| Vertical sash no longer operates smoothly p.44 ↗ | Cable is frayed or plastic protection is damaged. | Inspect cable and replace cable if worn or damaged immediately; otherwise injury could result. | Chapter 8 |
| Vertical sash no longer operates smoothly p.44 ↗ | Pulley bearing is damaged. | Replace pulley, bearing or add grease. | Chapter 8 |
| Vertical sash no longer operates smoothly p.44 ↗ | Cable or chain has slipped off the pulleys or sprockets. | Re-install, cable or chain must be replaced immediately if damaged. | Chapter 8 |
| Vertical sash no longer operates smoothly p.44 ↗ | Weight has broken pulleys or sprockets. | Replace pulleys or sprockets. | Chapter 8 |
| Combination A-Style Sash no longer operates smoothly p.44 ↗ | Horizontal glass panels have come off the tracks. | Re-install horizontal glass on tracks. | Chapter 8 |
| Combination A-Style Sash no longer operates smoothly p.45 ↗ | Vertical sash frame is distorted. | Place horizontal glass symmetrically and pull sash down to air foil. Straighten damaged frame. | Chapter 8 |
| Combination A-Style Sash no longer operates smoothly p.45 ↗ | Cable is frayed or has slipped off the pulleys. | Re-install, cable must be replaced immediately if damaged. | Chapter 8 |
| Electrical duplex outlets no longer have power p.45 ↗ | Wires not connected or faulty duplex. | Check wire connection or replace duplex. | Chapter 8 |
| Electrical duplex outlets no longer have power p.45 ↗ | Circuit breakers tripped in building electrical supply. | Reset circuit breakers. | Chapter 8 |
| Service valves no longer operate p.45 ↗ | Faulty building supply. | Inspect building supply shut off valves and appropriate pressures below the recommended 40 PSI. | Chapter 8 |
| Service valves no longer operate p.45 ↗ | Valve no longer operates. | Replace valve and check for leaks. | Chapter 8 |
| Service valves no longer operate p.45 ↗ | Supply line or outlet line has leaks. | Inspect line for leaks and fix any leaking plumbing connections. | Chapter 8 |