c190-e292

Technical Reference Summary

Covered Models & Document Scope

Sistema HPLC inerte Nexera lite; el documento también describe los módulos del sistema bomba LC-40i, SCL-40, CBM-40, CBM-40lite, pHM-40 y válvulas de conmutación de línea de flujo

Technical Data

Parameter Value Condición
Flow rate 0.5 mL/min p.3 ↗ analytical conditions, gel filtration
Column temperature 25 °C p.3 ↗ analytical conditions
Detection wavelength 280 nm p.3 ↗ UV detection
Retention time repeatability 0.08 % RSD p.3 ↗ n=5 injections
Number of degassed solvents 3 p.8 ↗ SCL-40
Number of degassed solvents 5 p.8 ↗ CBM-40
Analog board channels Up to 2 channels (option) p.8 ↗ SCL-40
Analog board channels Up to 1 channel p.8 ↗ CBM-40
Injection volume accuracy ≤ ±1 % p.9 ↗ µL injection
Temperature control range Room temperature −10 to 100 °C p.9 ↗ autosampler, standard equipment
Temperature control range Room temperature −10 to 85 °C p.9 ↗ autosampler without sample cooler
Temperature accuracy ± 0.5 °C p.9 ↗ standard equipment
Temperature accuracy ± 0.8 °C p.9 ↗ without sample cooler
Temperature precision ± 0.05 °C p.9 ↗ standard equipment
Temperature precision ± 0.1 °C p.9 ↗ without sample cooler
Containable column size and number Up to 250 mm L. column × 6 or 300 mm L. column × 3 mm p.9 ↗ standard equipment
Containable column size and number Up to 100 mm L. column × 6 or 300 mm L. column × 3 mm p.9 ↗ without sample cooler
Dimensions W 260 × D 500 × H 415 mm p.9 ↗ autosampler, standard equipment
Weight 27 kg p.9 ↗ autosampler, standard equipment
Dimensions W 260 × D 509 × H 415 mm p.9 ↗ autosampler without sample cooler
Weight 30 kg p.9 ↗ autosampler without sample cooler
Wavelength range 190 to 700 nm p.9 ↗ UV detector, deuterium lamp
Wavelength range 190 to 1000 nm p.9 ↗ UV detector, deuterium and tungsten lamps
Wavelength range 200 to 650 nm p.10 ↗ fluorescence detector, xenon lamp
Cell temperature setting range 4 to 40 in 1 °C steps °C p.10 ↗ fluorescence detector, low-pressure mercury lamp

Operating Limits

  • Temperature control range (autosampler, standard equipment): Room temperature −10 °C to 100 °C p.9 ↗
  • Temperature control range (autosampler, without sample cooler): Room temperature −10 °C to 85 °C p.9 ↗
  • Cell temperature setting range (fluorescence detector): 4 to 40 °C in 1 °C steps p.10 ↗
  • Cell temperature control range (fluorescence detector): 10 °C below R.T. to 40 °C p.10 ↗