Development of the Micro-ESI 8060 Micro-Flow Ionization Unit To Provide Both High Sensitivity and High Robustness
Technical Reference Summary
Covered Models & Document Scope
Informe técnico C146-E354 sobre la unidad de ionización de microflujo Micro-ESI 8060 / interfaz de ionización para sistemas de micro-LC/MS (Shimadzu Nexera Mikros). No se imprimen códigos de pedido ni variantes de tamaño.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| LC flow rate range | a few to a few tens µL/min p.1 ↗ | micro-LC/MS operation |
| Test flow rate (low) | 2 µL/min p.3 ↗ | sensitivity/robustness evaluation |
| Test flow rate (high) | 400 µL/min p.3 ↗ | sensitivity/robustness evaluation |
| Ion intensity (500 pg/mL) | 5,000 cps p.3 ↗ | 2 µL/min |
| Ion intensity (500 pg/mL) | 1,500 cps p.3 ↗ | 400 µL/min |
| Ion intensity (50 pg/mL) | 10,000 cps p.3 ↗ | 2 µL/min |
| Ion intensity (50 pg/mL) | 500 cps p.3 ↗ | 400 µL/min |
| Retention time window | 12.75–16.5 min p.3 ↗ | 500 pg/mL / 50 pg/mL runs |
| Retention time window | 11.5–13.5 min p.3 ↗ | LTD4, 50 pg/mL, 5 µL injection |