c190-e291
Technical Reference Summary
Covered Models & Document Scope
Sistemas de cromatografía de fluidos supercríticos analítica/preparativa y de extracción con fluidos supercríticos Nexera UC y Nexera UC Prep. El documento abarca el controlador de sistema CBM-40, las bombas de suministro de disolvente LC-40D, LC-40D XR, LC-40D XS, LC-40D X3, LC-20AR, LC-20AP, la unidad de desgasificación DGU-403, el automuestreador SIL-40C XR y los detectores SPD-40, SPD-40V, SPD-M40.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Allowable maximum pressure | 44 MPa p.2 ↗ | LC-40D |
| Allowable maximum pressure | 70 MPa p.2 ↗ | LC-40D XR |
| Allowable maximum pressure | 105 MPa p.2 ↗ | LC-40D XS |
| Allowable maximum pressure | 130 MPa p.2 ↗ | LC-40D X3 |
| Flow rate settings range | 0.0001 to 5.0000 mL/min p.2 ↗ | LC-40D, 1.0 to 44 MPa |
| Flow rate settings range | 5.0001 to 10.0000 mL/min p.2 ↗ | LC-40D, 1.0 to 22 MPa |
| Flow rate accuracy | ≤ ± 1% or ± 2 μL/min, whichever greater p.2 ↗ | under specified conditions |
| Flow rate precision | ≤ 0.06% RSD or 0.02 minSD, whichever greater p.2 ↗ | |
| Gradient range of set concentrations | 0 to 100 % p.2 ↗ | 0.1% step |
| Gradient concentration accuracy | ± 0.5 % p.2 ↗ | under specified conditions |
| Available pH range | 1 to 14 p.2 ↗ | |
| Weight | 10 kg p.2 ↗ | LC-40D |
| Weight | 12 kg p.2 ↗ | LC-40D XR |
| Dimensions | W 260 × D 500 × H 140 mm p.2 ↗ | LC-40D |
| Operating temperature range | 4 to 35 °C p.2 ↗ | LC-40D series |
| Power supply | 100 to 240 V AC, 150 VA, 50/60 Hz p.2 ↗ | LC-40D series |
| Plunger capacity | 47 μL p.2 ↗ | LC-20AR |
| Flow rate setting range | 0.001 to 5 mL/min p.2 ↗ | LC-20AR, 1.0 to 50 MPa |
| Flow rate setting range | 5.001 to 20 mL/min p.2 ↗ | LC-20AR, 1.0 to 25 MPa |
| Flow rate accuracy | No more than ±1% or ±10 μL/min, whichever is greater p.2 ↗ | LC-20AR, 0.1 to 5.0 mL/min |
| Flow rate precision | No more than 0.08% RSD or 0.02 min SD, whichever is greater p.2 ↗ | LC-20AR |
| Number of degassed solvents | 3 p.3 ↗ | DGU-403 |
| Degassed flow line capacity | 400 µL/1 line p.3 ↗ | DGU-403 |
| Injection volume accuracy | ≤ ± 1 % p.3 ↗ | SIL-40C XR, 5 µL injection, n = 20 |
| Wavelength range | 190 to 700 nm p.4 ↗ | SPD-40 |
Operating Limits
- Operating temperature range: 4 to 35°C p.2 ↗
- Allowable maximum pressure: 130 MPa p.2 ↗
- Available pH range: 1 to 14 p.2 ↗