Especificaciones del cromatógrafo líquido de ultra alta eficacia Nexera X4 (C196-E103A): controlador del sistema SCL-40, CBM-40, CBM-40lite; unidad de suministro de disolvente LC-40B X4; unidad de conmutación y desgasificación de disolvente SSU-40; automuestreador SIL-40C X4; detector SPD-M40 X4

Technical Data

Parameter Value Condición
Number of controllable units 8 p.3 ↗ SCL-40
Number of controllable units 8 (Using option: 12) p.3 ↗ CBM-40
Number of controllable units 4 (excluding built-in solvent delivery unit) p.3 ↗ CBM-40lite
Event input/output Input: 1, Output: 2 p.3 ↗ SCL-40
Analog signal input Up to two channels (option) p.3 ↗ SCL-40
Analog signal input One channel (option) p.3 ↗ CBM-40
Communication interface Ethernet p.3 ↗ SCL-40
Reservoir tray Included as standard p.3 ↗ SCL-40
Dimensions and weight W260 × D500 × H140 mm, 6 kg p.3 ↗ SCL-40
Dimensions and weight W260 × D500 × H72 mm, 5 kg p.3 ↗ CBM-40
Operating temperature range 4 to 35 °C p.3 ↗ SCL-40 / CBM-40
Power supply 100 to 240 V AC, 50 VA, 50/60 Hz p.3 ↗ SCL-40 / CBM-40
Pumping method Micro-volume double plunger pump p.3 ↗ LC-40B X4
Allowable maximum pressure 130 MPa p.3 ↗ LC-40B X4
Flow rate settings range 0.0001 to 2.0000 mL/min p.3 ↗ LC-40B X4
Flow rate accuracy ±1 % (at 1 mL/min, 80 MPa, with water, stable within a temperature range of 20 to 30 °C) p.3 ↗ LC-40B X4
Gradient mode Two-solvent high-pressure gradient p.3 ↗ LC-40B X4
Range of gradient mixing ratio 0 to 100 % (0.1 % step) p.3 ↗ LC-40B X4
Available pH range 1 to 14 p.3 ↗ LC-40B X4
Dimensions and weight W260 × D500 × H210 mm (excluding protrusions), 24 kg p.3 ↗ LC-40B X4
Power supply 100 to 240 V AC, 50/60 Hz, 150 VA p.3 ↗ LC-40B X4
Number of selectable solvents Up to 4 solvents per pump (maximum 2 pumps) p.3 ↗ SSU-40
Number of degassed solvent lines 5 p.3 ↗ SSU-40
Degassed flow line capacity 400 μL/1 line p.3 ↗ SSU-40
Dimensions and weight W260 × D500 × H72 mm, 5 kg p.3 ↗ SSU-40

Operating Limits

  • Operating temperature range: 4 to 35 °C p.3 ↗
  • Allowable maximum pressure: 130 MPa p.3 ↗
  • Available pH range: 1 to 14 p.3 ↗

Ordering Data

Ordering Code Part No. Variant / Code Aplica a Note
Nexlock II LD p.7 ↗ 228-87372-41 ID 0.1 × 600 mm

Modelos relacionados