Long Term Robustness of PFAS Analysis in Complex Matrices
Contenido
Overviewp. 2
Background and purpose of the PFAS robustness study
pp. 3-6
Peak area responses, MRM chromatograms, and matrix injection results
p. 7
Summary of long-term robustness findings for PFAS analysis
p. 2table, 9 rows
p. 3table, 6 rows
Table 1. Agilent 1290 Infinity II LC System method parameters.
Parameter
Value
Columns
Agilent ZORBAX RRHD Eclipse Plus C18, 2.1 × 30 mm, 1.8 μm
Agilent ZORBAX RRHD Eclipse Plus C18, 2.1 × 5 mm, 1.8 μm
Agilent InfinityLab PFC Delay Column, 4.6 × 30 mm
Sampler Temperature
4 °C
Mobile Phase A
Water + 2 mM NH4-acetate
Mobile Phase B
ACN
Flow Rate
0.8 mL/min
Injection Volume
2 μL
Column Temperature
45 °C
Gradient Program
Time (min) %B
0.0 20
1.0 50
1.5 90
2.5 90
2.51 20
Stop Time
3.0 min
Table 3. Agilent 6495D LC/TQ MRM parameters per PFAS.
Compound
Precursor m/z
Product m/z
IFunnel Mode
CAV (V)
CE (V)
Polarity
Measured Retention Time (min)
PFHxS
398.9
99.0
Large Mol.
5
43
Neg.
1.44
PFOA
413.0
369.0
Fragile
5
7
Neg.
1.39
PFOA-13C8
421.0
376.0
Fragile
5
7
Neg.
1.39
PFNA
463.0
419.0
Fragile
5
8
Neg.
1.52
PFOS
498.9
80.0
Large Mol.
5
52
Neg.
1.67
PFOS-13C8
507.0
80.0
Large Mol.
5
52
Neg.
1.67
Engineering Facts
12 facts · every claim linked to a page
Performance
The system sustained quantitative precision with peak area RSDs of ≤7.5% across all analytes without internal standard correction over 13,700 injections.
Demonstrates long-term signal stability that reduces the need for frequent recalibration in high-throughput PFAS workflows.
Source
“peak area %RSDs of ≤ 7.5% across all analytes without internal standard correction”
Operation
The instrument ran approximately three weeks of continuous operation with no maintenance or mass spectrometer calibration.
Indicates expected uptime and service intervals when planning lab capacity and maintenance schedules.
Source
“13,700 matrix injections were carried out over approximately three weeks of continuous operation with no maintenance or mass spectrometer calibration”
Performance
Even after exposure to more than 137 mL of salmon extract with visible matrix deposition on the inlet, no significant performance loss occurred.
Shows the source tolerates heavy matrix loading, informing decisions on sample cleanup requirements and maintenance frequency.
Source
“the instrument was exposed to more than 137 mL of salmon extract, which resulted in significant matrix deposition on the inlet of the mass spectrometer”
Maintenance
Mass axis and peak width calibrations were checked every ~2,500–3,000 matrix injections and stayed within manufacturer limits without retuning.
Defines a practical calibration verification interval for method validation and routine QC planning.
Source
“checked at regular intervals (after approximately every 2,500 to 3,000 matrix injections)”
Maintenance
No tune parameters needed readjustment and no significant contamination of inner ion optics was observed after the test.
Supports lower expected downtime and simpler service procedures for ion optics.
Source
“without need to readjust any tune parameters”
Maintenance
Routine maintenance of the ion source and optics is still recommended despite no observed performance loss.
Establishes that preventive maintenance should remain in the operating plan even with robust performance.
Source
“routine maintenance of ion source and optics is recommended to ensure prolonged instrument operation”
Installation
A dedicated PFC delay column is part of the LC configuration to manage PFAS background contamination.
Requires planning for an additional column in the flow path during installation to control PFAS background.