According to standard DIN 28 428, the base pressure is that measured in a leak-free test dome, 48 hours after the completion of test dome bake-out, with a Turbopump fitted with a ConFlat flange and using the recommended pre-vacuum pump.
Max gas throughput at full rotational speed (with recommended forepump) Air cooling/Water cooling
170 sccm
110 sccm
* Pumping speed measured with fluximeter.
Tab. 1
CONTROLLER PART NUMBER
PRODUCT NAME
Controller AG Rack base
X3506-64001
TwisTorr 304 FS AG Rack
Controller AG Rack with RS232/485
X3506-64002
TwisTorr 304 FS AG Rack RS232/485
Controller AG Rack with RS232/485 & Profibus
X3506-64003
TwisTorr 304 FS AG Rack Profibus
Controller V304 onboard 24Vdc
X3507-64002
TwisTorr 304 FS onboard 24V
Controller V304 onboard 100-240Vac
X3507-64003
TwisTorr 304 FS onboard 110/220V
Technical Specification
Parameter
Value
Operating position
Any
Operating ambient temperature
+5 °C to +35 °C
Relative humidity of air
0 – 90 % (not condensing)
Bakeout temperature
80 °C at inlet flange max.
Lubricant
Permanent lubrication
Cooling requirements
Natural convection (only with no gas load)
Forced air (5- 35 °C ambient temperature)
Water (mandatory if ambient temperature > 35 °C)
Coolant water
Minimum flow: 50 l/h (0.89 GPM)
Temperature: +15 °C to +30 °C
Pressure: 3 to 5 bar (45 to 75 psi)
Noise Pressure level
< 50 dB(A) at 1 meter
Pwr supply (mains):
Max input pwr:
300 VA
Stand-by pwr:
10 W
Max oper. power:
150 W with water or air cooling
Installation category
II
Pollution degree
2
Storage temperature
-40° C to +70° C
Max altitude
3000 m
Weight kg (lbs):
Pump ISO 100
5.5 kg (12.3)
Pump CFF 6’’
7.5 kg (16.5)
Pump ISO 160
5.7 kg (12.6)
Pump CFF 8’’
9.7 kg (20.9)
Tab. 5 Accessories and spare parts
DESCRIPTION
PART NUMBER
Mains cable NEMA Plug, 3m long
9699958
Mains cable European Plug, 3m long
9699957
Serial cable and Navigator Software
9699883
Inlet screen, ISO 100
X3500-68000
Inlet screen CFF 6’’
9699302
Inlet screen, ISO 160
X3500-68001
Inlet screen CFF 8’’
9699304
Water cooling kit
9699337
Plastic water cooling kit
9699347
Air cooling kit for On-board Controller
X3500-68010
Air cooling kit for Rack AG Controller
X3500-68011
Fan extension cable
9699949
Bracket for On-Board Controller side mounting
X3500-68012
Vibration isolator, ISO 100
9699344
Vibration isolator, CF 6"
9699334
Vibration isolator, ISO 160
9699345
Vibration isolator, CF 8"
9699335
Vent flange, NW 10 KF / M8
9699108
Vent valve for AG Rack Controller
9699844
Vent valve for on-board Controller
9699834
Purge valve 10 SCCM NW16KF – M12
9699239
Purge valve 10 SCCM ¼ Swagelok – M12
9699240
Purge valve 20 SCCM NW16KF – M12
9699241
Purge valve 20 SCCM ¼ Swagelok – M12
9699242
Purge valve 10 SCCM ¼ Swagelok – ¼ Swagelok
9699232
Purge valve 20 SCCM ¼ Swagelok – ¼ Swagelok
9699236
TwisTorr 304FS foreline flange KF25
X3500-68002
Active Gauges
Ask Agilent for details
Tab. 2
FLANGE
FIXING DEVICE
N.
FIXING TORQUE
ISO 100 K
M10 clamps
4
22 Nm
ISO 160 K
M10 clamps
4
22 Nm
Reduction in pumping speed (%)
Ar
N2
He
H2
X3500-68000
20%
18%
7%
5%
X3500-68001
Engineering Facts
14 facts · every claim linked to a page
Installation
The pump must be connected to a suitable Agilent controller and to a primary pump; it cannot be run standalone.
Determines controller selection and requires a backing pump in the system design.
Source
“Die Turbomolekularpumpen TwisTorr 304 FS dürfen nur mit einem der passenden Agilent-Controller verwendet werden und müssen an eine Primärpumpe angeschlossen werden”
Cumplimiento
The pump is installation category II (overvoltage) per EN 61010-1, so the mains supply must meet that category.
Affects electrical supply design and compliance of the installation.
Source
“Der TwisTorr 304 FS gehört zur zweiten Installationsklasse (Überdruck) die von den Normen EN 61010-1 vorgesehen ist.”
Operation
During bakeout the inlet flange must not exceed 120 °C with ConFlat or 80 °C with ISO flanges, and water cooling must always be used.
Sets bakeout limits and mandates water cooling to avoid overheating the pump.
Source
“la temperatura sulla flangia di ingresso non deve essere superiore a 120 °C con flangiatura ConFlat e a 80 °C con flangiatura ISO. Usare sempre il raffreddamento ad acqua durante le operazioni di riscaldamento.”
Safety
Never operate the pump unless the inlet flange is connected to the vacuum chamber or closed with the blank flange.
Prevents rotor damage and hazardous operation during commissioning.
Source
“Non far funzionare mai la pompa se la flangia di ingresso non è collegata alla camera a vuoto o non è chiusa con la flangia di chiusura.”
Operation
Vent pressure at the vent port must be below 2 bar above atmospheric, using dust-free air or inert gas.
Protects bearings and defines venting hardware requirements.
Source
“La pressione di ingresso attraverso l'apposita porta deve essere inferiore a 2 bar (oltre la pressione atmosferica).”
Operation
For gases with particulates or aggressive pollutants, an inert gas purge (N2, He, Ar) must be supplied through the integrated purge/vent port to protect bearings.
Requires purge gas supply and plumbing when pumping dirty or aggressive processes.
Source
“queste pompe sono dotate di una apposita porta (di purge/vent integrato) attraverso la quale è necessario fornire alla pompa un flusso di gas inerte (Azoto, Elio o Argon) per proteggere i cuscinetti della pompa”
Safety
Never use the pump with corrosive gases or vapors that could damage internal materials.
Excludes corrosive processes unless a compatible pump version is selected.
Source
“Non usare mai la pompa in presenza di gas o vapori corrosivi che possano danneggiare i materiali interni alla pompa.”
Installation
The pump can be installed in any position but must be fixed at the inlet flange to a counter-flange able to withstand 1000 Nm torque around its axis.
Defines mounting orientation freedom and the mechanical strength required of the chamber flange.
Source
“Fix the TwisTorr 304 FS in a stable position, mounting the inlet flange of the turbopump to the system counter-flange, with a connection capable of withstanding a torque of 1000 Nm around its axis.”
Installation
ISO-K flanges use 4 M10 high-strength clamps (e.g., Agilent IC63250DCMZ) tightened to 22 Nm.
Specifies clamp count and torque for leak-free, safe mounting.
Source
“ISO 100 K M10 clamps 4 22 Nm”
Installation
The pump cannot be fixed by means of its base; only the inlet flange mounting is allowed.
Prevents an invalid mounting approach that could lead to detachment.
Source
“The TwisTorr 304 FS cannot be fixed by means of its base.”
Operation
The pump must always be cooled before starting; running without cooling can damage internal components.
Requires cooling provision (water or forced air) in the installation design.
Source
“The pump must always be cooled before starting using it. Running the pump without cooling it could damage inside components.”
Operation
A thermistor near the upper bearing stops the pump with Error Code 2 above 60 °C and Error Code 1 below 0 °C.
Explains thermal trip behavior and cooling/ambient requirements.
Source
“If the thermistor reading exceeds 60 °C the pump fails and stops with Error Code = 2. If the thermistor reading is below 0 °C, the pump fails and stops with Error Code = 1”
Maintenance
Before any work, disconnect mains, vent the pump, wait for the rotor to stop and surface temperature below 50 °C.
Defines safe lockout procedure for service and maintenance.
Source
“Alvorens werkzaamheden aan het systeem uit te voeren, de voeding loskoppelen, de pomp met behulp van de hiervoor bestemde klep ontluchten en wachten totdat de rotor volledig stil staat en de oppervlaktetemperatuur van de pomp onder een temperatuur van 50 °C is gezakt.”
Operation
The pump is maintenance-free; any intervention must be performed by authorized personnel.
Sets service strategy: no field maintenance, only authorized repair/exchange.
Source
“De pompen TwisTorr 304 FS zijn onderhoudsvrij. Eventuele werkzaamheden moeten door bevoegd personeel worden uitgevoerd.”