The pump's titanium-based flow path tolerates high salt and extreme pH, so it suits biological and high-salt methods.
Determines whether the system can run your salt/pH method without corrosion damage.
Source
“The titanium-based pump offers highest corrosion resistance against high salt concentration (2 M) and offers a handling of a wide pH range (1 – 13 , short term 14 ).”
Selection
The pump reaches 600 bar and 10 mL/min at 200 bar, enabling conventional, sub-2 µm and superficially porous columns.
Sets the column particle size and flow rate you can use.
Source
“It features a pressure range of up to 600 bar and a flow rate up to 10 mL/min (at 200 bar).”
Selection
Standard PEEK capillaries are pressure-sensitive, so bio-inert systems use stainless-steel cladded PEEK rated to at least 600 bar.
Prevents selecting plain PEEK tubing that would fail at system pressure.
Source
“Agilent uses stainless-steel cladded PEEK capillaries (PK/ST), which keep the flow path free of metal as required for bio-inert applications and ensure pressure stability to at least 600 bar.”
Installation
UHP-FF fittings must be tightened to a maximum torque of 0.8 Nm using the fitting tool or a 5 mm hex wrench.
Over-torque damages the fitting and capillary during assembly.
Source
“Use 5043-0915 (Fitting mounting tool) or a 5 mm hex wrench for fixing the fitting (maximum torque 0.8 Nm).”
Maintenance
Fittings must not be removed from used bio-inert capillaries because the expanded PEEK front end can be ripped off.
Removing fittings destroys the capillary and forces replacement.
Source
“Do not remove fittings from used capillaries.”
Installation
The bio-inert ZDV union has two different ends and capillaries must be installed in the marked sequence to avoid leaks.
Wrong assembly order damages the union insert and causes leaks.
Source
“The 5067-4741 (ZDV union (Bio-inert)) has two different connectors where capillaries need to be installed in the correct sequence.”
Safety
The solvent cabinet holds a maximum of 8 L total and no single bottle may exceed 2 L.
Defines safe solvent storage limits for the bench setup.
Source
“The solvent cabinet is designed to store a maximum volume of 8 L solvent. The maximum volume for an individual bottle stored in the solvent cabinet should not exceed 2 L.”
Safety
The leak sensor in the leak plane stops the running system once the leak detection level is reached.
Explains the automatic shutdown behavior relied on for safe operation.
Source
“The leak sensor in the leak plane stops the running system as soon as the leak detection level is reached.”
Installation
Drain connectors must be fitted on every module except the lowest one, where a waste tube is connected instead.
Ensures low-flow high-viscosity leaks drain properly through the stack.
Source
“Make sure that dripping adapters are correctly installed on each module in the LC stack, excluding lowest module.”
Installation
Stacking modules in the prescribed order optimizes the flow path for minimum delay volume.
Affects gradient delay and bench footprint at installation.
Source
“This configuration optimizes the flow path for minimum delay volume and minimizes the bench space required.”
Operation
The manual injector offers a standard 20 µL injection volume with optional loops from 5 µL to 5 mL.
Determines whether manual injection volumes match your method needs.
Source
“It offers a standard injection volume of 20 µL (optional 5 µL to 5 mL) and ensures highest injection accuracy.”
Operation
The detector covers 190–950 nm with a 120 Hz acquisition rate suited to fast LC.
Confirms detector compatibility with fast, high-throughput methods.
Source
“With its 120 Hz data acquisition rate the detector is perfectly suited for fast LC applications.”
Cumplimiento
The instrument may only be operated by appropriately qualified and trained personnel.
Sets the personnel qualification requirement before commissioning.
Source
“The instrument may only be operated by appropriately qualified and trained personnel.”